Evaluation of Delivery Analysis to Detect Intrafractional Motion during Tomotherapy

K. Tarutani, M. Tanooka, K. Sano, Okada Wataru, Masayuki Fujiwara, K. Yamakado
{"title":"Evaluation of Delivery Analysis to Detect Intrafractional Motion during Tomotherapy","authors":"K. Tarutani, M. Tanooka, K. Sano, Okada Wataru, Masayuki Fujiwara, K. Yamakado","doi":"10.4236/ijmpcero.2019.84020","DOIUrl":null,"url":null,"abstract":"The purpose of this study was to investigate the ability of a management system (Delivery Analysis: DA) to detect intrafractional motion during intensity-modulated radiation therapy (IMRT) in tomotherapy mode. Tomotherapy has made it possible to manage internal movements during treatment using software DA, which quantifies using the information of the passing dose obtained during the radiation treatment of patients. First, three treatment plans for the test were created (lumbar spine, prostate, and femur). Second, a pelvis phantom was moved in the X, Y, and Z directions, and a sinogram was acquired. The magnitudes of the movements were 3 mm, 5 mm, and 10 mm, respectively. Finally, the ability of DA to detect the motion was evaluated by comparing the sinogram obtained by moving the phantom with a reference sinogram obtained without movement. The sensitivity of DA could be detected with a shift amount of 3 mm (gamma analysis tolerance 0.3 mm/0.3%). The average gamma analysis of each direction at 0.3 mm/0.3% tolerance at each treatment site was 96.1% for the prostate, 93.5% for the lumbar spine, and 94.4% for the femur. Additionally, the average gamma pass rate results for the pelvic phantom in the X, Y, Z directions for a 10 mm shift were 96.2%, 96.3%, and 95.9%, respectively. DA is a powerful tool with high detection sensitivity and ability to detect body movement during treatment.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2019-11-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/ijmpcero.2019.84020","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The purpose of this study was to investigate the ability of a management system (Delivery Analysis: DA) to detect intrafractional motion during intensity-modulated radiation therapy (IMRT) in tomotherapy mode. Tomotherapy has made it possible to manage internal movements during treatment using software DA, which quantifies using the information of the passing dose obtained during the radiation treatment of patients. First, three treatment plans for the test were created (lumbar spine, prostate, and femur). Second, a pelvis phantom was moved in the X, Y, and Z directions, and a sinogram was acquired. The magnitudes of the movements were 3 mm, 5 mm, and 10 mm, respectively. Finally, the ability of DA to detect the motion was evaluated by comparing the sinogram obtained by moving the phantom with a reference sinogram obtained without movement. The sensitivity of DA could be detected with a shift amount of 3 mm (gamma analysis tolerance 0.3 mm/0.3%). The average gamma analysis of each direction at 0.3 mm/0.3% tolerance at each treatment site was 96.1% for the prostate, 93.5% for the lumbar spine, and 94.4% for the femur. Additionally, the average gamma pass rate results for the pelvic phantom in the X, Y, Z directions for a 10 mm shift were 96.2%, 96.3%, and 95.9%, respectively. DA is a powerful tool with high detection sensitivity and ability to detect body movement during treatment.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
放射分析在断层治疗中检测病灶内运动的评价
本研究的目的是研究一种管理系统(传递分析:DA)在扫描治疗模式下检测调强放疗(IMRT)期间的病灶内运动的能力。使用DA软件,断层治疗使得在治疗期间管理内部运动成为可能,该软件使用患者放射治疗期间获得的通过剂量信息进行量化。首先,制定了三种治疗方案(腰椎、前列腺和股骨)。其次,在X, Y和Z方向移动骨盆幻像,并获得sinogram。运动幅度分别为3 mm、5 mm和10 mm。最后,通过比较运动幻影获得的正弦图与不运动时获得的参考正弦图来评估DA检测运动的能力。当位移量为3 mm时,可以检测到DA的灵敏度(伽马分析公差0.3 mm/0.3%)。在每个治疗部位0.3 mm/0.3%容忍度下,每个方向的平均伽马分析在前列腺为96.1%,腰椎为93.5%,股骨为94.4%。此外,骨盆幻像在X、Y、Z方向移位10毫米时的平均伽马通过率分别为96.2%、96.3%和95.9%。DA是一种强大的工具,具有很高的检测灵敏度和检测治疗过程中身体运动的能力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Assessment of Monitor Units and Gamma Pass Rate for 6 MV and Flattening Filter Free (FFF) Beams in Volumetric Modulated Arc Therapy (VMAT) The Journey of SRS: An Initial Experience of Starting Sterotactic Radiosurgery Facility at Purbanchal Cancer Hospital—First in Nepal Hereditary Multiple Exostoses (HME) with Peroneal Nerve Compresion: A Case Report Dosimetric Study of the Effect of Calypso-Compatible Couch Top for Spine Stereotactic Body Radiation Therapy Radiation-Associated Cardiotoxicity during Breast Cancer Treatment with Ionizing Radiation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1