A Noto, A Cortegiani, G Genoese, L Appendini, C Gregoretti, A Carlucci, C Crimi
{"title":"Performance of helmet CPAP using different configurations: Turbine-driven ventilators <i>vs</i> Venturi devices.","authors":"A Noto, A Cortegiani, G Genoese, L Appendini, C Gregoretti, A Carlucci, C Crimi","doi":"10.1016/j.pulmoe.2023.04.009","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Traditionally, Venturi-based flow generators have been preferred over mechanical ventilators to provide continuous positive airway pressure (CPAP) through the helmet (h-CPAP). Recently, modern turbine-driven ventilators (TDVs) showed to be safe and effective in delivering h-CPAP. We aimed to compare the pressure stability during h-CPAP delivered by Venturi devices and TDVs and assess the impact of High Efficiency Particulate Air (HEPA) filters on their performance.</p><p><strong>Methods: </strong>We performed a bench study using an artificial lung simulator set in a restrictive respiratory condition, simulating two different levels of patient effort (high and low) with and without the interposition of the HEPA filter. We calculated the average of minimal (Pmin), maximal (Pmax) and mean (Pmean) airway pressure and the time product measured on the airway pressure curve (PTPinsp). We defined the pressure swing (Pswing) as Pmax - Pmin and pressure drop (Pdrop) as End Expiratory Pressure - Pmin.</p><p><strong>Results: </strong>Pswing across CPAP levels varied widely among all the tested devices. During \"low effort\", no difference in Pswing and Pdrop was found between Venturi devices and TDVs; during high effort, Pswing (p<0.001) and Pdrop (p<0.001) were significantly higher in TDVs compared to Venturi devices, but the PTPinsp was lower (1.50 SD 0.54 vs 1.67 SD 0.55, p<0.001). HEPA filter addition almost doubled Pswing and PTPinsp (p<0.001) but left unaltered the differences among Venturi and TDVs systems in favor of the latter (p<0.001).</p><p><strong>Conclusions: </strong>TDVs performed better than Venturi systems in delivering a stable positive pressure level during h-CPAP in a bench setting.</p>","PeriodicalId":54237,"journal":{"name":"Pulmonology","volume":" ","pages":"2416789"},"PeriodicalIF":10.4000,"publicationDate":"2025-12-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pulmonology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.pulmoe.2023.04.009","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/24 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Traditionally, Venturi-based flow generators have been preferred over mechanical ventilators to provide continuous positive airway pressure (CPAP) through the helmet (h-CPAP). Recently, modern turbine-driven ventilators (TDVs) showed to be safe and effective in delivering h-CPAP. We aimed to compare the pressure stability during h-CPAP delivered by Venturi devices and TDVs and assess the impact of High Efficiency Particulate Air (HEPA) filters on their performance.
Methods: We performed a bench study using an artificial lung simulator set in a restrictive respiratory condition, simulating two different levels of patient effort (high and low) with and without the interposition of the HEPA filter. We calculated the average of minimal (Pmin), maximal (Pmax) and mean (Pmean) airway pressure and the time product measured on the airway pressure curve (PTPinsp). We defined the pressure swing (Pswing) as Pmax - Pmin and pressure drop (Pdrop) as End Expiratory Pressure - Pmin.
Results: Pswing across CPAP levels varied widely among all the tested devices. During "low effort", no difference in Pswing and Pdrop was found between Venturi devices and TDVs; during high effort, Pswing (p<0.001) and Pdrop (p<0.001) were significantly higher in TDVs compared to Venturi devices, but the PTPinsp was lower (1.50 SD 0.54 vs 1.67 SD 0.55, p<0.001). HEPA filter addition almost doubled Pswing and PTPinsp (p<0.001) but left unaltered the differences among Venturi and TDVs systems in favor of the latter (p<0.001).
Conclusions: TDVs performed better than Venturi systems in delivering a stable positive pressure level during h-CPAP in a bench setting.
PulmonologyMedicine-Pulmonary and Respiratory Medicine
CiteScore
14.30
自引率
5.10%
发文量
159
审稿时长
19 days
期刊介绍:
Pulmonology (previously Revista Portuguesa de Pneumologia) is the official journal of the Portuguese Society of Pulmonology (Sociedade Portuguesa de Pneumologia/SPP). The journal publishes 6 issues per year and focuses on respiratory system diseases in adults and clinical research. It accepts various types of articles including peer-reviewed original articles, review articles, editorials, and opinion articles. The journal is published in English and is freely accessible through its website, as well as Medline and other databases. It is indexed in Science Citation Index Expanded, Journal of Citation Reports, Index Medicus/MEDLINE, Scopus, and EMBASE/Excerpta Medica.