The influence of electrocardiogram-gated computed tomography reconstruction into 8 or 10 cardiac phases on cardiac-pulsatility-induced motion quantification of stent grafts in the aorta
Jaimy A. Simmering PhD , Deborah A. Zagers MSc , Robert H. Geelkerken PhD , Henny Kuipers BSc , Gerben A. te Riet o.g. Scholten MSc , Michel M.P.J. Reijnen PhD , Cornelis H. Slump PhD
{"title":"The influence of electrocardiogram-gated computed tomography reconstruction into 8 or 10 cardiac phases on cardiac-pulsatility-induced motion quantification of stent grafts in the aorta","authors":"Jaimy A. Simmering PhD , Deborah A. Zagers MSc , Robert H. Geelkerken PhD , Henny Kuipers BSc , Gerben A. te Riet o.g. Scholten MSc , Michel M.P.J. Reijnen PhD , Cornelis H. Slump PhD","doi":"10.1016/j.jvssci.2023.100131","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>The goal of this study was to determine to what extent aortic stent graft motion quantification is comparable between electrocardiogram (ECG)-gated computed tomography (CT) scans with reconstructions into 8 and 10 cardiac phases on CT scanners from two different vendors.</p></div><div><h3>Methods</h3><p>An experimental setup that induces motion of an aortic stent graft, according to a predefined aortic blood pressure wave, was placed in two CT scanners of different vendors. The stent graft motion was captured using an ECG-gated CT technique and quantified using dedicated analysis algorithms. The calculated motion amplitudes and total traveled path lengths of stent segmentations were compared between scans reconstructed into 8 and 10 phases and between the scanners, after validation with sensor measurements and repeated measurements.</p></div><div><h3>Results</h3><p>No difference in motion amplitudes in z-direction (craniocaudal direction) was observed between the reconstructions into 8 and 10 phases (0.02 mm; 95% confidence interval [CI], –0.01 to 0.05 mm; <em>P</em> = .358). The z-amplitudes differed by 0.04 mm (95% CI, 0.01-0.07 mm; <em>P</em> = .003) between the different CT scanners. Path lengths differed 0.07 mm (95% CI, 0.01-to 0.13 mm; <em>P</em> = .013) between the reconstructions into 8 and 10 phases and 0.13 mm (95% CI, 0.06-0.17 mm; <em>P</em> < .001) between the different scanners.</p></div><div><h3>Conclusions</h3><p>The motion amplitudes can accurately be compared between 8 and 10 phases and between the two scanners, without differences larger than the voxel size of 0.3 × 0.3 × 0.5 mm. Clinical motion analysis results of different ECG-gated CT scans and CT scanners can be compared up to the accuracy of the CT scan.</p></div>","PeriodicalId":74035,"journal":{"name":"JVS-vascular science","volume":"4 ","pages":"Article 100131"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666350323000354/pdfft?md5=436e7fbbab4cd50b922f2bef3645a308&pid=1-s2.0-S2666350323000354-main.pdf","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JVS-vascular science","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666350323000354","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1
Abstract
Objective
The goal of this study was to determine to what extent aortic stent graft motion quantification is comparable between electrocardiogram (ECG)-gated computed tomography (CT) scans with reconstructions into 8 and 10 cardiac phases on CT scanners from two different vendors.
Methods
An experimental setup that induces motion of an aortic stent graft, according to a predefined aortic blood pressure wave, was placed in two CT scanners of different vendors. The stent graft motion was captured using an ECG-gated CT technique and quantified using dedicated analysis algorithms. The calculated motion amplitudes and total traveled path lengths of stent segmentations were compared between scans reconstructed into 8 and 10 phases and between the scanners, after validation with sensor measurements and repeated measurements.
Results
No difference in motion amplitudes in z-direction (craniocaudal direction) was observed between the reconstructions into 8 and 10 phases (0.02 mm; 95% confidence interval [CI], –0.01 to 0.05 mm; P = .358). The z-amplitudes differed by 0.04 mm (95% CI, 0.01-0.07 mm; P = .003) between the different CT scanners. Path lengths differed 0.07 mm (95% CI, 0.01-to 0.13 mm; P = .013) between the reconstructions into 8 and 10 phases and 0.13 mm (95% CI, 0.06-0.17 mm; P < .001) between the different scanners.
Conclusions
The motion amplitudes can accurately be compared between 8 and 10 phases and between the two scanners, without differences larger than the voxel size of 0.3 × 0.3 × 0.5 mm. Clinical motion analysis results of different ECG-gated CT scans and CT scanners can be compared up to the accuracy of the CT scan.