{"title":"Extrapolation of the R-T Estimation in CoaguChek International Normalized Ratio Results Below 4","authors":"C. Richter","doi":"10.1097/POC.0000000000000189","DOIUrl":null,"url":null,"abstract":"Abstract While extensive literature has documented that the CoaguChek XS device results are not an accurate reflection of a patient's international normalized ratio (INR) when the values are elevated, only 1 equation has been able to consistently correct these results in the clinical setting. Data obtained from a previous study were used to apply the R-T Estimation to INR values below 4 to assess an extrapolation of this equation to an INR of 1. The results were consistent with previous studies using the R-T Estimation and demonstrate the correlation of this equation with venipuncture results with 95% confidence within 20% difference. This may be clinically relevant as a CoaguChek XS result of 3.8 is corrected using the R-T Estimation to an INR of 3.0.","PeriodicalId":20262,"journal":{"name":"Point of Care: The Journal of Near-Patient Testing & Technology","volume":"1 1","pages":"46 - 47"},"PeriodicalIF":0.0000,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Point of Care: The Journal of Near-Patient Testing & Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/POC.0000000000000189","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Abstract While extensive literature has documented that the CoaguChek XS device results are not an accurate reflection of a patient's international normalized ratio (INR) when the values are elevated, only 1 equation has been able to consistently correct these results in the clinical setting. Data obtained from a previous study were used to apply the R-T Estimation to INR values below 4 to assess an extrapolation of this equation to an INR of 1. The results were consistent with previous studies using the R-T Estimation and demonstrate the correlation of this equation with venipuncture results with 95% confidence within 20% difference. This may be clinically relevant as a CoaguChek XS result of 3.8 is corrected using the R-T Estimation to an INR of 3.0.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
CoaguChek国际归一化比率结果中R-T估计的外推低于4
虽然大量文献表明,CoaguChek XS装置的结果并不能准确反映患者的国际标准化比率(INR),当该值升高时,只有一个方程能够在临床环境中始终如一地纠正这些结果。从先前的研究中获得的数据用于将R-T估计应用于低于4的印度卢比值,以评估该方程外推到1的印度卢比。结果与前人使用R-T估计的研究结果一致,表明该方程与静脉穿刺结果的相关性为95%,差异在20%以内。这可能与临床相关,因为使用R-T估计将CoaguChek XS结果3.8校正为INR为3.0。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A Comparison of International Normalized Ratio Results by Point-of-Care Device and Clinical Laboratory Analyzers in a Vascular Surgery Department Point-of-Care Testing in Hypercoagulable Conditions Managed With Warfarin Critical Care Nurses' Views and Experiences of Preanalytical Factors Influencing Point-of-Care Testing Rapid Assessment of Coagulation at the Point of Care With the Hemochron Signature Elite System Invited Product Profile: Detecting Antibodies for Lyme Disease at the Point of Care
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1