{"title":"电子氡气监测仪的电磁干扰误差","authors":"L. Aciu, P. Ogrutan, G. Nicolae, C. Ursachi","doi":"10.1109/OPTIM.2014.6850914","DOIUrl":null,"url":null,"abstract":"Radon concentration measurements in air with an Electronic Radon Gas Monitor in a location close to the telecommunication antenna gave unacceptable values. This paper presents several sets of measurements of electric field at low and high frequency and Radon concentrations in studied location and at the reference location. The Radon concentration measurements were performed both with electronic devices and with the alpha track detectors. To determine susceptibility Radon detector, tests were carried out for galvanic and radiated disturbance coupling. Radiated Immunity tests were made in a TEM cell in the range 850MHz-900MHz, according to IEC 61000-4-6 in the frequency range 150kHz-80MHz. Tests revealed errors of measuring the concentration of Radon gas in the presence of disturbances.","PeriodicalId":298237,"journal":{"name":"2014 International Conference on Optimization of Electrical and Electronic Equipment (OPTIM)","volume":"12 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-05-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Errors in Electronic Radon Gas Monitors due to electromagnetic interferences\",\"authors\":\"L. Aciu, P. Ogrutan, G. Nicolae, C. Ursachi\",\"doi\":\"10.1109/OPTIM.2014.6850914\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Radon concentration measurements in air with an Electronic Radon Gas Monitor in a location close to the telecommunication antenna gave unacceptable values. This paper presents several sets of measurements of electric field at low and high frequency and Radon concentrations in studied location and at the reference location. The Radon concentration measurements were performed both with electronic devices and with the alpha track detectors. To determine susceptibility Radon detector, tests were carried out for galvanic and radiated disturbance coupling. Radiated Immunity tests were made in a TEM cell in the range 850MHz-900MHz, according to IEC 61000-4-6 in the frequency range 150kHz-80MHz. Tests revealed errors of measuring the concentration of Radon gas in the presence of disturbances.\",\"PeriodicalId\":298237,\"journal\":{\"name\":\"2014 International Conference on Optimization of Electrical and Electronic Equipment (OPTIM)\",\"volume\":\"12 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-05-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2014 International Conference on Optimization of Electrical and Electronic Equipment (OPTIM)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/OPTIM.2014.6850914\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 International Conference on Optimization of Electrical and Electronic Equipment (OPTIM)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/OPTIM.2014.6850914","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Errors in Electronic Radon Gas Monitors due to electromagnetic interferences
Radon concentration measurements in air with an Electronic Radon Gas Monitor in a location close to the telecommunication antenna gave unacceptable values. This paper presents several sets of measurements of electric field at low and high frequency and Radon concentrations in studied location and at the reference location. The Radon concentration measurements were performed both with electronic devices and with the alpha track detectors. To determine susceptibility Radon detector, tests were carried out for galvanic and radiated disturbance coupling. Radiated Immunity tests were made in a TEM cell in the range 850MHz-900MHz, according to IEC 61000-4-6 in the frequency range 150kHz-80MHz. Tests revealed errors of measuring the concentration of Radon gas in the presence of disturbances.