Association of x-ray velocimetry (XV) ventilation analysis compared to spirometry.

Jason P Kirkness, Jonathan Dusting, Nina Eikelis, Piraveen Pirakalathanan, John DeMarco, Stephen L Shiao, Andreas Fouras
{"title":"Association of x-ray velocimetry (XV) ventilation analysis compared to spirometry.","authors":"Jason P Kirkness,&nbsp;Jonathan Dusting,&nbsp;Nina Eikelis,&nbsp;Piraveen Pirakalathanan,&nbsp;John DeMarco,&nbsp;Stephen L Shiao,&nbsp;Andreas Fouras","doi":"10.3389/fmedt.2023.1148310","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>X-ray Velocimetry (XV) ventilation analysis is a 4-dimensional imaging-based method for quantifying regional ventilation, aiding in the assessment of lung function. We examined the performance characteristics of XV ventilation analysis by examining correlation to spirometry and measurement repeatability.</p><p><strong>Methods: </strong>XV analysis was assessed in 27 patients receiving thoracic radiotherapy for non-lung cancer malignancies. Measurements were obtained pre-treatment and at 4 and 12-months post-treatment. XV metrics such as ventilation defect percent (VDP) and regional ventilation heterogeneity (VH) were compared to spirometry at each time point, using correlation analysis. Repeatability was assessed between multiple runs of the analysis algorithm, as well as between multiple breaths in the same patient. Change in VH and VDP in a case series over 12 months was used to determine effect size and estimate sample sizes for future studies.</p><p><strong>Results: </strong>VDP and VH were found to significantly correlate with FEV<sub>1</sub> and FEV<sub>1</sub>/FVC (range: -0.36 to -0.57; <i>p </i>< 0.05). Repeatability tests demonstrated that VDP and VH had less than 2% variability within runs and less than 8% change in metrics between breaths. Three cases were used to illustrate the advantage of XV over spirometry, where XV indicated a change in lung function that was either undetectable or delayed in detection by spirometry. Case A demonstrated an improvement in XV metrics over time despite stable spirometric values. Case B demonstrated a decline in XV metrics as early as 4-months, although spirometric values did not change until 12-months. Case C demonstrated a decline in XV metrics at 12 months post-treatment while spirometric values remained normal throughout the study. Based on the effect sizes in each case, sample sizes ranging from 10 to 38 patients would provide 90% power for future studies aiming to detect similar changes.</p><p><strong>Conclusions: </strong>The performance and safety of XV analysis make it ideal for both clinical and research applications across most lung indications. Our results support continued research and provide a basis for powering future studies using XV as an endpoint to examine lung health and determine therapeutic efficacy.</p>","PeriodicalId":12599,"journal":{"name":"Frontiers in Medical Technology","volume":"5 ","pages":"1148310"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10335741/pdf/","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Medical Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fmedt.2023.1148310","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction: X-ray Velocimetry (XV) ventilation analysis is a 4-dimensional imaging-based method for quantifying regional ventilation, aiding in the assessment of lung function. We examined the performance characteristics of XV ventilation analysis by examining correlation to spirometry and measurement repeatability.

Methods: XV analysis was assessed in 27 patients receiving thoracic radiotherapy for non-lung cancer malignancies. Measurements were obtained pre-treatment and at 4 and 12-months post-treatment. XV metrics such as ventilation defect percent (VDP) and regional ventilation heterogeneity (VH) were compared to spirometry at each time point, using correlation analysis. Repeatability was assessed between multiple runs of the analysis algorithm, as well as between multiple breaths in the same patient. Change in VH and VDP in a case series over 12 months was used to determine effect size and estimate sample sizes for future studies.

Results: VDP and VH were found to significantly correlate with FEV1 and FEV1/FVC (range: -0.36 to -0.57; p < 0.05). Repeatability tests demonstrated that VDP and VH had less than 2% variability within runs and less than 8% change in metrics between breaths. Three cases were used to illustrate the advantage of XV over spirometry, where XV indicated a change in lung function that was either undetectable or delayed in detection by spirometry. Case A demonstrated an improvement in XV metrics over time despite stable spirometric values. Case B demonstrated a decline in XV metrics as early as 4-months, although spirometric values did not change until 12-months. Case C demonstrated a decline in XV metrics at 12 months post-treatment while spirometric values remained normal throughout the study. Based on the effect sizes in each case, sample sizes ranging from 10 to 38 patients would provide 90% power for future studies aiming to detect similar changes.

Conclusions: The performance and safety of XV analysis make it ideal for both clinical and research applications across most lung indications. Our results support continued research and provide a basis for powering future studies using XV as an endpoint to examine lung health and determine therapeutic efficacy.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
x射线测速(XV)通气分析与肺活量测定的比较。
简介:x射线测速(XV)通气分析是一种基于四维成像的量化区域通气的方法,有助于评估肺功能。我们通过检查与肺活量测定和测量重复性的相关性来检验XV通气分析的性能特征。方法:对27例接受胸部放疗的非肺癌恶性肿瘤患者进行XV分析。分别在治疗前、治疗后4个月和12个月进行测量。XV指标如通风缺陷百分比(VDP)和区域通风异质性(VH)在每个时间点与肺活量测定进行比较,采用相关性分析。在分析算法的多次运行之间以及同一患者的多次呼吸之间评估可重复性。在超过12个月的病例序列中,VH和VDP的变化被用来确定效应大小和估计未来研究的样本量。结果:VDP、VH与FEV1、FEV1/FVC显著相关(范围:-0.36 ~ -0.57;结论:XV分析的性能和安全性使其在大多数肺部适应症的临床和研究应用中都很理想。我们的结果支持继续研究,并为未来使用XV作为终点来检查肺部健康和确定治疗效果的研究提供基础。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Case Report: The effect of automated manual lymphatic drainage therapy on lymphatic contractility in 4 distinct cases Detection of natural autoimmunity to ghrelin in diabetes mellitus Ensuring safety and efficacy in combination products: regulatory challenges and best practices Augmented reality-based software (Echo-QR) for guiding the echographic probe toward the acoustic window: a pilot study Ethical considerations in the regulation and use of herbal medicines in the European Union
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1