新型Sysmex XR血液分析仪的自动体液分析性能评估

IF 1.1 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY Journal of Laboratory Medicine Pub Date : 2023-11-13 DOI:10.1515/labmed-2023-0064
Christine Van Laer, Lieselot Dedeene, Lien Gruwier, Bram Vanmechelen, Talent Hwandih, Konstantinos Mintzas, Marion Münster, Nancy Boeckx
{"title":"新型Sysmex XR血液分析仪的自动体液分析性能评估","authors":"Christine Van Laer, Lieselot Dedeene, Lien Gruwier, Bram Vanmechelen, Talent Hwandih, Konstantinos Mintzas, Marion Münster, Nancy Boeckx","doi":"10.1515/labmed-2023-0064","DOIUrl":null,"url":null,"abstract":"Abstract Objectives The cellular composition of body fluids (BF) provides insight into disease pathology and is an important diagnostic parameter. The well-established Sysmex XN haematology analyser (XN) offers automated BF analysis and is in routine use in many laboratories. In this study the performance of the new Sysmex XR analyser (XR) in testing BF is compared to the XN. Methods Cerebrospinal, pleural, peritoneal, and synovial fluids, as well as bronchoalveolar lavage (BAL) samples, were processed in BF mode on both analysers. Standard statistical methods were used to evaluate the performance of XR compared to the reference XN. Results Cell counts were generated from a total of 356 measurements from 307 patient BF samples. After application of exclusion criteria, 86 cerebrospinal, 77 peritoneal, 78 pleural, and 89 synovial fluid sample pairs were compared. An excellent correlation coefficient (r) between the two analysers was observed for all diagnostic parameters; WBC-BF: 0.996, RBC-BF: 0.974, TC-BF: 0.995, MN#: 0.994, MN%: 0.830, PMN#: 0.985 and PMN%: 0.835. The research parameters also performed equally well (r values of 0.900–0.994). The performance of diluted and undiluted samples was equally good. Similar results for diagnostic and research parameters were obtained from 62 BAL samples (r values of 0.899–0.992). Conclusions This study demonstrates that the automated BF counts on the new Sysmex XR analyser are equivalent to that of the XN for all BF types analysed.","PeriodicalId":55986,"journal":{"name":"Journal of Laboratory Medicine","volume":"53 33","pages":"0"},"PeriodicalIF":1.1000,"publicationDate":"2023-11-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance evaluation of the automated body fluid analysis of the new Sysmex XR haematology analyser\",\"authors\":\"Christine Van Laer, Lieselot Dedeene, Lien Gruwier, Bram Vanmechelen, Talent Hwandih, Konstantinos Mintzas, Marion Münster, Nancy Boeckx\",\"doi\":\"10.1515/labmed-2023-0064\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Objectives The cellular composition of body fluids (BF) provides insight into disease pathology and is an important diagnostic parameter. The well-established Sysmex XN haematology analyser (XN) offers automated BF analysis and is in routine use in many laboratories. In this study the performance of the new Sysmex XR analyser (XR) in testing BF is compared to the XN. Methods Cerebrospinal, pleural, peritoneal, and synovial fluids, as well as bronchoalveolar lavage (BAL) samples, were processed in BF mode on both analysers. Standard statistical methods were used to evaluate the performance of XR compared to the reference XN. Results Cell counts were generated from a total of 356 measurements from 307 patient BF samples. After application of exclusion criteria, 86 cerebrospinal, 77 peritoneal, 78 pleural, and 89 synovial fluid sample pairs were compared. An excellent correlation coefficient (r) between the two analysers was observed for all diagnostic parameters; WBC-BF: 0.996, RBC-BF: 0.974, TC-BF: 0.995, MN#: 0.994, MN%: 0.830, PMN#: 0.985 and PMN%: 0.835. The research parameters also performed equally well (r values of 0.900–0.994). The performance of diluted and undiluted samples was equally good. Similar results for diagnostic and research parameters were obtained from 62 BAL samples (r values of 0.899–0.992). Conclusions This study demonstrates that the automated BF counts on the new Sysmex XR analyser are equivalent to that of the XN for all BF types analysed.\",\"PeriodicalId\":55986,\"journal\":{\"name\":\"Journal of Laboratory Medicine\",\"volume\":\"53 33\",\"pages\":\"0\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2023-11-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1515/labmed-2023-0064\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/labmed-2023-0064","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

