Erroneous potassium results: preanalytical causes, detection, and corrective actions.

IF 6.6 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Critical reviews in clinical laboratory sciences Pub Date : 2023-09-01 DOI:10.1080/10408363.2023.2195936
Kathrin Schlüter, Janne Cadamuro
{"title":"Erroneous potassium results: preanalytical causes, detection, and corrective actions.","authors":"Kathrin Schlüter,&nbsp;Janne Cadamuro","doi":"10.1080/10408363.2023.2195936","DOIUrl":null,"url":null,"abstract":"<p><p>Potassium is one of the most requested laboratory tests. Its level is carefully monitored and maintained in a narrow physiological range. Even slightly altered potassium values may severely impact the patient's health, which is why an accurate and reliable result is of such importance. Even if high-quality analytics are available, there are still numerous ways in which potassium measurements may be biased, all of which occur in the preanalytical phase of the total laboratory testing process. As these results do not reflect the patient's <i>in-vivo</i> status, such results are referred to as pseudo-hyper/hypokalemia or indeed pseudo-normokalemia, depending on the true potassium result. Our goal in this review is to present an in-depth analysis of preanalytical errors that may result in inaccurate potassium results. After reviewing existing evidence on this topic, we classified preanalytical errors impacting potassium results into 4 categories: 1) patient factors like high platelet, leukocytes, or erythrocyte counts; 2) the sample type 3) the blood collection procedure, including inappropriate equipment, patient preparation, sample contamination and others and 4) the tube processing. The latter two include sample transport and storage conditions of whole blood, plasma, or serum as well as sample separation and subsequent preanalytical processes. In particular, we discuss the contribution of hemolysis, as one of the most frequent preanalytical errors, to pseudo-hyperkalemia. We provide a practical flow chart and a tabular overview of all the discussed preanalytical errors including possible underlying mechanisms, indicators for detection, suggestions for corrective actions, and references to the according evidence. We thereby hope that this manuscript will serve as a resource in the prevention and investigation of potentially biased potassium results.</p>","PeriodicalId":10760,"journal":{"name":"Critical reviews in clinical laboratory sciences","volume":"60 6","pages":"442-465"},"PeriodicalIF":6.6000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Critical reviews in clinical laboratory sciences","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/10408363.2023.2195936","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 1

Abstract

Potassium is one of the most requested laboratory tests. Its level is carefully monitored and maintained in a narrow physiological range. Even slightly altered potassium values may severely impact the patient's health, which is why an accurate and reliable result is of such importance. Even if high-quality analytics are available, there are still numerous ways in which potassium measurements may be biased, all of which occur in the preanalytical phase of the total laboratory testing process. As these results do not reflect the patient's in-vivo status, such results are referred to as pseudo-hyper/hypokalemia or indeed pseudo-normokalemia, depending on the true potassium result. Our goal in this review is to present an in-depth analysis of preanalytical errors that may result in inaccurate potassium results. After reviewing existing evidence on this topic, we classified preanalytical errors impacting potassium results into 4 categories: 1) patient factors like high platelet, leukocytes, or erythrocyte counts; 2) the sample type 3) the blood collection procedure, including inappropriate equipment, patient preparation, sample contamination and others and 4) the tube processing. The latter two include sample transport and storage conditions of whole blood, plasma, or serum as well as sample separation and subsequent preanalytical processes. In particular, we discuss the contribution of hemolysis, as one of the most frequent preanalytical errors, to pseudo-hyperkalemia. We provide a practical flow chart and a tabular overview of all the discussed preanalytical errors including possible underlying mechanisms, indicators for detection, suggestions for corrective actions, and references to the according evidence. We thereby hope that this manuscript will serve as a resource in the prevention and investigation of potentially biased potassium results.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
错误的钾结果:分析前的原因,检测和纠正措施。
钾是最需要的实验室测试之一。它的水平被仔细监测并维持在一个狭窄的生理范围内。即使是轻微的钾值变化也可能严重影响患者的健康,这就是为什么准确可靠的结果如此重要的原因。即使有高质量的分析,仍然有许多方法可以使钾测量有偏差,所有这些都发生在整个实验室测试过程的分析前阶段。由于这些结果不能反映患者的体内状态,因此根据钾的真实结果,这些结果被称为假性高/低钾血症或假性正钾血症。我们在这篇综述中的目的是对可能导致不准确的钾结果的分析前误差进行深入分析。在回顾了有关该主题的现有证据后,我们将影响钾结果的分析前错误分为4类:1)患者因素,如血小板、白细胞或红细胞计数高;2)样品类型3)采血程序,包括设备不当、患者准备、样品污染等4)管处理。后两者包括全血、血浆或血清的样品运输和储存条件,以及样品分离和随后的前分析过程。特别是,我们讨论溶血的贡献,作为最常见的分析前错误之一,假性高钾血症。我们提供了一个实用的流程图和所有讨论的分析前错误的表格概述,包括可能的潜在机制,检测指标,纠正措施建议,以及对相关证据的参考。因此,我们希望这篇手稿将作为预防和调查潜在偏钾结果的资源。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
20.00
自引率
0.00%
发文量
25
审稿时长
>12 weeks
期刊介绍: Critical Reviews in Clinical Laboratory Sciences publishes comprehensive and high quality review articles in all areas of clinical laboratory science, including clinical biochemistry, hematology, microbiology, pathology, transfusion medicine, genetics, immunology and molecular diagnostics. The reviews critically evaluate the status of current issues in the selected areas, with a focus on clinical laboratory diagnostics and latest advances. The adjective “critical” implies a balanced synthesis of results and conclusions that are frequently contradictory and controversial.
期刊最新文献
Unveiling Parkinson's disease through biomarker research: current insights and future prospects. Research progress of KL-6 in respiratory system diseases. The role of immunophenotyping in common variable immunodeficiency: a narrative review. KIF2C/MCAK a prognostic biomarker and its oncogenic potential in malignant progression, and prognosis of cancer patients: a systematic review and meta-analysis as biomarker. Extracellular vesicles in cancer: challenges and opportunities for clinical laboratories.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1