BIOANALYTICAL METHOD DEVELOPMENT AND VALIDATION FOR THE DETERMINATION OF SOFOSBUVIR FROM HUMAN PLASMA

S. Madhavi, A. Rani
{"title":"BIOANALYTICAL METHOD DEVELOPMENT AND VALIDATION FOR THE DETERMINATION OF SOFOSBUVIR FROM HUMAN PLASMA","authors":"S. Madhavi, A. Rani","doi":"10.22159/IJPPS.2017V9I3.16185","DOIUrl":null,"url":null,"abstract":"Objective: This study points to build up and validate a simple methodology to quantify the most used drug sofosbuvir for the treatment of hepatitis C virus (HCV) infection, in human plasma by using atazanavir as an Internal Standard (IS) for preclinical studies and validate as per USFDA guidelines. Methods: Sofosbuvir was isolated from plasma samples by liquid-liquid extraction method using acetonitrile; good chromatographic separation was achieved on Kromasil Column (250 mm ×4.6 mm, 5 µm). The mobile phase consisted of 0.1 % orthophosphoric acid (OPA) buffer pH 2 and acetonitrile in the ratio of (68:32, v/v), respectively. The analysis time was 7 min at a flow rate 1 ml/min. The photodiode array detector (PDA) detection was carried out at 228 nm. The suggested method was validated by performing linearity, system suitability, specificity and sensitivity, accuracy and precision, recovery, ruggedness, stability studies. The method was validated as per USFDA guidelines. Results: The developed method resulted in retention times of sofosbuvir and IS were found out to be 4.7 and 4.2 min respectively. The calibration curves are linear (r2 = 0.999) over the concentration range of 0.050-2.0 µg/ml of plasma analytes concentration. LOQ value was found to be 0.050 µg/ml with precision and accuracy. Within-batch % mean accuracy of the method ranged between 96.00% and 109.09%, and within-batch and total precision, expressed as the coefficient of variation, was 1.40–10.33%. Overall percentage mean recovery of sofosbuvir from spiked plasma was 84.14%. All the validated parameters were found to be within the limit. Conclusion: A simple, accurate, precise, linear, rugged and rapid RP-HPLC method was developed for quantitative estimation of sofosbuvir in human plasma and should be suitable for conducting pharmacokinetics studies and therapeutic drug monitoring.","PeriodicalId":14188,"journal":{"name":"International Journal of Pharmacy and Pharmaceutical Sciences","volume":"120 1","pages":"35-41"},"PeriodicalIF":0.0000,"publicationDate":"2017-02-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"28","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Pharmacy and Pharmaceutical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22159/IJPPS.2017V9I3.16185","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 28

Abstract

Objective: This study points to build up and validate a simple methodology to quantify the most used drug sofosbuvir for the treatment of hepatitis C virus (HCV) infection, in human plasma by using atazanavir as an Internal Standard (IS) for preclinical studies and validate as per USFDA guidelines. Methods: Sofosbuvir was isolated from plasma samples by liquid-liquid extraction method using acetonitrile; good chromatographic separation was achieved on Kromasil Column (250 mm ×4.6 mm, 5 µm). The mobile phase consisted of 0.1 % orthophosphoric acid (OPA) buffer pH 2 and acetonitrile in the ratio of (68:32, v/v), respectively. The analysis time was 7 min at a flow rate 1 ml/min. The photodiode array detector (PDA) detection was carried out at 228 nm. The suggested method was validated by performing linearity, system suitability, specificity and sensitivity, accuracy and precision, recovery, ruggedness, stability studies. The method was validated as per USFDA guidelines. Results: The developed method resulted in retention times of sofosbuvir and IS were found out to be 4.7 and 4.2 min respectively. The calibration curves are linear (r2 = 0.999) over the concentration range of 0.050-2.0 µg/ml of plasma analytes concentration. LOQ value was found to be 0.050 µg/ml with precision and accuracy. Within-batch % mean accuracy of the method ranged between 96.00% and 109.09%, and within-batch and total precision, expressed as the coefficient of variation, was 1.40–10.33%. Overall percentage mean recovery of sofosbuvir from spiked plasma was 84.14%. All the validated parameters were found to be within the limit. Conclusion: A simple, accurate, precise, linear, rugged and rapid RP-HPLC method was developed for quantitative estimation of sofosbuvir in human plasma and should be suitable for conducting pharmacokinetics studies and therapeutic drug monitoring.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
测定人血浆中索非布韦的生物分析方法的建立和验证
目的:本研究旨在建立并验证一种简单的方法,通过将阿扎那韦作为临床前研究的内标(IS)来量化人血浆中最常用的治疗丙型肝炎病毒(HCV)感染的药物索非布韦,并根据USFDA指南进行验证。方法:采用乙腈液-液萃取法从血浆样品中分离索非布韦;在Kromasil柱(250 mm ×4.6 mm, 5µm)上获得了良好的色谱分离。流动相为0.1%正磷酸(OPA)缓冲液pH 2和乙腈,比例为(68:32,v/v)。分析时间为7 min,流速为1 ml/min。采用光电二极管阵列检测器(PDA)在228 nm处进行检测。通过进行线性、系统适用性、特异性和敏感性、准确度和精密度、回收率、耐用性和稳定性研究验证了该方法。该方法按照USFDA指南进行了验证。结果:所建立的方法索非布韦和IS的保留时间分别为4.7 min和4.2 min。在0.050 ~ 2.0µg/ml血浆分析物浓度范围内,标定曲线呈线性关系(r2 = 0.999)。定量限为0.050µg/ml,精密度和准确度较高。该方法的批内平均准确度为96.00% ~ 109.09%,批内和总精密度(变异系数)为1.40 ~ 10.33%。索非布韦在加标血浆中的平均回收率为84.14%。所有验证参数均在限定范围内。结论:建立了一种简便、准确、精密度、线性、坚固、快速的反相高效液相色谱法定量测定人血浆中索非布韦的方法,适用于药代动力学研究和治疗药物监测。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
KNOWLEDGE, ATTITUDE AND APPLICATION OF HEMOVIGILANCE BY POSTGRADUATES AND INTERNS IN A TERTIARY CARE HOSPITAL GENERIC DRUG REGISTRATION AND COUNTRY-SPECIFIC REQUIREMENTS IN SAUDI ARABIA, QATAR, BAHRAIN, AND OMAN GASTRORETENTIVE EFFERVESCENT FLOATING TABLETS (GREFT) OF DRUGS ACTING ON CARDIOVASCULAR DISEASES REVIEW ON CHEMICAL CONSTITUENTS AND BIOLOGICAL ACTIVITIES OF GENUS JUNIPER A SIMULATION APPROACH FOR NOVEL 1,3,4 THIADIAZOLE ACETAMIDE MOIETIES AS POTENT ANTIMYCOBACTERIAL AGENTS
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1