V Ehrhardt, H Baadenhuijsen, S Brenna, M Browne, L Garcia-Beltrán, K Hellsing, W R Külpmann, K Paschen, C Pascual-Mostaza, P Pei
{"title":"【酶活化法测定钠、钾、氯离子试剂的多中心评价结果】。","authors":"V Ehrhardt, H Baadenhuijsen, S Brenna, M Browne, L Garcia-Beltrán, K Hellsing, W R Külpmann, K Paschen, C Pascual-Mostaza, P Pei","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>We present the results of a multicentre evaluation with Boehringer Mannheim/Hitachi instruments of new \"enzymatic\" methods for the determination of Na+, K+, and Cl- in serum or plasma. The between-day coefficient of variation was less than 1.4% (Na+), less than 2.6% (K+) and less than 1.7% (Cl-). The linear range of the assays were at least 80 to 200 mmol/l (Na+), 1.5 to 17 mmol/l (K+) and about 30 to at least 200 mmol/l (Cl-). The comparisons with routine flame atomic emission spectrometry and coulometry showed a satisfactory agreement of the test results. The \"enzymatic\" assays are insensitive to even grossly elevated levels of bilirubin and lipids (sodium, potassium, and chloride assays), NH4+ (potassium assay) and amylase (chloride assay). Interference by various drugs was not detected. Since the new methods can easily be adapted to photometric clinical chemistry instruments, they represent a valuable alternative to the use of ion-selective electrodes, flame atomic emission spectrometry and coulometry.</p>","PeriodicalId":76822,"journal":{"name":"Wiener klinische Wochenschrift. Supplementum","volume":"192 ","pages":"12-21"},"PeriodicalIF":0.0000,"publicationDate":"1992-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Results of multicenter evaluation of reagents for determination of sodium, potassium and chloride ions using enzyme activation].\",\"authors\":\"V Ehrhardt, H Baadenhuijsen, S Brenna, M Browne, L Garcia-Beltrán, K Hellsing, W R Külpmann, K Paschen, C Pascual-Mostaza, P Pei\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We present the results of a multicentre evaluation with Boehringer Mannheim/Hitachi instruments of new \\\"enzymatic\\\" methods for the determination of Na+, K+, and Cl- in serum or plasma. The between-day coefficient of variation was less than 1.4% (Na+), less than 2.6% (K+) and less than 1.7% (Cl-). The linear range of the assays were at least 80 to 200 mmol/l (Na+), 1.5 to 17 mmol/l (K+) and about 30 to at least 200 mmol/l (Cl-). The comparisons with routine flame atomic emission spectrometry and coulometry showed a satisfactory agreement of the test results. The \\\"enzymatic\\\" assays are insensitive to even grossly elevated levels of bilirubin and lipids (sodium, potassium, and chloride assays), NH4+ (potassium assay) and amylase (chloride assay). Interference by various drugs was not detected. Since the new methods can easily be adapted to photometric clinical chemistry instruments, they represent a valuable alternative to the use of ion-selective electrodes, flame atomic emission spectrometry and coulometry.</p>\",\"PeriodicalId\":76822,\"journal\":{\"name\":\"Wiener klinische Wochenschrift. Supplementum\",\"volume\":\"192 \",\"pages\":\"12-21\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1992-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Wiener klinische Wochenschrift. Supplementum\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Wiener klinische Wochenschrift. Supplementum","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
[Results of multicenter evaluation of reagents for determination of sodium, potassium and chloride ions using enzyme activation].
We present the results of a multicentre evaluation with Boehringer Mannheim/Hitachi instruments of new "enzymatic" methods for the determination of Na+, K+, and Cl- in serum or plasma. The between-day coefficient of variation was less than 1.4% (Na+), less than 2.6% (K+) and less than 1.7% (Cl-). The linear range of the assays were at least 80 to 200 mmol/l (Na+), 1.5 to 17 mmol/l (K+) and about 30 to at least 200 mmol/l (Cl-). The comparisons with routine flame atomic emission spectrometry and coulometry showed a satisfactory agreement of the test results. The "enzymatic" assays are insensitive to even grossly elevated levels of bilirubin and lipids (sodium, potassium, and chloride assays), NH4+ (potassium assay) and amylase (chloride assay). Interference by various drugs was not detected. Since the new methods can easily be adapted to photometric clinical chemistry instruments, they represent a valuable alternative to the use of ion-selective electrodes, flame atomic emission spectrometry and coulometry.