尿素对绵羊体外血液循环中电催化传感器测定葡萄糖的影响。

Biomedica biochimica acta Pub Date : 1991-01-01
S Saeger, W Preidel, I von Lucadou, L Ruprecht, W Lager
{"title":"尿素对绵羊体外血液循环中电催化传感器测定葡萄糖的影响。","authors":"S Saeger,&nbsp;W Preidel,&nbsp;I von Lucadou,&nbsp;L Ruprecht,&nbsp;W Lager","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>In an animal experiment with the electrocatalytic glucose sensor, measurements were carried out over one week in the extracorporeal circulation of a sheep. Glucose tolerance tests were performed, and the influence of increased urea concentrations in the blood on the glucose determination was investigated. The sensor constructed as a flow-through cell was integrated via a vascular graft outside the body into the carotid artery of the animal and activated by an external electronic unit of measurement. The glucose concentration was determined by measuring the impedance of the electrode/membrane system at various potentials. By means of a subsequent correlation analysis of the measured values obtained over one week, a calibration valid for the entire measurement period was established. After a zero adjustment, it was even possible to adopt the calibration from the glucose measurement of the preceding animal experiment. The investigations of the influence of urea on the glucose measurement showed that the error in measurement of the sensor, which is 20% on average, is only insignificantly increased when the urea level is raised beyond the maximum physiological concentration.</p>","PeriodicalId":8948,"journal":{"name":"Biomedica biochimica acta","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1991-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Influence of urea on the glucose measurement by electrocatalytic sensor in the extracorporeal blood circulation of a sheep.\",\"authors\":\"S Saeger,&nbsp;W Preidel,&nbsp;I von Lucadou,&nbsp;L Ruprecht,&nbsp;W Lager\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In an animal experiment with the electrocatalytic glucose sensor, measurements were carried out over one week in the extracorporeal circulation of a sheep. Glucose tolerance tests were performed, and the influence of increased urea concentrations in the blood on the glucose determination was investigated. The sensor constructed as a flow-through cell was integrated via a vascular graft outside the body into the carotid artery of the animal and activated by an external electronic unit of measurement. The glucose concentration was determined by measuring the impedance of the electrode/membrane system at various potentials. By means of a subsequent correlation analysis of the measured values obtained over one week, a calibration valid for the entire measurement period was established. After a zero adjustment, it was even possible to adopt the calibration from the glucose measurement of the preceding animal experiment. The investigations of the influence of urea on the glucose measurement showed that the error in measurement of the sensor, which is 20% on average, is only insignificantly increased when the urea level is raised beyond the maximum physiological concentration.</p>\",\"PeriodicalId\":8948,\"journal\":{\"name\":\"Biomedica biochimica acta\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biomedica biochimica acta\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedica biochimica acta","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在电催化葡萄糖传感器的动物实验中,在羊的体外循环中进行了一周的测量。进行葡萄糖耐量试验,并研究血液中尿素浓度升高对葡萄糖测定的影响。该传感器是一种流动细胞,通过体外的血管移植物植入动物的颈动脉,并由外部的电子测量单元激活。葡萄糖浓度是通过测量电极/膜系统在不同电位下的阻抗来确定的。通过随后对一周内获得的测量值进行相关分析,建立了一个适用于整个测量周期的校准。经过零调整后,甚至可以采用前面动物实验中葡萄糖测量的校准。尿素对葡萄糖测量影响的研究表明,当尿素水平超过最大生理浓度时,传感器的测量误差仅轻微增加,平均为20%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Influence of urea on the glucose measurement by electrocatalytic sensor in the extracorporeal blood circulation of a sheep.

In an animal experiment with the electrocatalytic glucose sensor, measurements were carried out over one week in the extracorporeal circulation of a sheep. Glucose tolerance tests were performed, and the influence of increased urea concentrations in the blood on the glucose determination was investigated. The sensor constructed as a flow-through cell was integrated via a vascular graft outside the body into the carotid artery of the animal and activated by an external electronic unit of measurement. The glucose concentration was determined by measuring the impedance of the electrode/membrane system at various potentials. By means of a subsequent correlation analysis of the measured values obtained over one week, a calibration valid for the entire measurement period was established. After a zero adjustment, it was even possible to adopt the calibration from the glucose measurement of the preceding animal experiment. The investigations of the influence of urea on the glucose measurement showed that the error in measurement of the sensor, which is 20% on average, is only insignificantly increased when the urea level is raised beyond the maximum physiological concentration.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
But what makes it doctoral?: taking on the traditionalists: interdisciplinary, practice-led doctoral research in the creative industries: a case study in academic politics, research, rigour and relevance How Homelessness Compromises the Exercise of the Rights of Citizenship in Australia XIth International Karlsburg Symposium on Problems of Diabetes. September 19-21, 1983. Phosphoinositide signalling system in red blood cells of patients with hereditary spherocytosis. Improved purification and characterization of membraneous and cytosolic inositol phospholipid-specific phospholipases C from porcine brain cortex.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1