Does continuous chemical disinfection affect mechanical properties of CAD/CAM PMMA?

IF 2.1 4区 医学 Q2 DENTISTRY, ORAL SURGERY & MEDICINE International Journal of Prosthodontics Pub Date : 2023-05-30 DOI:10.11607/ijp.8301
Oscar O Marcillo-Toala, Rodrigo Bm da Costa, Ana P Chappuis-Chocano, Helena Sandrini Venante, Joel F Junior Santiago, Vanessa S Lara, Karin H Neppelenbroek, Vinicius C Porto
{"title":"Does continuous chemical disinfection affect mechanical properties of CAD/CAM PMMA?","authors":"Oscar O Marcillo-Toala, Rodrigo Bm da Costa, Ana P Chappuis-Chocano, Helena Sandrini Venante, Joel F Junior Santiago, Vanessa S Lara, Karin H Neppelenbroek, Vinicius C Porto","doi":"10.11607/ijp.8301","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study was aimed at analyzing the effect of disinfectants on the roughness and mechanical properties of computer-aided design/computer-aided manufacturing (CAD/CAM) polymethylmethacrylate (PMMA) dentures.</p><p><strong>Materials and methods: </strong>Two groups of denture base resins were tested: heat-polymerized and milled blocks. The specimens were shaped (n=80) for flexural strength (FS), flexural modulus (FM), and (n=240) for the properties of roughness (Ra) and microhardness alterations (KHN). They were categorized into the following groups based on immersion: control (deionized water), H1 (1% sodium hypochlorite), H05 (0.5% sodium hypochlorite), and C2 (2% chlorhexidine) groups. The immersion periods were 0 (T0), 130 (T1), and 260 (T2) cycles. Statistical analyses were performed for flexural properties using the three-way analysis of variance (ANOVA). Microhardness and surface roughness were analyzed using repeatedmeasures ANOVA. A significance level of 5% was set.</p><p><strong>Results: </strong>CAD/CAM PMMA showed higher FS (<i>P</i> = .001) and FM (<i>P</i> < .001) than conventional PMMA. The KHN value was superior to the conventional PMMA (<i>P</i> < .001). The chemical solution affected the surface roughness of both resins (P = .007). The CAD/CAM PMMA block showed increased Ra values when H1 was used. Cycling separately increased the FS of conventional PMMA (T1 vs. baseline; <i>P</i> < .05). However, the FM of CAD/CAM PMMA was higher (T1 and T2 vs. baseline; <i>P</i> < .05). The time factor increased the microhardness of both resins (T2 vs. baseline; <i>P</i> < .05).</p><p><strong>Conclusion: </strong>The CAD/CAM resin showed higher values compared to conventional PMMA in all tests, regardless of the chemical solution used; however, the values obtained for both resins were clinically acceptable. Int J Prosthodont 2023. doi: 10.11607/ijp.8301.</p>","PeriodicalId":50292,"journal":{"name":"International Journal of Prosthodontics","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2023-05-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Prosthodontics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11607/ijp.8301","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study was aimed at analyzing the effect of disinfectants on the roughness and mechanical properties of computer-aided design/computer-aided manufacturing (CAD/CAM) polymethylmethacrylate (PMMA) dentures.

Materials and methods: Two groups of denture base resins were tested: heat-polymerized and milled blocks. The specimens were shaped (n=80) for flexural strength (FS), flexural modulus (FM), and (n=240) for the properties of roughness (Ra) and microhardness alterations (KHN). They were categorized into the following groups based on immersion: control (deionized water), H1 (1% sodium hypochlorite), H05 (0.5% sodium hypochlorite), and C2 (2% chlorhexidine) groups. The immersion periods were 0 (T0), 130 (T1), and 260 (T2) cycles. Statistical analyses were performed for flexural properties using the three-way analysis of variance (ANOVA). Microhardness and surface roughness were analyzed using repeatedmeasures ANOVA. A significance level of 5% was set.

