评价血糖质量的血糖危险指数与时间范围的比较。

IF 5.7 2区 医学 Q1 ENDOCRINOLOGY & METABOLISM Diabetes technology & therapeutics Pub Date : 2023-12-01 Epub Date: 2023-10-25 DOI:10.1089/dia.2023.0264
Ji Yoon Kim, Jee Hee Yoo, Jae Hyeon Kim
{"title":"评价血糖质量的血糖危险指数与时间范围的比较。","authors":"Ji Yoon Kim, Jee Hee Yoo, Jae Hyeon Kim","doi":"10.1089/dia.2023.0264","DOIUrl":null,"url":null,"abstract":"<p><p><b><i>Background:</i></b> The glycemia risk index (GRI) is a novel composite continuous glucose monitoring (CGM) metric that gives greater weight to hypoglycemia than to hyperglycemia and to extreme hypo/hyperglycemia over less extreme hypo/hyperglycemia. This study aimed at validating the effectiveness of GRI and at comparing it with time in range (TIR) in assessing glycemic quality in clinical practice. <b><i>Methods:</i></b> A total of 524 ninety-day CGM tracings of 194 insulin-treated adults with diabetes were included in the analysis. GRI was assessed according to standard metrics in ambulatory glucose profiles. Both cross-sectional and longitudinal analyses were performed to compare the GRI and TIR. <b><i>Results:</i></b> The GRI was strongly correlated not only with TIR (<i>r</i> = -0.974), but also with the coefficient of variation (<i>r</i> = 0.683). To identify whether the GRI differed by hypoglycemia even with a similar TIR, CGM tracings were grouped according to TIR (50% to <60%, 60% to <70%, 70% to <80%, and ≥80%). In each TIR group, the GRI increased as time below range (TBR)<sup><70 mg/dL</sup> increased (<i>P</i> < 0.001 for all TIR groups). In longitudinal analysis, as TBR<sup><70 mg/dL</sup> improved, the GRI improved significantly (<i>P</i> = 0.003) whereas TIR did not (<i>P</i> = 0.704). Both GRI and TIR improved as time above range (TAR)<sup>>180 mg/dL</sup> improved (<i>P</i> < 0.001 for both). The longitudinal change was easily identifiable on a GRI grid. <b><i>Conclusions:</i></b> The GRI is a useful tool for assessing glycemic quality in clinical practice and reflects hypoglycemia better than does TIR.</p>","PeriodicalId":11159,"journal":{"name":"Diabetes technology & therapeutics","volume":null,"pages":null},"PeriodicalIF":5.7000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison of Glycemia Risk Index with Time in Range for Assessing Glycemic Quality.\",\"authors\":\"Ji Yoon Kim, Jee Hee Yoo, Jae Hyeon Kim\",\"doi\":\"10.1089/dia.2023.0264\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b><i>Background:</i></b> The glycemia risk index (GRI) is a novel composite continuous glucose monitoring (CGM) metric that gives greater weight to hypoglycemia than to hyperglycemia and to extreme hypo/hyperglycemia over less extreme hypo/hyperglycemia. This study aimed at validating the effectiveness of GRI and at comparing it with time in range (TIR) in assessing glycemic quality in clinical practice. <b><i>Methods:</i></b> A total of 524 ninety-day CGM tracings of 194 insulin-treated adults with diabetes were included in the analysis. GRI was assessed according to standard metrics in ambulatory glucose profiles. Both cross-sectional and longitudinal analyses were performed to compare the GRI and TIR. <b><i>Results:</i></b> The GRI was strongly correlated not only with TIR (<i>r</i> = -0.974), but also with the coefficient of variation (<i>r</i> = 0.683). To identify whether the GRI differed by hypoglycemia even with a similar TIR, CGM tracings were grouped according to TIR (50% to <60%, 60% to <70%, 70% to <80%, and ≥80%). In each TIR group, the GRI increased as time below range (TBR)<sup><70 mg/dL</sup> increased (<i>P</i> < 0.001 for all TIR groups). In longitudinal analysis, as TBR<sup><70 mg/dL</sup> improved, the GRI improved significantly (<i>P</i> = 0.003) whereas TIR did not (<i>P</i> = 0.704). Both GRI and TIR improved as time above range (TAR)<sup>>180 mg/dL</sup> improved (<i>P</i> < 0.001 for both). The longitudinal change was easily identifiable on a GRI grid. <b><i>Conclusions:</i></b> The GRI is a useful tool for assessing glycemic quality in clinical practice and reflects hypoglycemia better than does TIR.</p>\",\"PeriodicalId\":11159,\"journal\":{\"name\":\"Diabetes technology & therapeutics\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":5.7000,\"publicationDate\":\"2023-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Diabetes technology & therapeutics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1089/dia.2023.0264\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/25 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diabetes technology & therapeutics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1089/dia.2023.0264","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/25 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

