Investigation of the cleaning performance of commercial orthodontic cleaning tablets regarding biofilm removal on PMMA test specimens : An ex vivo study.

IF 1.3 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie Pub Date : 2025-03-01 Epub Date: 2023-06-02 DOI:10.1007/s00056-023-00474-2
A Arndt-Fink, P-G Jost-Brinkmann
{"title":"Investigation of the cleaning performance of commercial orthodontic cleaning tablets regarding biofilm removal on PMMA test specimens : An ex vivo study.","authors":"A Arndt-Fink, P-G Jost-Brinkmann","doi":"10.1007/s00056-023-00474-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The purpose of this ex vivo study was to compare the cleaning performance of three commercially available orthodontic cleaners on polymethyl methacrylate (PMMA) test specimens covered with biofilm.</p><p><strong>Methods: </strong>Twenty subjects wore an individually manufactured vacuum-formed maxillary splint with four integrated PMMA test specimens for 7 days. The four test specimens were located on the buccal surfaces of the maxillary molars. After a 7-day wearing period, the PMMA test specimens colonized by biofilm were divided into two halves. One half was placed in 150 ml of tap water or in 150 ml of cleaning solution of the cleaners Retainer Brite® (Dentsply International Raintree Essix, Sarasota, FL, USA), Kukis® Xpress (Reckitt Benckiser, Heidelberg, Germany) or Dontodent (Propack, Heidelberg, Germany) while the other half remained uncleaned. The modified o‑phthaldialdehyde (OPA) method was used to determine the amount of protein on both halves of the test specimens. The difference was tested for significance as a measure of the cleaning effect using a paired sample t‑test.</p><p><strong>Results: </strong>The cleaning performance of the three orthodontic cleaners was higher than the cleaning performance of tap water (mean 25.9 ± 6.5%). While Retainer Brite® (mean 54.5 ± 7.1%) removed significantly more biofilm than Dontodent (mean 41.5 ± 9.2%, p < 0.001) and Kukis® Xpress (mean 39.9 ± 11.5%, p < 0.001), there was no significant difference in the cleaning performance between Kukis® Xpress and Dontodent (p = 1).</p><p><strong>Conclusion: </strong>Seven-day-old biofilm is only removed partially by the investigated orthodontic cleaners, so that they are not suitable as the only measure for removing established biofilms.</p>","PeriodicalId":54776,"journal":{"name":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","volume":" ","pages":"89-97"},"PeriodicalIF":1.3000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11861341/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00056-023-00474-2","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/6/2 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The purpose of this ex vivo study was to compare the cleaning performance of three commercially available orthodontic cleaners on polymethyl methacrylate (PMMA) test specimens covered with biofilm.

Methods: Twenty subjects wore an individually manufactured vacuum-formed maxillary splint with four integrated PMMA test specimens for 7 days. The four test specimens were located on the buccal surfaces of the maxillary molars. After a 7-day wearing period, the PMMA test specimens colonized by biofilm were divided into two halves. One half was placed in 150 ml of tap water or in 150 ml of cleaning solution of the cleaners Retainer Brite® (Dentsply International Raintree Essix, Sarasota, FL, USA), Kukis® Xpress (Reckitt Benckiser, Heidelberg, Germany) or Dontodent (Propack, Heidelberg, Germany) while the other half remained uncleaned. The modified o‑phthaldialdehyde (OPA) method was used to determine the amount of protein on both halves of the test specimens. The difference was tested for significance as a measure of the cleaning effect using a paired sample t‑test.

Results: The cleaning performance of the three orthodontic cleaners was higher than the cleaning performance of tap water (mean 25.9 ± 6.5%). While Retainer Brite® (mean 54.5 ± 7.1%) removed significantly more biofilm than Dontodent (mean 41.5 ± 9.2%, p < 0.001) and Kukis® Xpress (mean 39.9 ± 11.5%, p < 0.001), there was no significant difference in the cleaning performance between Kukis® Xpress and Dontodent (p = 1).

Conclusion: Seven-day-old biofilm is only removed partially by the investigated orthodontic cleaners, so that they are not suitable as the only measure for removing established biofilms.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
市售正畸清洁片去除PMMA生物膜清洁性能的离体研究。
目的:本离体研究的目的是比较三种市售正畸清洁剂对覆盖生物膜的聚甲基丙烯酸甲酯(PMMA)测试标本的清洁性能。方法:20名受试者分别佩戴自制的真空成型上颌夹板和4个集成PMMA试件,持续7天。4个试验标本均位于上颌磨牙的颊面。经过7天的佩戴期后,将生物膜定殖的PMMA试样分成两半。其中一半放置在150 毫升的自来水中,或150 毫升的清洗液中,这些清洗液分别是:洁牙器®(Dentsply International Raintree Essix,美国佛罗里达州Sarasota)、Kukis®Xpress(德国海德堡利洁时公司)或Dontodent(德国海德堡Propack公司),而另一半则未清洗。改进的邻苯二醛(OPA)方法用于确定两半试样上的蛋白质量。使用配对样本t检验检验差异的显著性,以衡量清洁效果。结果:三种正畸清洗剂的清洁性能均高于自来水的清洁性能(平均25.9 ±6.5%)。牙釉质洁牙剂(Retainer Brite®)去除的生物膜数量(平均54.5 ±7.1%)明显高于Dontodent(平均41.5 ±9.2%,p )。结论:7日龄生物膜仅部分被正畸洁牙剂去除,因此不适合作为去除已形成生物膜的唯一措施。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.90
自引率
0.00%
发文量
64
审稿时长
>12 weeks
期刊介绍: The Journal of Orofacial Orthopedics provides orthodontists and dentists who are also actively interested in orthodontics, whether in university clinics or private practice, with highly authoritative and up-to-date information based on experimental and clinical research. The journal is one of the leading publications for the promulgation of the results of original work both in the areas of scientific and clinical orthodontics and related areas. All articles undergo peer review before publication. The German Society of Orthodontics (DGKFO) also publishes in the journal important communications, statements and announcements.
期刊最新文献
Is simultaneous placement of orthodontic mini-implants and skeletally anchored appliances advisable? : CBCT-based cadaver study comparing the accuracy of the digital workflow with methods requiring two visits. Mitteilungen der DGKFO. Investigation of the cleaning performance of commercial orthodontic cleaning tablets regarding biofilm removal on PMMA test specimens : An ex vivo study. Comparison of six different CAD/CAM retainers vs. the stainless steel twistflex retainer: an in vitro investigation of survival rate and stability. Changes in the distribution of occlusal forces in the course of the orthodontic retention phase : A prospective cohort study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1