Backscattered electron effects in a high-angle EDXS

A. Wilson, L. Lambrianidis
{"title":"Backscattered electron effects in a high-angle EDXS","authors":"A. Wilson, L. Lambrianidis","doi":"10.1088/0022-3735/22/9/009","DOIUrl":null,"url":null,"abstract":"The performance of a high take-off angle energy dispersive X-ray spectrometer (EDXS) on a 200 keV STEM has been investigated. A simple modification to the multichannel analyser (MCA) was used to extend the energy range beyond 200 keV to confirm that high energy, backscattered electrons were entering the detector. Up to 50% of the total count rate was observed to be due to these electrons. The spectrometer performance was determined both with and without a carbon filter which was used to stop the backscattered electrons entering the detector. It is shown that these electrons reduce the resolution and increase the dead-time and background noise in the X-ray spectra. The performance of the high take-off angle detector is compared with that of a horizontally mounted detector on a similar 200 keV STEM and a 300 keV STEM.","PeriodicalId":16791,"journal":{"name":"Journal of Physics E: Scientific Instruments","volume":"55 1","pages":"726-729"},"PeriodicalIF":0.0000,"publicationDate":"1989-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Physics E: Scientific Instruments","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/0022-3735/22/9/009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3

Abstract

The performance of a high take-off angle energy dispersive X-ray spectrometer (EDXS) on a 200 keV STEM has been investigated. A simple modification to the multichannel analyser (MCA) was used to extend the energy range beyond 200 keV to confirm that high energy, backscattered electrons were entering the detector. Up to 50% of the total count rate was observed to be due to these electrons. The spectrometer performance was determined both with and without a carbon filter which was used to stop the backscattered electrons entering the detector. It is shown that these electrons reduce the resolution and increase the dead-time and background noise in the X-ray spectra. The performance of the high take-off angle detector is compared with that of a horizontally mounted detector on a similar 200 keV STEM and a 300 keV STEM.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
高角度EDXS中的背散射电子效应
研究了高发射角能量色散x射线光谱仪(EDXS)在200kev STEM上的性能。对多通道分析仪(MCA)进行了简单的修改,将能量范围扩展到200 keV以上,以确认高能背散射电子正在进入探测器。总计数率的50%被观察到是由于这些电子。在有碳过滤器和没有碳过滤器的情况下测定了光谱仪的性能,碳过滤器用于阻止背散射电子进入探测器。结果表明,这些电子降低了x射线光谱的分辨率,增加了死区时间和背景噪声。将高起飞角探测器的性能与水平安装在类似的200 keV STEM和300 keV STEM上的探测器进行了比较。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Apparatus for the measurement of electro-optical rotation changes in polymer solutions DESIGN NOTE: Simple oven design for highly reactive metal beam applications Ratio transformer bridge for the measurement of dielectric constant of lossy liquids Temperature resistant and vacuum tight LiF connections Accounting for the effects of gas species on piston gauge calculated area: absolute mode
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1