Preload and friction in an implant-abutment-screw complex including a carbon-coated titanium alloy abutment screw: an in vitro study.

IF 3.1 3区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE International Journal of Implant Dentistry Pub Date : 2023-03-22 DOI:10.1186/s40729-023-00473-3
K Sagheb, C-I Görgen, S Döll, I Schmidtmann, S Wentaschek
{"title":"Preload and friction in an implant-abutment-screw complex including a carbon-coated titanium alloy abutment screw: an in vitro study.","authors":"K Sagheb, C-I Görgen, S Döll, I Schmidtmann, S Wentaschek","doi":"10.1186/s40729-023-00473-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>An experimental approach was designed to measure the preload force, the coefficient of friction and the component of the tightening torque that is needed to surmount the thread-friction in an implant-abutment-screw complex that includes a carbon-coated screw. With the determined preload values the coefficient of friction was calculated.</p><p><strong>Methods: </strong>25 unused complexes, containing an implant, an abutment and a carbon-coated titanium alloy abutment screw, were tested. A custom load frame with two load cells and associated electronics was used. The threads were not lubricated. All abutment screws were torqued ten times to 25 Ncm. The produced preload values and a force that was proportional to the thread-friction component of the tightening torque were recorded.</p><p><strong>Results: </strong>Mean preload values decreased significantly with the number of repetitions (p < 0.0001) from initially 329.9 N ± 33.3 (range 255.7 to 383.9) to 253.7 N ± 36.8 (range 200.1 to 332.5) for the last tightening procedure. The corresponding change in the calculated coefficient of friction was 0.33 ± 0.04 (range 0.28 to 0.43) to 0.44 ± 0.07 (range 0.32 to 0.56). For the thread-friction no corresponding trend for consecutive tightening repetitions could be noticed.</p><p><strong>Conclusions: </strong>In the investigated implant-abutment units, repeated use of a coated abutment screw appears to increase the friction of the screw head and thereby decrease the preload. These results indicate that a pre-used coated implant-abutment-screw will fail reaching optimal screw preload.</p>","PeriodicalId":14076,"journal":{"name":"International Journal of Implant Dentistry","volume":null,"pages":null},"PeriodicalIF":3.1000,"publicationDate":"2023-03-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10033796/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Implant Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40729-023-00473-3","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: An experimental approach was designed to measure the preload force, the coefficient of friction and the component of the tightening torque that is needed to surmount the thread-friction in an implant-abutment-screw complex that includes a carbon-coated screw. With the determined preload values the coefficient of friction was calculated.

Methods: 25 unused complexes, containing an implant, an abutment and a carbon-coated titanium alloy abutment screw, were tested. A custom load frame with two load cells and associated electronics was used. The threads were not lubricated. All abutment screws were torqued ten times to 25 Ncm. The produced preload values and a force that was proportional to the thread-friction component of the tightening torque were recorded.

Results: Mean preload values decreased significantly with the number of repetitions (p < 0.0001) from initially 329.9 N ± 33.3 (range 255.7 to 383.9) to 253.7 N ± 36.8 (range 200.1 to 332.5) for the last tightening procedure. The corresponding change in the calculated coefficient of friction was 0.33 ± 0.04 (range 0.28 to 0.43) to 0.44 ± 0.07 (range 0.32 to 0.56). For the thread-friction no corresponding trend for consecutive tightening repetitions could be noticed.

Conclusions: In the investigated implant-abutment units, repeated use of a coated abutment screw appears to increase the friction of the screw head and thereby decrease the preload. These results indicate that a pre-used coated implant-abutment-screw will fail reaching optimal screw preload.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
包括碳涂层钛合金基台螺钉在内的种植体-基台-螺钉复合体的预紧力和摩擦力:体外研究。
目的:设计了一种实验方法来测量种植体-基台-螺丝复合体(包括碳涂层螺丝)的预紧力、摩擦系数以及克服螺纹摩擦所需的拧紧扭矩。方法:测试了 25 个未使用的复合体,包括一个种植体、一个基台和一个碳涂层钛合金基台螺丝。测试中使用了一个带有两个传感器和相关电子设备的定制载荷框架。螺纹未进行润滑。所有基台螺钉均以 25 Ncm 扭力拧紧十次。记录产生的预紧力值和与拧紧扭矩的螺纹摩擦力成正比的力:结果:随着重复次数的增加,平均预紧力值明显降低(p 结论:随着重复次数的增加,平均预紧力值明显降低:在所研究的种植体基台中,重复使用涂层基台螺钉似乎会增加螺钉头部的摩擦力,从而降低预紧力。这些结果表明,重复使用的涂层基台螺钉无法达到最佳的螺钉预紧力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
International Journal of Implant Dentistry
International Journal of Implant Dentistry DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
1.70
自引率
7.40%
发文量
53
审稿时长
13 weeks
期刊介绍: The International Journal of Implant Dentistry is a peer-reviewed open access journal published under the SpringerOpen brand. The journal is dedicated to promoting the exchange and discussion of all research areas relevant to implant dentistry in the form of systematic literature or invited reviews, prospective and retrospective clinical studies, clinical case reports, basic laboratory and animal research, and articles on material research and engineering.
期刊最新文献
Effect of osteopenia and osteoporosis on failure of first and second dental implants: a retrospective observational study. Influence of gingival phenotype on crestal bone loss at implants : A long-term 2 to 20-year cohort study in periodontally compromised patient. An eight-year retrospective study on the clinical outcomes of laser surface-treated implants. Long-term follow-up of a patient with Parkinson's disease under nursing care after replacement of fixed implant-supported prostheses with an implant overdenture: a case report. Accuracy of computer-guided implant surgery in partially edentulous patients: a prospective observational study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1