Pedro Fragoso Costa, Walter Jentzen, Finja SÜßELBECK, Wolfgang P Fendler, Christoph Rischpler, Ken Herrmann, Maurizio Conti, David Kersting, Manuel Weber
{"title":"Reduction of emission time for [68Ga]Ga-PSMA PET/CT using the digital biograph vision: a phantom study.","authors":"Pedro Fragoso Costa, Walter Jentzen, Finja SÜßELBECK, Wolfgang P Fendler, Christoph Rischpler, Ken Herrmann, Maurizio Conti, David Kersting, Manuel Weber","doi":"10.23736/S1824-4785.21.03300-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The aim of this phantom study was to optimize the [<sup>68</sup>Ga]Ga-PSMA PET/CT examination in terms of scan time duration and image reconstruction parameters, in combination with PSF and TOF modelling, in a digital Biograph Vision PET/CT scanner.</p><p><strong>Methods: </strong>Three types of phantoms were used: 1) soft-tissue tumor phantom consisting of six spheres mounted in a torso phantom; 2) bone-lung tumor phantom; 3) resolution phantom. Phantom inserts were filled with activity concentrations (ACs) that were derived from clinical data. Phantom data were acquired in list-mode at one bed position. Images with emission data ranging from 30 to 210 s in 30-s increments were reconstructed from a reference image acquired with 3.5-min emission. Iterative image reconstruction (OSEM), point-spread-function (PSF) and time-of-flight (TOF) options were applied using different iterations, Gaussian filters, and voxel sizes. The criteria for image quality was lesion detectability and lesion quantification, evaluated as contrast-to-noise ratio (CNR) and maximum AC (peak AC), respectively. A threshold value of CNR above 6 and percentage maximum AC (peak AC) deviation range of ±20% of the reference image were considered acceptable. The proposed single-bed scan time reduction was projected to a whole-body examination (patient validation scan) using the continuous-bed-motion mode.</p><p><strong>Results: </strong>Sphere and background ACs of 20 kBq/mL and 1 kBq/mL were selected, respectively. The optimized single-bed scan time was approximately 60 s using OSEM-TOF or OSEM-TOF+PSF (four iterations, 4.0-mm Gaussian filter and almost isotropic voxel size of 3.0-mm side length), resulting in a PET spatial resolution of 6.3 mm for OSEM-TOF and 5.5 mm for OSEM-TOF+PSF. In the patient validation, the maximum percentage difference in lesion quantification between standard and optimized protocol (whole-body scan time of 15 vs. 5 min) was below 19%.</p><p><strong>Conclusions: </strong>A reduction of single-bed and whole-body scan time for [<sup>68</sup>Ga]Ga-PSMA PET/CT compared to current recommended clinical acquisition protocols is postulated. Clinical studies are warranted to validate the applicability of this protocol.</p>","PeriodicalId":23069,"journal":{"name":"The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of...","volume":"67 1","pages":"57-68"},"PeriodicalIF":0.0000,"publicationDate":"2023-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of...","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.23736/S1824-4785.21.03300-8","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8
Abstract
Background: The aim of this phantom study was to optimize the [68Ga]Ga-PSMA PET/CT examination in terms of scan time duration and image reconstruction parameters, in combination with PSF and TOF modelling, in a digital Biograph Vision PET/CT scanner.
Methods: Three types of phantoms were used: 1) soft-tissue tumor phantom consisting of six spheres mounted in a torso phantom; 2) bone-lung tumor phantom; 3) resolution phantom. Phantom inserts were filled with activity concentrations (ACs) that were derived from clinical data. Phantom data were acquired in list-mode at one bed position. Images with emission data ranging from 30 to 210 s in 30-s increments were reconstructed from a reference image acquired with 3.5-min emission. Iterative image reconstruction (OSEM), point-spread-function (PSF) and time-of-flight (TOF) options were applied using different iterations, Gaussian filters, and voxel sizes. The criteria for image quality was lesion detectability and lesion quantification, evaluated as contrast-to-noise ratio (CNR) and maximum AC (peak AC), respectively. A threshold value of CNR above 6 and percentage maximum AC (peak AC) deviation range of ±20% of the reference image were considered acceptable. The proposed single-bed scan time reduction was projected to a whole-body examination (patient validation scan) using the continuous-bed-motion mode.
Results: Sphere and background ACs of 20 kBq/mL and 1 kBq/mL were selected, respectively. The optimized single-bed scan time was approximately 60 s using OSEM-TOF or OSEM-TOF+PSF (four iterations, 4.0-mm Gaussian filter and almost isotropic voxel size of 3.0-mm side length), resulting in a PET spatial resolution of 6.3 mm for OSEM-TOF and 5.5 mm for OSEM-TOF+PSF. In the patient validation, the maximum percentage difference in lesion quantification between standard and optimized protocol (whole-body scan time of 15 vs. 5 min) was below 19%.
Conclusions: A reduction of single-bed and whole-body scan time for [68Ga]Ga-PSMA PET/CT compared to current recommended clinical acquisition protocols is postulated. Clinical studies are warranted to validate the applicability of this protocol.