摘要

背景:血糖风险指数(GRI)是一种新的复合连续血糖监测(CGM)指标,它对低血糖的重视程度高于对高血糖的重视程度,对极端低/高血糖的关注程度高于对不太极端的低/高糖的重视程度。本研究旨在验证GRI的有效性,并将其与临床实践中评估血糖质量的时间范围(TIR)进行比较。方法:对194例接受胰岛素治疗的成年糖尿病患者进行了524次90天CGM追踪分析。GRI根据动态血糖谱中的标准指标进行评估。进行横截面和纵向分析以比较GRI和TIR。结果:GRI不仅与TIR呈正相关(r = -0.974),但也与变异系数(r = 0.683)。为了确定GRI是否因低血糖而不同,即使TIR相似,CGM追踪也根据TIR分组(50%至增加(P  GRI明显改善(P = 0.003),而TIR没有(P = 0.704)。GRI和TIR都随着时间超过范围(TAR)>180而改善 mg/dL改善(P 结论:GRI在临床实践中是评估血糖质量的有用工具,并且比TIR更好地反映低血糖。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Comparison of Glycemia Risk Index with Time in Range for Assessing Glycemic Quality.

Background: The glycemia risk index (GRI) is a novel composite continuous glucose monitoring (CGM) metric that gives greater weight to hypoglycemia than to hyperglycemia and to extreme hypo/hyperglycemia over less extreme hypo/hyperglycemia. This study aimed at validating the effectiveness of GRI and at comparing it with time in range (TIR) in assessing glycemic quality in clinical practice. Methods: A total of 524 ninety-day CGM tracings of 194 insulin-treated adults with diabetes were included in the analysis. GRI was assessed according to standard metrics in ambulatory glucose profiles. Both cross-sectional and longitudinal analyses were performed to compare the GRI and TIR. Results: The GRI was strongly correlated not only with TIR (r = -0.974), but also with the coefficient of variation (r = 0.683). To identify whether the GRI differed by hypoglycemia even with a similar TIR, CGM tracings were grouped according to TIR (50% to <60%, 60% to <70%, 70% to <80%, and ≥80%). In each TIR group, the GRI increased as time below range (TBR)<70 mg/dL increased (P < 0.001 for all TIR groups). In longitudinal analysis, as TBR<70 mg/dL improved, the GRI improved significantly (P = 0.003) whereas TIR did not (P = 0.704). Both GRI and TIR improved as time above range (TAR)>180 mg/dL improved (P < 0.001 for both). The longitudinal change was easily identifiable on a GRI grid. Conclusions: The GRI is a useful tool for assessing glycemic quality in clinical practice and reflects hypoglycemia better than does TIR.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Diabetes technology & therapeutics
Diabetes technology & therapeutics 医学-内分泌学与代谢
CiteScore
10.60
自引率
14.80%
发文量
145
审稿时长
3-8 weeks
期刊介绍: Diabetes Technology & Therapeutics is the only peer-reviewed journal providing healthcare professionals with information on new devices, drugs, drug delivery systems, and software for managing patients with diabetes. This leading international journal delivers practical information and comprehensive coverage of cutting-edge technologies and therapeutics in the field, and each issue highlights new pharmacological and device developments to optimize patient care.
期刊最新文献
GLP-1 Receptor Agonist Therapy in Cystic Fibrosis-Related Diabetes: A Case Report. Online Classification of Unstructured Free-Living Exercise Sessions in People with Type 1 Diabetes. Response to: MiniMed 780G System Outperforms Other Automated Insulin Systems Due to Algorithm Design, Not Bias-Response to Inaccurate Allegations. The Potential Impact of Continuous Glucose Monitoring Use on Diabetes-Related Attitudes and Behaviors in Adults with Type 2 Diabetes: A Qualitative Investigation of the Patient Experience. MiniMed 780G System Outperforms Other Automated Insulin Systems Due to Algorithm Design, Not Bias: Response to Inaccurate Allegations.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1