{"title":"基于菱形机构的骨传导植入体压电换能器的磁共振相容性改善设计","authors":"Dong-Ho Shin, Hyung-Gyu Lim, Myoung-Nam Kim, Ki-Woong Seong","doi":"10.9717/kmms.2023.26.8.956","DOIUrl":null,"url":null,"abstract":"This study introduces a novel piezoelectric transducer for bone conduction implants that combines piezoelectric elements with a rhombus mechanism to enhance compatibility with magnetic resonance environments. To derive the optimal design of the rhombus structure, various parameters were investigated using theoretical analysis and finite element analysis. A theoretical model of the rhombus structure was employed to identify the parameters affecting displacement amplification magnitude. Based on this, a parametric analysis was performed to calculate the displacement amplification ratio according to these parameters. The results showed that as the beam thickness and width increased, the amplification ratio reduced, while with an increased length, the amplification ratio was increased. Therefore, the optimal rhombus structure for the transducer featured beam dimensions of 0.15 mm thickness, 2 mm width, 3.5 mm length, and 5.5° inclination. This configuration amplified piezoelectric element displacement by a factor of 7.02. The amplification ratio remained constant as long as the mass applied to the rhombus frame to control mechanical resonance did not exceed the blocking force of the piezoelectric element. When a mass of 0.3 g was applied to the frame, mechanical resonance occurred at a frequency of 2 kHz, making it suitable as a transducer for bone conduction implants.","PeriodicalId":16316,"journal":{"name":"Journal of Korea Multimedia Society","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Design of Bone Conduction Implants Piezoelectric Transducer Based on Rhombus Mechanism for Magnetic Resonance Compatibility Improvement\",\"authors\":\"Dong-Ho Shin, Hyung-Gyu Lim, Myoung-Nam Kim, Ki-Woong Seong\",\"doi\":\"10.9717/kmms.2023.26.8.956\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This study introduces a novel piezoelectric transducer for bone conduction implants that combines piezoelectric elements with a rhombus mechanism to enhance compatibility with magnetic resonance environments. To derive the optimal design of the rhombus structure, various parameters were investigated using theoretical analysis and finite element analysis. A theoretical model of the rhombus structure was employed to identify the parameters affecting displacement amplification magnitude. Based on this, a parametric analysis was performed to calculate the displacement amplification ratio according to these parameters. The results showed that as the beam thickness and width increased, the amplification ratio reduced, while with an increased length, the amplification ratio was increased. Therefore, the optimal rhombus structure for the transducer featured beam dimensions of 0.15 mm thickness, 2 mm width, 3.5 mm length, and 5.5° inclination. This configuration amplified piezoelectric element displacement by a factor of 7.02. The amplification ratio remained constant as long as the mass applied to the rhombus frame to control mechanical resonance did not exceed the blocking force of the piezoelectric element. When a mass of 0.3 g was applied to the frame, mechanical resonance occurred at a frequency of 2 kHz, making it suitable as a transducer for bone conduction implants.\",\"PeriodicalId\":16316,\"journal\":{\"name\":\"Journal of Korea Multimedia Society\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-08-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Korea Multimedia Society\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.9717/kmms.2023.26.8.956\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Korea Multimedia Society","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.9717/kmms.2023.26.8.956","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本研究介绍了一种新型骨传导植入物的压电换能器,该换能器将压电元件与菱形机构相结合,以增强与磁共振环境的兼容性。采用理论分析和有限元分析相结合的方法对各参数进行了研究,得出了菱形结构的最优设计方案。采用菱形结构的理论模型,确定了影响位移放大幅度的参数。在此基础上,进行了参数化分析,根据这些参数计算位移放大比。结果表明:随着光束厚度和宽度的增加,放大比减小,随着光束长度的增加,放大比增大;因此,换能器的最佳菱形结构的光束尺寸为0.15 mm厚,2mm宽,3.5 mm长,5.5°倾角。这种结构将压电元件的位移放大了7.02倍。只要施加在菱形框架上控制机械共振的质量不超过压电元件的阻挡力,放大比保持不变。当0.3 g的质量施加到框架上时,机械共振频率为2khz,使其适合作为骨传导植入物的换能器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Design of Bone Conduction Implants Piezoelectric Transducer Based on Rhombus Mechanism for Magnetic Resonance Compatibility Improvement
This study introduces a novel piezoelectric transducer for bone conduction implants that combines piezoelectric elements with a rhombus mechanism to enhance compatibility with magnetic resonance environments. To derive the optimal design of the rhombus structure, various parameters were investigated using theoretical analysis and finite element analysis. A theoretical model of the rhombus structure was employed to identify the parameters affecting displacement amplification magnitude. Based on this, a parametric analysis was performed to calculate the displacement amplification ratio according to these parameters. The results showed that as the beam thickness and width increased, the amplification ratio reduced, while with an increased length, the amplification ratio was increased. Therefore, the optimal rhombus structure for the transducer featured beam dimensions of 0.15 mm thickness, 2 mm width, 3.5 mm length, and 5.5° inclination. This configuration amplified piezoelectric element displacement by a factor of 7.02. The amplification ratio remained constant as long as the mass applied to the rhombus frame to control mechanical resonance did not exceed the blocking force of the piezoelectric element. When a mass of 0.3 g was applied to the frame, mechanical resonance occurred at a frequency of 2 kHz, making it suitable as a transducer for bone conduction implants.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Usability Study of GAN-based Webtoon Background Image Data Augmentation A Smart Sensor for Sleep Posture Measurement Using Pressure Sensors LNG and HFO Fuel Consumption Forecasting Modeling Using LightGBM Input Data Processing Methods to Improve Point Cloud Completion Model for Dental Prosthesis Low-Resolution Image Upsampling Method Using Super Resolution Based Adaptive Pixel Shuffle
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1