{"title":"确定终末期肺部疾病肺低温活检的最佳设置。人体,离体,病变肺临床试验。","authors":"Vanessa Menezes, Clare Pollock, Pasquale Ferraro, Basil Nasir, Charles Leduc, Julie Morisset, Moishe Liberman","doi":"10.1097/LBR.0000000000000948","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To evaluate optimal settings of probe size, freezing time, and distance to the pleura that influence the size and quality of biopsy specimens during transbronchial lung cryobiopsies in ESPD.</p><p><strong>Methods: </strong>We prospectively recruited 17 patients undergoing lung transplantation. We created a nonperfused ex vivo bronchoscopy setting to perform multiple cryobiopsies with different probe sizes (1.7, 1.9, and 2.4 mm), freezing times (3, 5, 7, 10, 20, 30 seconds), and probe distance from pleura (5, 10, and 20 mm). Alveolated pulmonary parenchyma area≥50% in histology was considered a good quality biopsy, with a minimum procedural artifact. We used logistic regression to identify independent parameters as risk factors for histologic adequacy.</p><p><strong>Results: </strong>A total of 545 cryobiopsies were obtained from 34 explanted lungs after pneumonectomy for lung transplantation. The mean maximum diameter of the specimen achieved with the 1.7 probe was larger (13.5 mm) than those obtained with 1.9 and 2.4 mm probes (11.3 and 10.7 mm, P= 0.07). More pleural macroscopic damage and pleural tissue in histology occurred with the 2.4 mm probe ( P <0.001). There was no difference in the quality of specimens between the different freezing times and the distance from the pleura.</p><p><strong>Conclusions: </strong>Freezing time and distance from the pleura did not affect the histologic quality for diagnosing ESPD in severely damaged lungs. Smaller cryoprobe size did not negatively affect sample adequacy.</p>","PeriodicalId":15268,"journal":{"name":"Journal of Bronchology & Interventional Pulmonology","volume":null,"pages":null},"PeriodicalIF":3.3000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Defining Optimal Settings for Lung Cryobiopsy in End-Stage Pulmonary Disease. A Human, Ex Vivo, Diseased Lung Clinical Trial.\",\"authors\":\"Vanessa Menezes, Clare Pollock, Pasquale Ferraro, Basil Nasir, Charles Leduc, Julie Morisset, Moishe Liberman\",\"doi\":\"10.1097/LBR.0000000000000948\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>To evaluate optimal settings of probe size, freezing time, and distance to the pleura that influence the size and quality of biopsy specimens during transbronchial lung cryobiopsies in ESPD.</p><p><strong>Methods: </strong>We prospectively recruited 17 patients undergoing lung transplantation. We created a nonperfused ex vivo bronchoscopy setting to perform multiple cryobiopsies with different probe sizes (1.7, 1.9, and 2.4 mm), freezing times (3, 5, 7, 10, 20, 30 seconds), and probe distance from pleura (5, 10, and 20 mm). Alveolated pulmonary parenchyma area≥50% in histology was considered a good quality biopsy, with a minimum procedural artifact. We used logistic regression to identify independent parameters as risk factors for histologic adequacy.</p><p><strong>Results: </strong>A total of 545 cryobiopsies were obtained from 34 explanted lungs after pneumonectomy for lung transplantation. The mean maximum diameter of the specimen achieved with the 1.7 probe was larger (13.5 mm) than those obtained with 1.9 and 2.4 mm probes (11.3 and 10.7 mm, P= 0.07). More pleural macroscopic damage and pleural tissue in histology occurred with the 2.4 mm probe ( P <0.001). There was no difference in the quality of specimens between the different freezing times and the distance from the pleura.</p><p><strong>Conclusions: </strong>Freezing time and distance from the pleura did not affect the histologic quality for diagnosing ESPD in severely damaged lungs. Smaller cryoprobe size did not negatively affect sample adequacy.</p>\",\"PeriodicalId\":15268,\"journal\":{\"name\":\"Journal of Bronchology & Interventional Pulmonology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.3000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Bronchology & Interventional Pulmonology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1097/LBR.0000000000000948\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RESPIRATORY SYSTEM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Bronchology & Interventional Pulmonology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/LBR.0000000000000948","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0
摘要
背景:评估ESPD经支气管肺低温活检中探针大小、冷冻时间和胸膜距离对活检标本大小和质量的影响。方法:前瞻性招募17例肺移植患者。我们创建了一个非灌注的离体支气管镜设置,使用不同的探针大小(1.7、1.9和2.4 mm)、冷冻时间(3、5、7、10、20、30秒)和探针与胸膜的距离(5、10和20 mm)进行多次冷冻活检。组织学上肺泡实质面积≥50%被认为是高质量的活检,具有最小的程序伪影。我们使用逻辑回归来确定独立参数作为组织学充分性的危险因素。结果:肺移植全肺切除术后的34只肺共获得545个冷冻切片。使用1.7 mm探针测得的标本平均最大直径(13.5 mm)大于使用1.9和2.4 mm探针测得的标本(11.3和10.7 mm, P=0.07)。结论:冰冻时间和距离胸膜的距离对严重损伤肺ESPD的诊断没有影响组织学质量。较小的冷冻探针尺寸对样品充分性没有负面影响。
Defining Optimal Settings for Lung Cryobiopsy in End-Stage Pulmonary Disease. A Human, Ex Vivo, Diseased Lung Clinical Trial.
Background: To evaluate optimal settings of probe size, freezing time, and distance to the pleura that influence the size and quality of biopsy specimens during transbronchial lung cryobiopsies in ESPD.
Methods: We prospectively recruited 17 patients undergoing lung transplantation. We created a nonperfused ex vivo bronchoscopy setting to perform multiple cryobiopsies with different probe sizes (1.7, 1.9, and 2.4 mm), freezing times (3, 5, 7, 10, 20, 30 seconds), and probe distance from pleura (5, 10, and 20 mm). Alveolated pulmonary parenchyma area≥50% in histology was considered a good quality biopsy, with a minimum procedural artifact. We used logistic regression to identify independent parameters as risk factors for histologic adequacy.
Results: A total of 545 cryobiopsies were obtained from 34 explanted lungs after pneumonectomy for lung transplantation. The mean maximum diameter of the specimen achieved with the 1.7 probe was larger (13.5 mm) than those obtained with 1.9 and 2.4 mm probes (11.3 and 10.7 mm, P= 0.07). More pleural macroscopic damage and pleural tissue in histology occurred with the 2.4 mm probe ( P <0.001). There was no difference in the quality of specimens between the different freezing times and the distance from the pleura.
Conclusions: Freezing time and distance from the pleura did not affect the histologic quality for diagnosing ESPD in severely damaged lungs. Smaller cryoprobe size did not negatively affect sample adequacy.