{"title":"Phase calibration of oscilloscopes at low frequencies","authors":"Joo-Gwang Lee, Chihyun Cho","doi":"10.1109/IRMMW-THZ.2015.7327761","DOIUrl":null,"url":null,"abstract":"This paper describes how to calibrate phase response of oscilloscopes, especially at low frequencies from 10 MHz to 1 GHz. The method uses the phase response of a standard mixer calibrated by three-mixer method as the reference phase. We describe basic principles and show results obtained using a balanced mixer and a real-time oscilloscope (1 GHz BW).","PeriodicalId":6577,"journal":{"name":"2015 40th International Conference on Infrared, Millimeter, and Terahertz waves (IRMMW-THz)","volume":"22 1","pages":"1-2"},"PeriodicalIF":0.0000,"publicationDate":"2015-11-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 40th International Conference on Infrared, Millimeter, and Terahertz waves (IRMMW-THz)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IRMMW-THZ.2015.7327761","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
This paper describes how to calibrate phase response of oscilloscopes, especially at low frequencies from 10 MHz to 1 GHz. The method uses the phase response of a standard mixer calibrated by three-mixer method as the reference phase. We describe basic principles and show results obtained using a balanced mixer and a real-time oscilloscope (1 GHz BW).