Results: CAD/CAM PMMA showed higher FS (P = .001) and FM (P < .001) than conventional PMMA. The KHN value was superior to the conventional PMMA (P < .001). The chemical solution affected the surface roughness of both resins (P = .007). The CAD/CAM PMMA block showed increased Ra values when H1 was used. Cycling separately increased the FS of conventional PMMA (T1 vs. baseline; P < .05). However, the FM of CAD/CAM PMMA was higher (T1 and T2 vs. baseline; P < .05). The time factor increased the microhardness of both resins (T2 vs. baseline; P < .05).

Conclusion: The CAD/CAM resin showed higher values compared to conventional PMMA in all tests, regardless of the chemical solution used; however, the values obtained for both resins were clinically acceptable. Int J Prosthodont 2023. doi: 10.11607/ijp.8301.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
连续化学消毒是否会影响 CAD/CAM PMMA 的机械性能?
目的:本研究旨在分析消毒剂对计算机辅助设计/计算机辅助制造(CAD/CAM)聚甲基丙烯酸甲酯(PMMA)义齿的粗糙度和机械性能的影响:测试了两组义齿基托树脂:热聚合块和研磨块。对试样(n=80)进行弯曲强度(FS)、弯曲模量(FM)和粗糙度(Ra)及显微硬度变化(KHN)性能的测定。根据浸泡情况分为以下几组:对照组(去离子水)、H1 组(1% 次氯酸钠)、H05 组(0.5% 次氯酸钠)和 C2 组(2% 洗必泰)。浸泡周期分别为 0(T0)、130(T1)和 260(T2)次。采用三因素方差分析法(ANOVA)对挠曲性能进行了统计分析。显微硬度和表面粗糙度采用重复测量方差分析。显著性水平设定为 5%:与传统 PMMA 相比,CAD/CAM PMMA 显示出更高的 FS(P = .001)和 FM(P < .001)。KHN 值优于传统 PMMA(P < .001)。化学溶液影响了两种树脂的表面粗糙度(P = .007)。使用 H1 时,CAD/CAM PMMA 块的 Ra 值增加。循环分别增加了传统 PMMA 的 FS(T1 与基线相比;P < .05)。然而,CAD/CAM PMMA 的 FM 较高(T1 和 T2 与基线相比;P < .05)。时间因素增加了两种树脂的显微硬度(T2 与基线相比;P < .05):结论:与传统的 PMMA 相比,CAD/CAM 树脂在所有测试中都显示出更高的数值,无论使用哪种化学溶液;不过,两种树脂获得的数值在临床上都是可以接受的。国际口腔修复学杂志,2023 年。doi: 10.11607/ijp.8301。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
International Journal of Prosthodontics
International Journal of Prosthodontics 医学-牙科与口腔外科
CiteScore
3.20
自引率
4.30%
发文量
82
审稿时长
6-12 weeks
期刊介绍: Official Journal of the European Association for Osseointegration (EAO), the International College of Prosthodontists (ICP), the German Society of Prosthodontics and Dental Materials Science (DGPro), and the Italian Academy of Prosthetic Dentistry (AIOP) Prosthodontics demands a clinical research emphasis on patient- and dentist-mediated concerns in the management of oral rehabilitation needs. It is about making and implementing the best clinical decisions to enhance patients'' quality of life via applied biologic architecture - a role that far exceeds that of traditional prosthetic dentistry, with its emphasis on materials and techniques. The International Journal of Prosthodontics is dedicated to exploring and developing this conceptual shift in the role of today''s prosthodontist, clinician, and educator alike. The editorial board is composed of a distinguished team of leading international scholars.
期刊最新文献
A Comparison of Abutment Screw Loosening in 24-Degree Angulation-Correcting and Straight Implants: An In Vitro Study. Analysis of Topography, Flexural Strength, and Microstructure of a Lithium Disilicate Glass- Ceramic after Surface Finishing. Aging and Fracture Resistance of Screw-Retained Implant-Supported Molar Crowns Fabricated from Lithium Disilicate Containing Virgilite. Comparison Between Interocclusal Registration Using Silicone Bite Registration Material and Intraoral Scanner on Clenching Strength. Satisfaction with Labial Reinforcement of Custom-Made Mouthguards Among a Cohort of Rugby Union Players: A Randomized Crossover Trial.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1