W A Winter, M R Esseveld, P Warners, E Frensdorff, Y T King, H Kruiswijk, W P Verdegaal
{"title":"超声检测胆道梗阻的敏感性。","authors":"W A Winter, M R Esseveld, P Warners, E Frensdorff, Y T King, H Kruiswijk, W P Verdegaal","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>In 28 patient with varying degrees of biliary obstruction, the ultrasonic findings were correlated with the serum bilirubin levels and with the size of the bile ducts as measured by percutaneous transhepatic cholangiography (PTC). With ultrasound, study of the intrahepatic ducts achieved the greatest accuracy (71%); examination of common bile ducts and the gallbladder were found to be less sensitive (54 and 53%). Ultrasound proved to be very accurate in biliary obstruction, when the serum bilirubin was above 4.8 mg/100 ml (80 mumol/1).</p>","PeriodicalId":76405,"journal":{"name":"Radiologia clinica","volume":"47 5","pages":"321-9"},"PeriodicalIF":0.0000,"publicationDate":"1978-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Sensitivity of ultrasound in the detection of biliary tract obstruction.\",\"authors\":\"W A Winter, M R Esseveld, P Warners, E Frensdorff, Y T King, H Kruiswijk, W P Verdegaal\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In 28 patient with varying degrees of biliary obstruction, the ultrasonic findings were correlated with the serum bilirubin levels and with the size of the bile ducts as measured by percutaneous transhepatic cholangiography (PTC). With ultrasound, study of the intrahepatic ducts achieved the greatest accuracy (71%); examination of common bile ducts and the gallbladder were found to be less sensitive (54 and 53%). Ultrasound proved to be very accurate in biliary obstruction, when the serum bilirubin was above 4.8 mg/100 ml (80 mumol/1).</p>\",\"PeriodicalId\":76405,\"journal\":{\"name\":\"Radiologia clinica\",\"volume\":\"47 5\",\"pages\":\"321-9\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1978-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiologia clinica\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiologia clinica","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
在28例不同程度胆道梗阻患者中,超声表现与血清胆红素水平及经皮经肝胆管造影(PTC)测量的胆管大小相关。超声检查肝内导管的准确率最高(71%);检查胆总管和胆囊时发现敏感度较低(分别为54%和53%)。当血清胆红素高于4.8 mg/100 ml (80 μ mol/1)时,超声诊断胆道梗阻非常准确。
Sensitivity of ultrasound in the detection of biliary tract obstruction.
In 28 patient with varying degrees of biliary obstruction, the ultrasonic findings were correlated with the serum bilirubin levels and with the size of the bile ducts as measured by percutaneous transhepatic cholangiography (PTC). With ultrasound, study of the intrahepatic ducts achieved the greatest accuracy (71%); examination of common bile ducts and the gallbladder were found to be less sensitive (54 and 53%). Ultrasound proved to be very accurate in biliary obstruction, when the serum bilirubin was above 4.8 mg/100 ml (80 mumol/1).