Yun Alelyunas, Gregory T Roman, Jay S. Johnson, C. Doneanu, M. Wrona
{"title":"High throughput analysis at microscale:performance of ionKey/MS with Xevo G2-XS QTof under rapid gradient conditions","authors":"Yun Alelyunas, Gregory T Roman, Jay S. Johnson, C. Doneanu, M. Wrona","doi":"10.17145/JAB.15.021","DOIUrl":null,"url":null,"abstract":"In this paper, high throughput analysis with 3 minute, rapid gradient conditions is described using the ionKey/MS™ System with an integrated ACQUITY UPLC® M-Class System and the Xevo® G2-XS QTof Mass Spectrometer. Extensive testing of representative small molecules and a peptide shows that the system is well-tolerated and exhibits excellent reproducibility and linear response. The iKey™ HSS T3 Separation Device used is robust, withstanding ~2200 injections of prepared human plasma with excellent peak shape and system pressure profile. A 99% solvent savings was realized when compared with an analytical system using a 2.1 mm column with flow rate ranging from 0.6 mL/min to 1.5 mL/min. These data, coupled with examples from the literature, illustrate that the ionKey/MS System with Xevo G2-XS QTof can be used as a full service platform for high throughput analysis and high sensitivity analysis to support all phases of drug discovery and development.","PeriodicalId":15014,"journal":{"name":"Journal of Applied Bioanalysis","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2015-10-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Bioanalysis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17145/JAB.15.021","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
In this paper, high throughput analysis with 3 minute, rapid gradient conditions is described using the ionKey/MS™ System with an integrated ACQUITY UPLC® M-Class System and the Xevo® G2-XS QTof Mass Spectrometer. Extensive testing of representative small molecules and a peptide shows that the system is well-tolerated and exhibits excellent reproducibility and linear response. The iKey™ HSS T3 Separation Device used is robust, withstanding ~2200 injections of prepared human plasma with excellent peak shape and system pressure profile. A 99% solvent savings was realized when compared with an analytical system using a 2.1 mm column with flow rate ranging from 0.6 mL/min to 1.5 mL/min. These data, coupled with examples from the literature, illustrate that the ionKey/MS System with Xevo G2-XS QTof can be used as a full service platform for high throughput analysis and high sensitivity analysis to support all phases of drug discovery and development.