Thibaut Lucena, Etienne Cavaignac, Louis Courtot, Vincent Marot, Philippe Chiron, Nicolas Reina
{"title":"Implant breakage and revision factors for modular fluted tapered stems in revision total hip arthroplasty.","authors":"Thibaut Lucena, Etienne Cavaignac, Louis Courtot, Vincent Marot, Philippe Chiron, Nicolas Reina","doi":"10.1177/11207000231180065","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Modular stems in revision total hip arthroplasties allow diaphyseal fixation and optimal restoration of the architecture of the proximal femur. Several studies report metaphyseal implant breakage having a negative impact on survivorship. The aim of the study was to evaluate the outcome of an uncemented modular fluted tapered stem (MFT) in revision surgery.</p><p><strong>Methods: </strong>In a retrospective study, 316 patients who had revision surgery using the same design of MFT implant (Modular Revision Stem [MRS], Lima Corporate, Italy) between 2012 and 2017 were identified. Patients were male in 51% of cases and mean age was 74 years. Indications were 110 periprosthetic fractures, 98 periprosthetic joint infection, 97 aseptic loosening, 10 instability and 1 other cause. Survivorship, complications, clinical and radiographic outcomes, were assessed. Mean follow-up was 5 years.</p><p><strong>Results: </strong>There was no implant breakage. At 5-year follow-up, the survivorship, free from revision for aseptic loosening and free of revision for any reason were 96% and 87%, respectively. At last follow-up (8 years), these figures were 92% and 71%, respectively. 31 implants were revised. The risk of revision for any cause was higher for extreme length metaphyseal implants, hazard ratio was 3.7 (95% CI, 1.82-7.52). A mean stem subsidence of 9 mm was noted in 37 cases; 4 were revised for aseptic loosening. The Harris Hip Score at final follow-up was 82.</p><p><strong>Conclusions: </strong>At 5-year follow-up, the MFT implant showed a good survivorship and outcomes with no specific complications. Unlike literature reports, no specific complications occurred with this design. Stem junction positioning and therefore metaphyseal length may be key to optimise long-term survivorship. However, a longer follow-up is needed as implant breakage is more often seen after long implantation times.</p>","PeriodicalId":12911,"journal":{"name":"HIP International","volume":" ","pages":"207-214"},"PeriodicalIF":1.3000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"HIP International","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/11207000231180065","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/6/12 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Modular stems in revision total hip arthroplasties allow diaphyseal fixation and optimal restoration of the architecture of the proximal femur. Several studies report metaphyseal implant breakage having a negative impact on survivorship. The aim of the study was to evaluate the outcome of an uncemented modular fluted tapered stem (MFT) in revision surgery.
Methods: In a retrospective study, 316 patients who had revision surgery using the same design of MFT implant (Modular Revision Stem [MRS], Lima Corporate, Italy) between 2012 and 2017 were identified. Patients were male in 51% of cases and mean age was 74 years. Indications were 110 periprosthetic fractures, 98 periprosthetic joint infection, 97 aseptic loosening, 10 instability and 1 other cause. Survivorship, complications, clinical and radiographic outcomes, were assessed. Mean follow-up was 5 years.
Results: There was no implant breakage. At 5-year follow-up, the survivorship, free from revision for aseptic loosening and free of revision for any reason were 96% and 87%, respectively. At last follow-up (8 years), these figures were 92% and 71%, respectively. 31 implants were revised. The risk of revision for any cause was higher for extreme length metaphyseal implants, hazard ratio was 3.7 (95% CI, 1.82-7.52). A mean stem subsidence of 9 mm was noted in 37 cases; 4 were revised for aseptic loosening. The Harris Hip Score at final follow-up was 82.
Conclusions: At 5-year follow-up, the MFT implant showed a good survivorship and outcomes with no specific complications. Unlike literature reports, no specific complications occurred with this design. Stem junction positioning and therefore metaphyseal length may be key to optimise long-term survivorship. However, a longer follow-up is needed as implant breakage is more often seen after long implantation times.
期刊介绍:
HIP International is the official journal of the European Hip Society. It is the only international, peer-reviewed, bi-monthly journal dedicated to diseases of the hip. HIP International considers contributions relating to hip surgery, traumatology of the hip, prosthetic surgery, biomechanics, and basic sciences relating to the hip. HIP International invites reviews from leading specialists with the aim of informing its readers of current evidence-based best practice.
The journal also publishes supplements containing proceedings of symposia, special meetings or articles of special educational merit.
HIP International is divided into six independent sections led by editors of the highest scientific merit. These sections are:
• Biomaterials
• Biomechanics
• Conservative Hip Surgery
• Paediatrics
• Primary and Revision Hip Arthroplasty
• Traumatology