【摘要】目的体液的细胞组成(BF)为了解疾病病理提供了依据,是一个重要的诊断参数。成熟的Sysmex XN血液学分析仪(XN)提供自动化BF分析,并在许多实验室中常规使用。在本研究中,比较了Sysmex新型XR分析仪(XR)与XN检测BF的性能。方法在这两种分析仪上以BF模式处理脑脊髓、胸膜、腹膜和滑液以及支气管肺泡灌洗液(BAL)样本。采用标准统计方法比较XR与参考XN的性能。结果从307例患者BF样本中共356次测量产生细胞计数。应用排除标准后,对86对脑脊液、77对腹膜、78对胸膜和89对滑液样本进行比较。两种分析仪之间的所有诊断参数均具有良好的相关系数(r);WBC-BF: 0.996, RBC-BF: 0.974, TC-BF: 0.995, MN#: 0.994, MN%: 0.830, PMN#: 0.985, PMN%: 0.835。研究参数也同样表现良好(r值为0.900 ~ 0.994)。稀释样品和未稀释样品的性能同样好。62份BAL样本的诊断参数和研究参数结果相似(r值为0.899 ~ 0.992)。本研究表明,新Sysmex XR分析仪上的自动BF计数与所分析的所有BF类型的XN相等。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Performance evaluation of the automated body fluid analysis of the new Sysmex XR haematology analyser
Abstract Objectives The cellular composition of body fluids (BF) provides insight into disease pathology and is an important diagnostic parameter. The well-established Sysmex XN haematology analyser (XN) offers automated BF analysis and is in routine use in many laboratories. In this study the performance of the new Sysmex XR analyser (XR) in testing BF is compared to the XN. Methods Cerebrospinal, pleural, peritoneal, and synovial fluids, as well as bronchoalveolar lavage (BAL) samples, were processed in BF mode on both analysers. Standard statistical methods were used to evaluate the performance of XR compared to the reference XN. Results Cell counts were generated from a total of 356 measurements from 307 patient BF samples. After application of exclusion criteria, 86 cerebrospinal, 77 peritoneal, 78 pleural, and 89 synovial fluid sample pairs were compared. An excellent correlation coefficient (r) between the two analysers was observed for all diagnostic parameters; WBC-BF: 0.996, RBC-BF: 0.974, TC-BF: 0.995, MN#: 0.994, MN%: 0.830, PMN#: 0.985 and PMN%: 0.835. The research parameters also performed equally well (r values of 0.900–0.994). The performance of diluted and undiluted samples was equally good. Similar results for diagnostic and research parameters were obtained from 62 BAL samples (r values of 0.899–0.992). Conclusions This study demonstrates that the automated BF counts on the new Sysmex XR analyser are equivalent to that of the XN for all BF types analysed.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Laboratory Medicine
Journal of Laboratory Medicine Mathematics-Discrete Mathematics and Combinatorics
CiteScore
2.50
自引率
0.00%
发文量
39
审稿时长
10 weeks
期刊介绍: The Journal of Laboratory Medicine (JLM) is a bi-monthly published journal that reports on the latest developments in laboratory medicine. Particular focus is placed on the diagnostic aspects of the clinical laboratory, although technical, regulatory, and educational topics are equally covered. The Journal specializes in the publication of high-standard, competent and timely review articles on clinical, methodological and pathogenic aspects of modern laboratory diagnostics. These reviews are critically reviewed by expert reviewers and JLM’s Associate Editors who are specialists in the various subdisciplines of laboratory medicine. In addition, JLM publishes original research articles, case reports, point/counterpoint articles and letters to the editor, all of which are peer reviewed by at least two experts in the field.
期刊最新文献
Assessing the stability of uncentrifuged serum and plasma analytes at various post-collection intervals How Gaussian mixture modelling can help to verify reference intervals from laboratory data with a high proportion of pathological values Using machine learning techniques for exploration and classification of laboratory data Automated sex and age partitioning for the estimation of reference intervals using a regression tree model Serum LDH and its isoenzymes (LDH2 and LDH5) associated with predictive value for refractory mycoplasma pneumoniae pneumonia in children
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1