Ikki Tateishi, M. Furukawa, H. Katsumata, S. Kaneco
{"title":"Metal and Molecular Vapor Separation Analysis for Direct Determination of Mn and Cu by Atomic Absorption Detection, Free of Background Absorption","authors":"Ikki Tateishi, M. Furukawa, H. Katsumata, S. Kaneco","doi":"10.3390/suschem3040029","DOIUrl":null,"url":null,"abstract":"The metal and molecular vapor separation analysis (MMVSA) of solid samples with an atomic absorption detector (AA) was investigated for the direct determination of manganese and copper in biological materials. An open column made with a molybdenum tube (i.d. 1.22 mm) with three-ring supporters was developed. Pure argon as a carrier gas flowed at a flow rate of 4.0 mL min−1. An ultrasonic agitation method was used for suspending NIST standard reference material powders in water. Manganese and copper in the biological powders were completely separated from Al, Ca, Fe, K, Mg, Na, and Zn elements by MMVSA under optimal experimental conditions. Several NIST biological samples were directly analyzed with satisfactory results. It was found that manganese and copper in biological materials without interferences from matrix elements could be directly determined after only an ultrasonic agitation of the biological powders. The advantages of the slurry sampling of MMVSA are simplicity, low cost, a high speed of analysis, and rapid calibration.","PeriodicalId":22103,"journal":{"name":"Sustainable Chemistry","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-11-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sustainable Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/suschem3040029","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2
Abstract
The metal and molecular vapor separation analysis (MMVSA) of solid samples with an atomic absorption detector (AA) was investigated for the direct determination of manganese and copper in biological materials. An open column made with a molybdenum tube (i.d. 1.22 mm) with three-ring supporters was developed. Pure argon as a carrier gas flowed at a flow rate of 4.0 mL min−1. An ultrasonic agitation method was used for suspending NIST standard reference material powders in water. Manganese and copper in the biological powders were completely separated from Al, Ca, Fe, K, Mg, Na, and Zn elements by MMVSA under optimal experimental conditions. Several NIST biological samples were directly analyzed with satisfactory results. It was found that manganese and copper in biological materials without interferences from matrix elements could be directly determined after only an ultrasonic agitation of the biological powders. The advantages of the slurry sampling of MMVSA are simplicity, low cost, a high speed of analysis, and rapid calibration.
采用原子吸收检测器(AA)对固体样品进行金属和分子气相分离分析(MMVSA),用于生物材料中锰和铜的直接测定。研制了一种由直径1.22 mm的钼管制成的带三环支架的开孔柱。纯氩气作为载气以4.0 mL min - 1的流速流动。采用超声搅拌法将NIST标准物质粉末悬浮在水中。在最佳实验条件下,采用MMVSA技术将生物粉中的锰、铜与Al、Ca、Fe、K、Mg、Na、Zn等元素完全分离。几个NIST生物样品直接分析,结果令人满意。发现生物材料中的锰和铜在不受基质元素干扰的情况下,只需对生物粉末进行超声搅拌即可直接测定。MMVSA浆液取样的优点是简单、成本低、分析速度快、校准速度快。