在肯萨拉斯医院放射治疗装置中使用勇猛豹5.10在鼻咽癌直线加速器(LINAC)平面上的质量保证计划

Pandu Kurnianto, G. Maslebu, J. Muninggar, Muhammad Hidayatullah
{"title":"在肯萨拉斯医院放射治疗装置中使用勇猛豹5.10在鼻咽癌直线加速器(LINAC)平面上的质量保证计划","authors":"Pandu Kurnianto, G. Maslebu, J. Muninggar, Muhammad Hidayatullah","doi":"10.26877/asset.v5i2.15247","DOIUrl":null,"url":null,"abstract":"Radiotherapy is the treatment of cancer by using electromagnetic and particle radiation. Nasopharyngeal Cancer (NPC) is one of the difficult case to be treated in radiotherapy because its anatomical location. Precision of radiation dose is part of Quality Assurance Program as the key factor in this treatment. Thus, it is very important to ensure that the output dose of Linear Accelerator (LINAC) matched with result of Treatment Planning System (TPS). This study used Siemens LINAC Type Primus MACH Series 5633 and a set of detectors for Nasopharynx Cancer analysis with 8 field. The total dose of 5,000 cGy divided into 25 fractions with 200 cGy dose per fraction. The 8 fields are irradiated with a target on a detector device. It is then accumulated with PTW-Verisoft software by plotting the results obtained from the detector tool with PTW (phantom) which we have CT Scanned first in PTW-Verisoft. From the total detectors exposed to the radiation, the detector corresponding to PTW is 372 detectors (100%) with unsuitable detector of 0 detectors (0.00%). It is proved that the planning is 100% match for NPC with 8 fields of radiation. Thus, this method is recomended to be implemented for NPC treatment","PeriodicalId":414022,"journal":{"name":"Advance Sustainable Science Engineering and Technology","volume":"42 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-07-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Quality Assurance Plan on a Linear Accelerator (LINAC) Plane on Nasopharynx Cancer (NFC) by using Prowess Panther 5.10 at Radiotherapy Installation Ken Saras Hospital\",\"authors\":\"Pandu Kurnianto, G. Maslebu, J. Muninggar, Muhammad Hidayatullah\",\"doi\":\"10.26877/asset.v5i2.15247\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Radiotherapy is the treatment of cancer by using electromagnetic and particle radiation. Nasopharyngeal Cancer (NPC) is one of the difficult case to be treated in radiotherapy because its anatomical location. Precision of radiation dose is part of Quality Assurance Program as the key factor in this treatment. Thus, it is very important to ensure that the output dose of Linear Accelerator (LINAC) matched with result of Treatment Planning System (TPS). This study used Siemens LINAC Type Primus MACH Series 5633 and a set of detectors for Nasopharynx Cancer analysis with 8 field. The total dose of 5,000 cGy divided into 25 fractions with 200 cGy dose per fraction. The 8 fields are irradiated with a target on a detector device. It is then accumulated with PTW-Verisoft software by plotting the results obtained from the detector tool with PTW (phantom) which we have CT Scanned first in PTW-Verisoft. From the total detectors exposed to the radiation, the detector corresponding to PTW is 372 detectors (100%) with unsuitable detector of 0 detectors (0.00%). It is proved that the planning is 100% match for NPC with 8 fields of radiation. Thus, this method is recomended to be implemented for NPC treatment\",\"PeriodicalId\":414022,\"journal\":{\"name\":\"Advance Sustainable Science Engineering and Technology\",\"volume\":\"42 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-07-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Advance Sustainable Science Engineering and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.26877/asset.v5i2.15247\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advance Sustainable Science Engineering and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.26877/asset.v5i2.15247","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

放射治疗是利用电磁和粒子辐射治疗癌症。鼻咽癌因其解剖位置,是放疗治疗的难点之一。放射剂量的准确性是质量保证程序的一部分,是该治疗的关键因素。因此,确保直线加速器(Linear Accelerator, LINAC)的输出剂量与治疗计划系统(Treatment Planning System, TPS)的结果相匹配是非常重要的。本研究采用西门子LINAC型Primus MACH系列5633和一套8场鼻咽癌检测仪器。总剂量为5,000 cGy,分为25份,每份剂量为200 cGy。用探测器装置上的目标照射8个场。然后用PTW(幻影)绘制检测器工具获得的结果,我们先在PTW- verisoft中进行CT扫描,然后用PTW- verisoft软件进行累积。从暴露于辐射的探测器总数来看,PTW对应的探测器为372个(100%),不合适的探测器为0个(0.00%)。结果表明,该规划与NPC的8场辐射100%匹配。因此,这种方法被推荐用于鼻咽癌的治疗
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Quality Assurance Plan on a Linear Accelerator (LINAC) Plane on Nasopharynx Cancer (NFC) by using Prowess Panther 5.10 at Radiotherapy Installation Ken Saras Hospital
Radiotherapy is the treatment of cancer by using electromagnetic and particle radiation. Nasopharyngeal Cancer (NPC) is one of the difficult case to be treated in radiotherapy because its anatomical location. Precision of radiation dose is part of Quality Assurance Program as the key factor in this treatment. Thus, it is very important to ensure that the output dose of Linear Accelerator (LINAC) matched with result of Treatment Planning System (TPS). This study used Siemens LINAC Type Primus MACH Series 5633 and a set of detectors for Nasopharynx Cancer analysis with 8 field. The total dose of 5,000 cGy divided into 25 fractions with 200 cGy dose per fraction. The 8 fields are irradiated with a target on a detector device. It is then accumulated with PTW-Verisoft software by plotting the results obtained from the detector tool with PTW (phantom) which we have CT Scanned first in PTW-Verisoft. From the total detectors exposed to the radiation, the detector corresponding to PTW is 372 detectors (100%) with unsuitable detector of 0 detectors (0.00%). It is proved that the planning is 100% match for NPC with 8 fields of radiation. Thus, this method is recomended to be implemented for NPC treatment
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Synthesis and Characterization Materials Modern (CMC-Fe3O4-Chitosan-TiO2) As Portable Adsorbent Toxic Metal (Hg) and Dye Substance (Rh B) Implementation of DenseNet121 Architecture for Waste Type Classification Improving the Accuracy of House Price Prediction using Catboost Regression with Random Search Hyperparameter Tuning: A Comparative Analysis Harnessing Quantum SVR on Quantum Turing Machine for Drug Compounds Corrosion Inhibitors Analysis Classification of Movie Recommendation on Netflix Using Random Forest Algorithm
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1