用声学方法测定探针管插入深度的实耳-耦合器差的可重复性

T. Ching, L. Britton
{"title":"用声学方法测定探针管插入深度的实耳-耦合器差的可重复性","authors":"T. Ching, L. Britton","doi":"10.1375/AUDI.30.2.91","DOIUrl":null,"url":null,"abstract":"This article reports the real-ear-to-coupler differences (RECD) of 24 children (46 ears) and the repeatability of measures for 30 ears. For measuring real-ear sound pressure level (SPL), the probe tube must be positioned close to the eardrum in order to be representative of the SPL at the eardrum. In this study, an acoustic method that used a 6 kHz standing wave minimum in the real ear to place the probe tube tip at 9 mm from the eardrum was used with custom earmoulds for measuring the real-ear portion of the RECD. The results showed a mean insertion depth from the inter-tragal notch of 24.4 mm. Across frequencies between 0.25 to 4 kHz, the test-retest difference was less than 1.0 dB, and the reliability coefficients ranged from 0.7 to 0.9. The finding on insertion depths from the inter-tragal notch confirms the appropriateness of recommendations for an insertion depth of 25 mm in measuring RECD of children of this age. Variability in measured insertion depths across individuals to achieve the same relative distance from the eardrums supports the use of the 6 kHz notch method for accurate real-ear SPL measurements. Intersubject variability is several times greater than variability associated with repeated measurements of a single subject. The repeatability of this method compares favourably with previous literature on measures using foam/immitance tips with constant probe tube insertion depth.","PeriodicalId":114768,"journal":{"name":"Australian and New Zealand Journal of Audiology","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2008-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Repeatability of Real-ear-to-coupler Differences Measured by an Acoustic Method for Determining Probe Tube Insertion Depth\",\"authors\":\"T. Ching, L. Britton\",\"doi\":\"10.1375/AUDI.30.2.91\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This article reports the real-ear-to-coupler differences (RECD) of 24 children (46 ears) and the repeatability of measures for 30 ears. For measuring real-ear sound pressure level (SPL), the probe tube must be positioned close to the eardrum in order to be representative of the SPL at the eardrum. In this study, an acoustic method that used a 6 kHz standing wave minimum in the real ear to place the probe tube tip at 9 mm from the eardrum was used with custom earmoulds for measuring the real-ear portion of the RECD. The results showed a mean insertion depth from the inter-tragal notch of 24.4 mm. Across frequencies between 0.25 to 4 kHz, the test-retest difference was less than 1.0 dB, and the reliability coefficients ranged from 0.7 to 0.9. The finding on insertion depths from the inter-tragal notch confirms the appropriateness of recommendations for an insertion depth of 25 mm in measuring RECD of children of this age. Variability in measured insertion depths across individuals to achieve the same relative distance from the eardrums supports the use of the 6 kHz notch method for accurate real-ear SPL measurements. Intersubject variability is several times greater than variability associated with repeated measurements of a single subject. The repeatability of this method compares favourably with previous literature on measures using foam/immitance tips with constant probe tube insertion depth.\",\"PeriodicalId\":114768,\"journal\":{\"name\":\"Australian and New Zealand Journal of Audiology\",\"volume\":\"1 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2008-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Australian and New Zealand Journal of Audiology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1375/AUDI.30.2.91\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australian and New Zealand Journal of Audiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1375/AUDI.30.2.91","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

本文报道了24例儿童(46耳)的实际耳-耦合器差异(RECD)和30耳测量的可重复性。测量实耳声压级(SPL)时,探头必须靠近鼓膜,以代表鼓膜处的声压级。在本研究中,使用一种声学方法,在真耳中使用6 kHz驻波最小值,将探针管尖端放置在距离鼓膜9毫米的位置,并使用定制耳模测量RECD的实耳部分。结果显示,平均插入深度为24.4毫米。在0.25 ~ 4 kHz频率范围内,重测差异小于1.0 dB,信度系数范围为0.7 ~ 0.9。从瓣间缺口的插入深度的发现证实了在测量这个年龄段儿童的RECD时,推荐的插入深度为25mm的适宜性。在个体之间测量插入深度的可变性,以达到与鼓膜相同的相对距离,支持使用6 kHz陷波法进行准确的实耳声压级测量。被试间的变异性比单一被试重复测量的变异性大几倍。该方法的可重复性优于先前使用泡沫/阻抗尖端恒定探头插入深度的测量文献。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Repeatability of Real-ear-to-coupler Differences Measured by an Acoustic Method for Determining Probe Tube Insertion Depth
This article reports the real-ear-to-coupler differences (RECD) of 24 children (46 ears) and the repeatability of measures for 30 ears. For measuring real-ear sound pressure level (SPL), the probe tube must be positioned close to the eardrum in order to be representative of the SPL at the eardrum. In this study, an acoustic method that used a 6 kHz standing wave minimum in the real ear to place the probe tube tip at 9 mm from the eardrum was used with custom earmoulds for measuring the real-ear portion of the RECD. The results showed a mean insertion depth from the inter-tragal notch of 24.4 mm. Across frequencies between 0.25 to 4 kHz, the test-retest difference was less than 1.0 dB, and the reliability coefficients ranged from 0.7 to 0.9. The finding on insertion depths from the inter-tragal notch confirms the appropriateness of recommendations for an insertion depth of 25 mm in measuring RECD of children of this age. Variability in measured insertion depths across individuals to achieve the same relative distance from the eardrums supports the use of the 6 kHz notch method for accurate real-ear SPL measurements. Intersubject variability is several times greater than variability associated with repeated measurements of a single subject. The repeatability of this method compares favourably with previous literature on measures using foam/immitance tips with constant probe tube insertion depth.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Rehabilitation Efforts and Stress in Parents of Children With Cochlear Implants Communicating with Australian Deaf People about Communication Technology Student Audiologists' Impressions of a Simulation Training Program An Independent Review of Neuromonics Tinnitus Treatment Controlled Clinical Trials A Pilot Investigation into the Provision of Hearing Services Using Tele-audiology to Remote Areas
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1