Luigi Di Biase, Jacopo Marazzato, Fengwei Zou, Aung Lin, Vito Grupposo, Nilarun Chowdhuri, Jennifer Maffre, Salman Farshchi-Heydari, Tushar Sharma, Christopher Beeckler, Assaf Govari, Rahul Bhardwaj, Sanghamitra Mohanty, Andrea Natale, Hiroshi Nakagawa, Xiaodong Zhang
{"title":"使用多模态发生器和接触式力感应消融导管的逐点脉冲场消融:与射频消融在重测慢性猪心脏中的比较。","authors":"Luigi Di Biase, Jacopo Marazzato, Fengwei Zou, Aung Lin, Vito Grupposo, Nilarun Chowdhuri, Jennifer Maffre, Salman Farshchi-Heydari, Tushar Sharma, Christopher Beeckler, Assaf Govari, Rahul Bhardwaj, Sanghamitra Mohanty, Andrea Natale, Hiroshi Nakagawa, Xiaodong Zhang","doi":"10.1161/CIRCEP.123.012344","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Pulsed field ablation (PFA) has emerged as an alternative to radiofrequency ablation. However, data on focal point-by-point PFA are scarce. The aim of this study was to compare lesion durability and collateral damage between focally delivered unipolar/biphasic PFA versus radiofrequency in swine.</p><p><strong>Methods: </strong>Eighteen swine were randomized to low-dose PFA, high-dose PFA, and radiofrequency using a multimodality generator. Radiofrequency delivered by market-available generator served as control group. A contact force-sensing catheter was used to focally deliver PFA/radiofrequency at the pulmonary veins and other predefined sites in the atria. Animals were remapped postprocedurally and 28 days postablation to test lesion durability followed by gross necroscopy and histology.</p><p><strong>Results: </strong>All targeted sites were successfully ablated (contact force value, 13.9±4.1 g). Follow-up remapping showed persistent pulmonary vein isolation in all animals (100%) with lesion durability at nonpulmonary vein sites proven in most (98%). Regardless of the energy source used, the lesion size was similar across the study groups. Transmurality was achieved in 95% of targeted sites and 100% at pulmonary veins. On histology, PFA animals showed more mature scar formation than their radiofrequency counterpart without myocardial necrosis or inflammation. Finally, no sign of collateral damage was observed in any of the groups.</p><p><strong>Conclusions: </strong>In a randomized preclinical study, focally delivered unipolar/biphasic PFA guided by contact force values was associated with durable lesions on chronic remapping and with mature scar formation on histology without signs of collateral injury on necroscopy. Further studies are needed to investigate the long-term feasibility of this new approach to atrial fibrillation treatment.</p>","PeriodicalId":10319,"journal":{"name":"Circulation. Arrhythmia and electrophysiology","volume":null,"pages":null},"PeriodicalIF":9.1000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10734778/pdf/","citationCount":"0","resultStr":"{\"title\":\"Point-by-Point Pulsed Field Ablation Using a Multimodality Generator and a Contact Force-Sensing Ablation Catheter: Comparison With Radiofrequency Ablation in a Remapped Chronic Swine Heart.\",\"authors\":\"Luigi Di Biase, Jacopo Marazzato, Fengwei Zou, Aung Lin, Vito Grupposo, Nilarun Chowdhuri, Jennifer Maffre, Salman Farshchi-Heydari, Tushar Sharma, Christopher Beeckler, Assaf Govari, Rahul Bhardwaj, Sanghamitra Mohanty, Andrea Natale, Hiroshi Nakagawa, Xiaodong Zhang\",\"doi\":\"10.1161/CIRCEP.123.012344\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Pulsed field ablation (PFA) has emerged as an alternative to radiofrequency ablation. However, data on focal point-by-point PFA are scarce. The aim of this study was to compare lesion durability and collateral damage between focally delivered unipolar/biphasic PFA versus radiofrequency in swine.</p><p><strong>Methods: </strong>Eighteen swine were randomized to low-dose PFA, high-dose PFA, and radiofrequency using a multimodality generator. Radiofrequency delivered by market-available generator served as control group. A contact force-sensing catheter was used to focally deliver PFA/radiofrequency at the pulmonary veins and other predefined sites in the atria. Animals were remapped postprocedurally and 28 days postablation to test lesion durability followed by gross necroscopy and histology.</p><p><strong>Results: </strong>All targeted sites were successfully ablated (contact force value, 13.9±4.1 g). Follow-up remapping showed persistent pulmonary vein isolation in all animals (100%) with lesion durability at nonpulmonary vein sites proven in most (98%). Regardless of the energy source used, the lesion size was similar across the study groups. Transmurality was achieved in 95% of targeted sites and 100% at pulmonary veins. On histology, PFA animals showed more mature scar formation than their radiofrequency counterpart without myocardial necrosis or inflammation. Finally, no sign of collateral damage was observed in any of the groups.</p><p><strong>Conclusions: </strong>In a randomized preclinical study, focally delivered unipolar/biphasic PFA guided by contact force values was associated with durable lesions on chronic remapping and with mature scar formation on histology without signs of collateral injury on necroscopy. Further studies are needed to investigate the long-term feasibility of this new approach to atrial fibrillation treatment.</p>\",\"PeriodicalId\":10319,\"journal\":{\"name\":\"Circulation. Arrhythmia and electrophysiology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":9.1000,\"publicationDate\":\"2023-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10734778/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Circulation. Arrhythmia and electrophysiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1161/CIRCEP.123.012344\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/11/23 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation. Arrhythmia and electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1161/CIRCEP.123.012344","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/11/23 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Point-by-Point Pulsed Field Ablation Using a Multimodality Generator and a Contact Force-Sensing Ablation Catheter: Comparison With Radiofrequency Ablation in a Remapped Chronic Swine Heart.
Background: Pulsed field ablation (PFA) has emerged as an alternative to radiofrequency ablation. However, data on focal point-by-point PFA are scarce. The aim of this study was to compare lesion durability and collateral damage between focally delivered unipolar/biphasic PFA versus radiofrequency in swine.
Methods: Eighteen swine were randomized to low-dose PFA, high-dose PFA, and radiofrequency using a multimodality generator. Radiofrequency delivered by market-available generator served as control group. A contact force-sensing catheter was used to focally deliver PFA/radiofrequency at the pulmonary veins and other predefined sites in the atria. Animals were remapped postprocedurally and 28 days postablation to test lesion durability followed by gross necroscopy and histology.
Results: All targeted sites were successfully ablated (contact force value, 13.9±4.1 g). Follow-up remapping showed persistent pulmonary vein isolation in all animals (100%) with lesion durability at nonpulmonary vein sites proven in most (98%). Regardless of the energy source used, the lesion size was similar across the study groups. Transmurality was achieved in 95% of targeted sites and 100% at pulmonary veins. On histology, PFA animals showed more mature scar formation than their radiofrequency counterpart without myocardial necrosis or inflammation. Finally, no sign of collateral damage was observed in any of the groups.
Conclusions: In a randomized preclinical study, focally delivered unipolar/biphasic PFA guided by contact force values was associated with durable lesions on chronic remapping and with mature scar formation on histology without signs of collateral injury on necroscopy. Further studies are needed to investigate the long-term feasibility of this new approach to atrial fibrillation treatment.
期刊介绍:
Circulation: Arrhythmia and Electrophysiology is a journal dedicated to the study and application of clinical cardiac electrophysiology. It covers a wide range of topics including the diagnosis and treatment of cardiac arrhythmias, as well as research in this field. The journal accepts various types of studies, including observational research, clinical trials, epidemiological studies, and advancements in translational research.