{"title":"Determination of sodium cromoglycate in human plasma by liquid chromatography–mass spectrometry in the turbo ion spray mode","authors":"M.L Ozoux , J Girault , J.M Malgouyat , O Pasquier","doi":"10.1016/S0378-4347(01)00420-0","DOIUrl":null,"url":null,"abstract":"<div><p>A highly sensitivity liquid chromatography–tandem mass spectrometry method has been developed for the quantitation of sodium cromoglycate (SCG) in human plasma. The method was validated over a linear range of 0.100–50.0 ng/ml, using <sup>13</sup>C<sub>4</sub> sodium cromoglycate as the internal standard. Compounds were extracted from 1.0 ml of lithium heparin plasma by methanol elution of C<sub>18</sub> solid-phase extraction cartridges. The dried residue was reconstituted with 100 μl of 0.01 N HCl, and 30 μl was injected onto the LC–MS–MS system. Chromatographic separation was achieved on a C8 (3.5 μm) column with an isocratic mobile phase of methanol–water–0.5 <em>M</em> ammonium acetate (35:64.8:0.2, v/v/v). The analytes were detected with a PE Sciex API 3000 mass spectrometer using turbo ion spray with positive ionization. Ions monitored in the multiple reaction monitoring (MRM) mode were <em>m</em>/<em>z</em> 469.2 (precursor ion) to <em>m</em>/<em>z</em> 245.1 (product ion) for SCG and <em>m</em>/<em>z</em> 473.2 (precursor ion) to <em>m</em>/<em>z</em> 247.1 (product ion) for <sup>13</sup>C<sub>4</sub> SCG (I.S.). The average recoveries of SCG and the I.S. from human plasma were 91 and 87%, respectively. The low limit of quantitation was 0.100 ng/ml. Results from a 4-day validation study demonstrated excellent precision (C.V.% values were between 1.9 and 6.5%) and accuracy (−5.4 to −1.2%) across the calibration range of 0.100–50.0 ng/ml.</p></div>","PeriodicalId":15463,"journal":{"name":"Journal of Chromatography B: Biomedical Sciences and Applications","volume":"765 2","pages":"Pages 179-185"},"PeriodicalIF":0.0000,"publicationDate":"2001-12-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0378-4347(01)00420-0","citationCount":"24","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Chromatography B: Biomedical Sciences and Applications","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0378434701004200","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 24
Abstract
A highly sensitivity liquid chromatography–tandem mass spectrometry method has been developed for the quantitation of sodium cromoglycate (SCG) in human plasma. The method was validated over a linear range of 0.100–50.0 ng/ml, using 13C4 sodium cromoglycate as the internal standard. Compounds were extracted from 1.0 ml of lithium heparin plasma by methanol elution of C18 solid-phase extraction cartridges. The dried residue was reconstituted with 100 μl of 0.01 N HCl, and 30 μl was injected onto the LC–MS–MS system. Chromatographic separation was achieved on a C8 (3.5 μm) column with an isocratic mobile phase of methanol–water–0.5 M ammonium acetate (35:64.8:0.2, v/v/v). The analytes were detected with a PE Sciex API 3000 mass spectrometer using turbo ion spray with positive ionization. Ions monitored in the multiple reaction monitoring (MRM) mode were m/z 469.2 (precursor ion) to m/z 245.1 (product ion) for SCG and m/z 473.2 (precursor ion) to m/z 247.1 (product ion) for 13C4 SCG (I.S.). The average recoveries of SCG and the I.S. from human plasma were 91 and 87%, respectively. The low limit of quantitation was 0.100 ng/ml. Results from a 4-day validation study demonstrated excellent precision (C.V.% values were between 1.9 and 6.5%) and accuracy (−5.4 to −1.2%) across the calibration range of 0.100–50.0 ng/ml.