Optimization of the Position of the CR-39 Polycarbonate Sheet Inside the Solid State Track Detector “Measuring Device” Through Computational Fluid Dynamics Technique

A. Lima Flores, R. Palomino-Merino, V. Castaño, J. Golzarri, G. Espinosa
{"title":"Optimization of the Position of the CR-39 Polycarbonate Sheet Inside the Solid State Track Detector “Measuring Device” Through Computational Fluid Dynamics Technique","authors":"A. Lima Flores, R. Palomino-Merino, V. Castaño, J. Golzarri, G. Espinosa","doi":"10.15415/JNP.2021.82013","DOIUrl":null,"url":null,"abstract":"The “measuring device” is one of the most reliable, efficient and economic indoor radon dosimeters that exist. This device was developed by the Proyecto de Aplicaciones de la Dosimetria (PAD) at the Physics Institute of UNAM (IF-UNAM) and consists of a transparent rigid plastic cup, a CR-39 polycarbonate sheet and a standard size metal clip that is used to hold the polycarbonate in the center of the cup. The cup is wrapped and covered with a low-density polyurethane protector in order to prevent the detector from being irradiated by ionizing particles found in the environment. In this work, an analysis was carried out that allowed to understand how the radon concentration on the polycarbonate sheet varies when its height is changed with respect to the base of the plastic cup, in order to understand what position increase the probability of interaction between radon and the surface of the detector. For the development of this work, four computational simulations were performed with the technique called Computational Fluid Dynamics (CFD). The results shows that as the CR-39 is positioned more closed to the base of the cup, the probability of interaction of the radon and the detector increase. Based on these results it is concluded that, when there is a limit in the time in which a measuring device can be placed in the zone where it is desired to quantify indoor radon, it is recommended to collocated the CR-39 at 1 cm with respect to the base of the cup.","PeriodicalId":16534,"journal":{"name":"Journal of Nuclear Physics, Material Sciences, Radiation and Applications","volume":"55 1","pages":"109-114"},"PeriodicalIF":0.0000,"publicationDate":"2021-02-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Nuclear Physics, Material Sciences, Radiation and Applications","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15415/JNP.2021.82013","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The “measuring device” is one of the most reliable, efficient and economic indoor radon dosimeters that exist. This device was developed by the Proyecto de Aplicaciones de la Dosimetria (PAD) at the Physics Institute of UNAM (IF-UNAM) and consists of a transparent rigid plastic cup, a CR-39 polycarbonate sheet and a standard size metal clip that is used to hold the polycarbonate in the center of the cup. The cup is wrapped and covered with a low-density polyurethane protector in order to prevent the detector from being irradiated by ionizing particles found in the environment. In this work, an analysis was carried out that allowed to understand how the radon concentration on the polycarbonate sheet varies when its height is changed with respect to the base of the plastic cup, in order to understand what position increase the probability of interaction between radon and the surface of the detector. For the development of this work, four computational simulations were performed with the technique called Computational Fluid Dynamics (CFD). The results shows that as the CR-39 is positioned more closed to the base of the cup, the probability of interaction of the radon and the detector increase. Based on these results it is concluded that, when there is a limit in the time in which a measuring device can be placed in the zone where it is desired to quantify indoor radon, it is recommended to collocated the CR-39 at 1 cm with respect to the base of the cup.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
基于计算流体力学技术的CR-39聚碳酸酯片材在固体轨迹探测器“测量装置”内的位置优化
该“测量装置”是目前最可靠、最高效、最经济的室内氡剂量计之一。该装置由UNAM物理研究所(IF-UNAM)的应用剂量学项目(PAD)开发,由一个透明的刚性塑料杯,CR-39聚碳酸酯片和一个标准尺寸的金属夹组成,用于将聚碳酸酯固定在杯中心。杯子用低密度聚氨酯保护器包裹和覆盖,以防止探测器被环境中发现的电离粒子照射。在这项工作中,进行了一项分析,以了解当聚碳酸酯片的高度相对于塑料杯的底座发生变化时,聚碳酸酯片上的氡浓度是如何变化的,以便了解什么位置增加了氡与探测器表面相互作用的概率。为了开展这项工作,使用计算流体动力学(CFD)技术进行了四次计算模拟。结果表明,CR-39的位置越靠近杯底,氡与探测器相互作用的概率越大。根据这些结果,得出的结论是,当测量装置可以放置在想要量化室内氡的区域的时间有限制时,建议将CR-39放置在距离杯底1厘米的位置。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Radii of Thorium Nuclides Lying in Between the Drip Lines Charge Radius And Neutron Skin Thickness Of Platinum And Osmium Isotopes Near The Nuclear Drip Lines Evaluation of Natural Radioactivity Levels and Exhalation rate of 222Rn and 220Rn in the Soil Samples from the Kuthiran Hills, Kerala, India Deformation Effect on Proton Bubble Structure in N = 28 Isotones Phase Shift Analysis for Neutron-Alpha Elastic Scattering Using Phase Function Method with Local Gaussian Potential
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1