应用六西格玛指标评价临床化学实验室检测性能的初步研究

Pier Angeli Medina, J. Matibag, Sarah Jane Datay-Lim, Elizabeth Arcellana-Nuqui
{"title":"应用六西格玛指标评价临床化学实验室检测性能的初步研究","authors":"Pier Angeli Medina, J. Matibag, Sarah Jane Datay-Lim, Elizabeth Arcellana-Nuqui","doi":"10.21141/pjp.2019.13","DOIUrl":null,"url":null,"abstract":"Introduction Six sigma has been used over the years, initially in manufacturing industries to improve quality by reducing the number of wastes and defects. In the laboratory, it can also provide measurement of quality using the sigma scale. Objectives The main objective of the study is to evaluate the performance of tests in two chemistry analyzers using the six sigma scale. Methodology A total of twenty (28) tests were evaluated on two Abbott Architect c8000 chemistry analyzers from September 2014 to July 2019 using results of quality control mean, coefficient of variation, bias and total allowable error to compute for the six sigma value. Both level one and level two third party quality controls were included in the evaluation. Results OResults of the study showed the tests that were >6 sigma for both levels 1 and 2 throughout the 5 years. Di-Bil, CK, HLD, TG and UA were consistently >6 sigma for one machine while CK, Di-Bil, HDL, Mg, TG and UA were consistently >6 sigma for the other. Level 1 and Level 2 sigma scores were noted to be incongruent in some analytes as follows: ALB, ALT, K, TP for one instrument and ALB, ALP and AST for the other instrument. Electrolytes Ca, Cl, and Na were generally low ( Conclusion Using six sigma metrics allowed the laboratory to evaluate the performance of the chemistry tests objectively. Tests that are >6.0 sigma signifies world class performance and entail application of fewer Westgard rules with fewer number of runs while those that are","PeriodicalId":166708,"journal":{"name":"Philippine Journal of Pathology","volume":"1245 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-12-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"A Pilot Study on the Evaluation of Clinical Chemistry Laboratory Test Performance using Six Sigma Metrics\",\"authors\":\"Pier Angeli Medina, J. Matibag, Sarah Jane Datay-Lim, Elizabeth Arcellana-Nuqui\",\"doi\":\"10.21141/pjp.2019.13\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction Six sigma has been used over the years, initially in manufacturing industries to improve quality by reducing the number of wastes and defects. In the laboratory, it can also provide measurement of quality using the sigma scale. Objectives The main objective of the study is to evaluate the performance of tests in two chemistry analyzers using the six sigma scale. Methodology A total of twenty (28) tests were evaluated on two Abbott Architect c8000 chemistry analyzers from September 2014 to July 2019 using results of quality control mean, coefficient of variation, bias and total allowable error to compute for the six sigma value. Both level one and level two third party quality controls were included in the evaluation. Results OResults of the study showed the tests that were >6 sigma for both levels 1 and 2 throughout the 5 years. Di-Bil, CK, HLD, TG and UA were consistently >6 sigma for one machine while CK, Di-Bil, HDL, Mg, TG and UA were consistently >6 sigma for the other. Level 1 and Level 2 sigma scores were noted to be incongruent in some analytes as follows: ALB, ALT, K, TP for one instrument and ALB, ALP and AST for the other instrument. Electrolytes Ca, Cl, and Na were generally low ( Conclusion Using six sigma metrics allowed the laboratory to evaluate the performance of the chemistry tests objectively. Tests that are >6.0 sigma signifies world class performance and entail application of fewer Westgard rules with fewer number of runs while those that are\",\"PeriodicalId\":166708,\"journal\":{\"name\":\"Philippine Journal of Pathology\",\"volume\":\"1245 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-12-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Philippine Journal of Pathology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.21141/pjp.2019.13\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Philippine Journal of Pathology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.21141/pjp.2019.13","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

六西格玛已经使用了多年,最初是在制造业中,通过减少浪费和缺陷的数量来提高质量。在实验室中,它还可以使用西格玛标度提供质量测量。本研究的主要目的是使用六西格玛量表评估两种化学分析仪的测试性能。2014年9月至2019年7月,对两台雅培Architect c8000化学分析仪进行了共20(28)次测试,使用质量控制平均值、变异系数、偏差和总允许误差的结果来计算六西格玛值。评估中包括一级和二级第三方质量控制。结果研究结果显示,在整个5年中,1级和2级的测试均>6西格玛。一台机器的Di-Bil、CK、HLD、TG和UA一致>6西格玛,而另一台机器的CK、Di-Bil、HDL、Mg、TG和UA一致>6西格玛。1级和2级西格玛评分在以下分析中存在不一致:一种仪器的ALB、ALT、K、TP和另一种仪器的ALB、ALP和AST。电解质Ca、Cl和Na普遍较低(结论使用六西格玛指标可以使实验室客观地评估化学测试的性能。>6.0 sigma的测试意味着世界级的性能,并且需要应用较少的Westgard规则和较少的运行次数
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A Pilot Study on the Evaluation of Clinical Chemistry Laboratory Test Performance using Six Sigma Metrics
Introduction Six sigma has been used over the years, initially in manufacturing industries to improve quality by reducing the number of wastes and defects. In the laboratory, it can also provide measurement of quality using the sigma scale. Objectives The main objective of the study is to evaluate the performance of tests in two chemistry analyzers using the six sigma scale. Methodology A total of twenty (28) tests were evaluated on two Abbott Architect c8000 chemistry analyzers from September 2014 to July 2019 using results of quality control mean, coefficient of variation, bias and total allowable error to compute for the six sigma value. Both level one and level two third party quality controls were included in the evaluation. Results OResults of the study showed the tests that were >6 sigma for both levels 1 and 2 throughout the 5 years. Di-Bil, CK, HLD, TG and UA were consistently >6 sigma for one machine while CK, Di-Bil, HDL, Mg, TG and UA were consistently >6 sigma for the other. Level 1 and Level 2 sigma scores were noted to be incongruent in some analytes as follows: ALB, ALT, K, TP for one instrument and ALB, ALP and AST for the other instrument. Electrolytes Ca, Cl, and Na were generally low ( Conclusion Using six sigma metrics allowed the laboratory to evaluate the performance of the chemistry tests objectively. Tests that are >6.0 sigma signifies world class performance and entail application of fewer Westgard rules with fewer number of runs while those that are
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Defining the Terms of Reference of the PSP’s Committee on Academic and Research Pathology Mucosal Melanoma of the Male Urethra: A Case Report Correlation of Clinicopathologic Features of Filipino Primary Breast Cancer Patients with HER2 Subgroups Classified according to the ASCO/CAP 2018 Breast Cancer HER2 Testing Guidelines Needs Assessment for Establishment of Telepathology in the Philippines Ethical Concerns and Recommendations for Sharing Anatomic Pathology Images on Online Social Media Networks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1