Influence of stress-relieving heat treatments on the efficacy of Co-Cr-Mo-W alloy copings fabricated using selective laser melting.

IF 4.3 3区 材料科学 Q1 ENGINEERING, ELECTRICAL & ELECTRONIC ACS Applied Electronic Materials Pub Date : 2024-04-08 Epub Date: 2023-07-13 DOI:10.2186/jpr.JPR_D_22_00283
Hein Linn Htat, Atsushi Takaichi, Yuka Kajima, Nuttaphon Kittikundecha, Shingo Kamijo, Takao Hanawa, Noriyuki Wakabayashi
{"title":"Influence of stress-relieving heat treatments on the efficacy of Co-Cr-Mo-W alloy copings fabricated using selective laser melting.","authors":"Hein Linn Htat, Atsushi Takaichi, Yuka Kajima, Nuttaphon Kittikundecha, Shingo Kamijo, Takao Hanawa, Noriyuki Wakabayashi","doi":"10.2186/jpr.JPR_D_22_00283","DOIUrl":null,"url":null,"abstract":"<p><p>Purpose This study aimed to evaluate the influence of stress-relieving heat treatments on the metal-ceramic bond strength and fitness accuracy of selective laser melting (SLM)-fabricated Co-Cr alloy copings.Methods SLM-manufactured Co-Cr samples were stress-relieved at 750 (Ht-750) and 1150 °C (Ht-1150). The microstructure, surface roughness, metal-ceramic bond strength, marginal and internal fit, Vickers hardness, and residual stress were then compared with those of the non-heat-treated group (As-built). The results were analyzed using one-way ANOVA and post-hoc tests (Tukey's or Student's t test) (P = 0.05).Results The microstructure of the Ht-1150 samples had a brittle oxide layer and lower surface roughness, resulting in significantly lower bond strength values than those of the other groups. The As-built group exhibited significantly lower marginal gap values than the Ht-750 and Ht-1150 groups. Therefore, the post-heat treatments degraded the marginal fitness. The surface residual stress in all sample groups were compressive because of the sandblasting effect. The compressive stresses were larger in Ht-1150 than in As-built and Ht-750 owing to their low hardness values.Conclusions Stress-relief annealing porcelain-fused-to-metal single crowns does not improve bond strength and degrades fitness accuracy because additional post-heat treatments induce thermal distortion. These findings are expected to facilitate the direct application of As-built SLM single crowns in dentistry to minimize post-manufacturing costs and time.</p>","PeriodicalId":3,"journal":{"name":"ACS Applied Electronic Materials","volume":null,"pages":null},"PeriodicalIF":4.3000,"publicationDate":"2024-04-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Electronic Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2186/jpr.JPR_D_22_00283","RegionNum":3,"RegionCategory":"材料科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/7/13 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"ENGINEERING, ELECTRICAL & ELECTRONIC","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose This study aimed to evaluate the influence of stress-relieving heat treatments on the metal-ceramic bond strength and fitness accuracy of selective laser melting (SLM)-fabricated Co-Cr alloy copings.Methods SLM-manufactured Co-Cr samples were stress-relieved at 750 (Ht-750) and 1150 °C (Ht-1150). The microstructure, surface roughness, metal-ceramic bond strength, marginal and internal fit, Vickers hardness, and residual stress were then compared with those of the non-heat-treated group (As-built). The results were analyzed using one-way ANOVA and post-hoc tests (Tukey's or Student's t test) (P = 0.05).Results The microstructure of the Ht-1150 samples had a brittle oxide layer and lower surface roughness, resulting in significantly lower bond strength values than those of the other groups. The As-built group exhibited significantly lower marginal gap values than the Ht-750 and Ht-1150 groups. Therefore, the post-heat treatments degraded the marginal fitness. The surface residual stress in all sample groups were compressive because of the sandblasting effect. The compressive stresses were larger in Ht-1150 than in As-built and Ht-750 owing to their low hardness values.Conclusions Stress-relief annealing porcelain-fused-to-metal single crowns does not improve bond strength and degrades fitness accuracy because additional post-heat treatments induce thermal distortion. These findings are expected to facilitate the direct application of As-built SLM single crowns in dentistry to minimize post-manufacturing costs and time.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
应力消除热处理对利用选择性激光熔化技术制造的 Co-Cr-Mo-W 合金盖板功效的影响。
目的 本研究旨在评估消除应力热处理对选择性激光熔化(SLM)制造的 Co-Cr 合金盖板的金属陶瓷结合强度和配合精度的影响。方法 SLM 制造的 Co-Cr 样品在 750 (Ht-750) 和 1150 °C (Ht-1150) 下消除应力。然后将微观结构、表面粗糙度、金属陶瓷结合强度、边缘和内部配合、维氏硬度和残余应力与未热处理组(As-built)进行比较。结果 Ht-1150 样品的微观结构具有脆性氧化层和较低的表面粗糙度,因此其结合强度值明显低于其他组别。As-built组的边缘间隙值明显低于Ht-750和Ht-1150组。因此,后热处理降低了边际适配性。由于喷砂效应,所有样品组的表面残余应力都是压应力。由于硬度值较低,Ht-1150 组的压应力大于 As-built 组和 Ht-750 组。这些发现将有助于在牙科中直接应用 "原位 "SLM 单冠,以最大限度地减少后期制造成本和时间。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
7.20
自引率
4.30%
发文量
567
期刊最新文献
Hyperbaric oxygen treatment promotes tendon-bone interface healing in a rabbit model of rotator cuff tears. Oxygen-ozone therapy for myocardial ischemic stroke and cardiovascular disorders. Comparative study on the anti-inflammatory and protective effects of different oxygen therapy regimens on lipopolysaccharide-induced acute lung injury in mice. Heme oxygenase/carbon monoxide system and development of the heart. Hyperbaric oxygen for moderate-to-severe traumatic brain injury: outcomes 5-8 years after injury.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1