Huu Loi Le, Hoang Long Nguyen, Quang Son Ong, M. Ho, Van Hoang Dao, Truong Son Truong, Van Tien Vo, Quang Vuong Le
{"title":"Establishment and quality assessment of X-ray narrow spectrum series according to ISO 4037:2019 at SSDL of center for nuclear technologies","authors":"Huu Loi Le, Hoang Long Nguyen, Quang Son Ong, M. Ho, Van Hoang Dao, Truong Son Truong, Van Tien Vo, Quang Vuong Le","doi":"10.53747/nst.v12i1.374","DOIUrl":null,"url":null,"abstract":"Narrow spectrum series from N-40 to N-120 were established according to ISO 4037:2019. This study was made through experimental measurements and the software modeling toolkit SpekPy v2.0. Experiments were performed to determine Inherent filtration (0.332 & 0.492 mm Al at 100 cm & 250 cm, respectively). The maximum deviation compared to ISO 4037:2019 of the measured HVLs values are not more then 0.10 mm Al and for modeling not exceed 0.09 mm Al; h range 0.91 to 0.97 for experimently and 0.89 to 0.96 for modeling (0.88 - 1, ISO 4037-1:2019). The maximum contribution of scattering is value 4.0% (< 5%, ISO 4037-1:2019). Radiation field uniformity of N-40 to N-120 should list all categories in detail 12.5, 11.0, 10.0, 9.0, 8.5 cm at 100 cm distance; and 35.5, 31.5, 29.0, 26.5, 24.5 cm at 250 cm beam center axis distance. On the basis of ISO 4037:2019, air kerma values at 100 cm and 250 cm have also been determined with uncertainties 3.85% to 5.18% for experimently and 3.41% to 4.19% for modeling (not exceed 10%, ISO 4037:2019). Therefore, the X-ray radiation field has been well established in compliance with the ISO 4037:2019, it can be used for calibrating survey meters and personal dosimeters.","PeriodicalId":19445,"journal":{"name":"Nuclear Science and Technology","volume":"51 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-01-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.53747/nst.v12i1.374","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Narrow spectrum series from N-40 to N-120 were established according to ISO 4037:2019. This study was made through experimental measurements and the software modeling toolkit SpekPy v2.0. Experiments were performed to determine Inherent filtration (0.332 & 0.492 mm Al at 100 cm & 250 cm, respectively). The maximum deviation compared to ISO 4037:2019 of the measured HVLs values are not more then 0.10 mm Al and for modeling not exceed 0.09 mm Al; h range 0.91 to 0.97 for experimently and 0.89 to 0.96 for modeling (0.88 - 1, ISO 4037-1:2019). The maximum contribution of scattering is value 4.0% (< 5%, ISO 4037-1:2019). Radiation field uniformity of N-40 to N-120 should list all categories in detail 12.5, 11.0, 10.0, 9.0, 8.5 cm at 100 cm distance; and 35.5, 31.5, 29.0, 26.5, 24.5 cm at 250 cm beam center axis distance. On the basis of ISO 4037:2019, air kerma values at 100 cm and 250 cm have also been determined with uncertainties 3.85% to 5.18% for experimently and 3.41% to 4.19% for modeling (not exceed 10%, ISO 4037:2019). Therefore, the X-ray radiation field has been well established in compliance with the ISO 4037:2019, it can be used for calibrating survey meters and personal dosimeters.