Miao Zhang, Lei Zheng, Qi Sun, Xiao Lin, Liang Ding
{"title":"[德国ICP-MS外部质量评价方案中全血铅的影响因素]。","authors":"Miao Zhang, Lei Zheng, Qi Sun, Xiao Lin, Liang Ding","doi":"10.19813/j.cnki.weishengyanjiu.2023.01.023","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To investigate the influence factors of lead in whole blood in the German External Quality Assessment Scheme(G-EQUAS) biological materials by inductively coupled plasma mass spectrometry(ICP-MS) from 2018 to 2020.</p><p><strong>Methods: </strong>ICP-MS was used to determine lead in whole blood of G-EQUAS by optimizing the reaction gas mode, sample dilution ratio, internal standard element, isotope and so on, analyzed the result.</p><p><strong>Results: </strong>This study indicated that selection of internal standard elements, isotopes, matrix effect and physical effect may affect the accuracy of ICP-MS determination of lead in biological samples.</p><p><strong>Conclusion: </strong>Avoided the factors that may affect the result when determining lead in whole blood by ICP-MS.</p>","PeriodicalId":23789,"journal":{"name":"Wei sheng yan jiu = Journal of hygiene research","volume":"52 1","pages":"136-141"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Influence factors of lead in whole blood in the German External Quality Assessment Scheme by ICP-MS].\",\"authors\":\"Miao Zhang, Lei Zheng, Qi Sun, Xiao Lin, Liang Ding\",\"doi\":\"10.19813/j.cnki.weishengyanjiu.2023.01.023\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To investigate the influence factors of lead in whole blood in the German External Quality Assessment Scheme(G-EQUAS) biological materials by inductively coupled plasma mass spectrometry(ICP-MS) from 2018 to 2020.</p><p><strong>Methods: </strong>ICP-MS was used to determine lead in whole blood of G-EQUAS by optimizing the reaction gas mode, sample dilution ratio, internal standard element, isotope and so on, analyzed the result.</p><p><strong>Results: </strong>This study indicated that selection of internal standard elements, isotopes, matrix effect and physical effect may affect the accuracy of ICP-MS determination of lead in biological samples.</p><p><strong>Conclusion: </strong>Avoided the factors that may affect the result when determining lead in whole blood by ICP-MS.</p>\",\"PeriodicalId\":23789,\"journal\":{\"name\":\"Wei sheng yan jiu = Journal of hygiene research\",\"volume\":\"52 1\",\"pages\":\"136-141\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Wei sheng yan jiu = Journal of hygiene research\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.19813/j.cnki.weishengyanjiu.2023.01.023\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Wei sheng yan jiu = Journal of hygiene research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.19813/j.cnki.weishengyanjiu.2023.01.023","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
[Influence factors of lead in whole blood in the German External Quality Assessment Scheme by ICP-MS].
Objective: To investigate the influence factors of lead in whole blood in the German External Quality Assessment Scheme(G-EQUAS) biological materials by inductively coupled plasma mass spectrometry(ICP-MS) from 2018 to 2020.
Methods: ICP-MS was used to determine lead in whole blood of G-EQUAS by optimizing the reaction gas mode, sample dilution ratio, internal standard element, isotope and so on, analyzed the result.
Results: This study indicated that selection of internal standard elements, isotopes, matrix effect and physical effect may affect the accuracy of ICP-MS determination of lead in biological samples.
Conclusion: Avoided the factors that may affect the result when determining lead in whole blood by ICP-MS.