Biomechanical and Histologic Evaluation of Osseointegration of Titanium Dental Implants Modified by Various Combinations of Sandblasting, Acid-Etching, Hydroxyapatite, and Hyaluronic Acid Coating Techniques.
{"title":"Biomechanical and Histologic Evaluation of Osseointegration of Titanium Dental Implants Modified by Various Combinations of Sandblasting, Acid-Etching, Hydroxyapatite, and Hyaluronic Acid Coating Techniques.","authors":"Emre Yurttutan, Ömür Dereci, Mine Alkaya Karagöz","doi":"10.11607/jomi.9935","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare the osseointegration process of the titanium dental implants with five different surface characteristics-sandblasted, sandblasted and acid-etched, hyaluronic acid-coated (HYA), hydroxyapatite-coated (HA), and machined-in an experimental sheep model at 1- and 3-month examinations.</p><p><strong>Materials and methods: </strong>One hundred sixty dental implants were placed in the left and right tibias of 16 sheep. Five experimental groups were designed. Eight animals (80 implants) were used for biomechanical tests of reverse torque analysis and resonance frequency analysis (RFA). Eight of them (80 implants) were used for the evaluation of bone-to-implant contact (BIC) percentage in histomorphometric analysis. Forty of 80 implants (8 implants for each group) were used at 1-month examinations, and the remaining 40 (8 implants for each group) implants were used at 3-month examinations in the biomechanical test group and histomorphometric examination group, separately.</p><p><strong>Results: </strong>Intergroup analysis at the 3-month followup showed that the increase in the implant stability quotient (ISQ) value was statistically significant for only the HYA group (<i>P</i> < .05). According to ISQ values at 1 and 3 months, group HYA showed statistically higher values at the 1 and 3-month examinations (<i>P</i> < .05). Groups HYA and HA had statistically higher reverse torque values than other groups at the 1-month examination (<i>P</i> < .05). At the 3-month evaluation, the HYA group showed significantly higher reverse torque values compared to other groups (<i>P</i> < .05). The BIC values of the sandblasted and acid-etched, HYA, and HA groups were significantly higher than the sandblasted and machined groups at the 1- and 3-month examinations (<i>P</i> < .05). The BIC value for the HA group showed decreased value at the 3-month examination compared to the 1-month examination (<i>P</i> < .05).</p><p><strong>Conclusion: </strong>The RFA, reverse torque, and histomorphometric analysis at 1- and 3-month examinations show that dental implants coated with HYA may have increased potential for osseointegration compared to dental implants with sandblasted, sandblasted and acid-etched, machined, and HA-coated surfaces. Int J Oral Maxillofac Implants 2023;38:583-590. doi: 10.11607/jomi.9935.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2023-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11607/jomi.9935","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To compare the osseointegration process of the titanium dental implants with five different surface characteristics-sandblasted, sandblasted and acid-etched, hyaluronic acid-coated (HYA), hydroxyapatite-coated (HA), and machined-in an experimental sheep model at 1- and 3-month examinations.
Materials and methods: One hundred sixty dental implants were placed in the left and right tibias of 16 sheep. Five experimental groups were designed. Eight animals (80 implants) were used for biomechanical tests of reverse torque analysis and resonance frequency analysis (RFA). Eight of them (80 implants) were used for the evaluation of bone-to-implant contact (BIC) percentage in histomorphometric analysis. Forty of 80 implants (8 implants for each group) were used at 1-month examinations, and the remaining 40 (8 implants for each group) implants were used at 3-month examinations in the biomechanical test group and histomorphometric examination group, separately.
Results: Intergroup analysis at the 3-month followup showed that the increase in the implant stability quotient (ISQ) value was statistically significant for only the HYA group (P < .05). According to ISQ values at 1 and 3 months, group HYA showed statistically higher values at the 1 and 3-month examinations (P < .05). Groups HYA and HA had statistically higher reverse torque values than other groups at the 1-month examination (P < .05). At the 3-month evaluation, the HYA group showed significantly higher reverse torque values compared to other groups (P < .05). The BIC values of the sandblasted and acid-etched, HYA, and HA groups were significantly higher than the sandblasted and machined groups at the 1- and 3-month examinations (P < .05). The BIC value for the HA group showed decreased value at the 3-month examination compared to the 1-month examination (P < .05).
Conclusion: The RFA, reverse torque, and histomorphometric analysis at 1- and 3-month examinations show that dental implants coated with HYA may have increased potential for osseointegration compared to dental implants with sandblasted, sandblasted and acid-etched, machined, and HA-coated surfaces. Int J Oral Maxillofac Implants 2023;38:583-590. doi: 10.11607/jomi.9935.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.