{"title":"采用火焰密封或不采用火焰密封核磁共振管:汽液平衡对变温实验精度的影响。","authors":"Derek Morrelli, Santanu Maitra, V. V. Krishnan","doi":"10.1002/mrc.5411","DOIUrl":null,"url":null,"abstract":"<p>In NMR experiments, it is crucial to control the temperature of the sample, especially when measuring kinetic parameters. Usually, it takes 2 to 5 min for the temperature of the sample inside the NMR probe to stabilize at a fixed value set for the experiment. However, the NMR sample tubes are flame-sealed in some cases, such as when working with volatile solvents, atmosphere-sensitive samples, or calibration samples for long-term use. When these samples are placed inside the NMR probe, the spectrometer controls the lower portion (liquid phase) of the NMR sample tube with a gas flow at a fixed temperature, while the upper portion (vapor) is at ambient temperature. This probe design creates a unique temperature gradient across the sample, leading to vapor pressure build-up, particularly inside a sealed NMR tube. By analyzing the temperature-dependent spectral line shape changes of a chemical exchange process, we report that under standard experimental conditions, the sample temperature can take up to 2 to 3 h (instead of minutes) to stabilize. The time scale of the liquid–vapor equilibrium process is much slower, with a half-life exceeding 35 min, in contrast to the 2-min duration required to obtain each spectrum. This phenomenon is exclusively due to the liquid–vapor equilibrium process of the flame-sealed NMR tube and is not observable otherwise.</p>","PeriodicalId":18142,"journal":{"name":"Magnetic Resonance in Chemistry","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2023-11-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"To flame-seal or not to flame-seal NMR tubes: The role of liquid–vapor equilibria on the accuracy of variable temperature experiments\",\"authors\":\"Derek Morrelli, Santanu Maitra, V. V. Krishnan\",\"doi\":\"10.1002/mrc.5411\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p>In NMR experiments, it is crucial to control the temperature of the sample, especially when measuring kinetic parameters. Usually, it takes 2 to 5 min for the temperature of the sample inside the NMR probe to stabilize at a fixed value set for the experiment. However, the NMR sample tubes are flame-sealed in some cases, such as when working with volatile solvents, atmosphere-sensitive samples, or calibration samples for long-term use. When these samples are placed inside the NMR probe, the spectrometer controls the lower portion (liquid phase) of the NMR sample tube with a gas flow at a fixed temperature, while the upper portion (vapor) is at ambient temperature. This probe design creates a unique temperature gradient across the sample, leading to vapor pressure build-up, particularly inside a sealed NMR tube. By analyzing the temperature-dependent spectral line shape changes of a chemical exchange process, we report that under standard experimental conditions, the sample temperature can take up to 2 to 3 h (instead of minutes) to stabilize. The time scale of the liquid–vapor equilibrium process is much slower, with a half-life exceeding 35 min, in contrast to the 2-min duration required to obtain each spectrum. This phenomenon is exclusively due to the liquid–vapor equilibrium process of the flame-sealed NMR tube and is not observable otherwise.</p>\",\"PeriodicalId\":18142,\"journal\":{\"name\":\"Magnetic Resonance in Chemistry\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.9000,\"publicationDate\":\"2023-11-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Magnetic Resonance in Chemistry\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/mrc.5411\",\"RegionNum\":3,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Chemistry","FirstCategoryId":"92","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/mrc.5411","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
To flame-seal or not to flame-seal NMR tubes: The role of liquid–vapor equilibria on the accuracy of variable temperature experiments
In NMR experiments, it is crucial to control the temperature of the sample, especially when measuring kinetic parameters. Usually, it takes 2 to 5 min for the temperature of the sample inside the NMR probe to stabilize at a fixed value set for the experiment. However, the NMR sample tubes are flame-sealed in some cases, such as when working with volatile solvents, atmosphere-sensitive samples, or calibration samples for long-term use. When these samples are placed inside the NMR probe, the spectrometer controls the lower portion (liquid phase) of the NMR sample tube with a gas flow at a fixed temperature, while the upper portion (vapor) is at ambient temperature. This probe design creates a unique temperature gradient across the sample, leading to vapor pressure build-up, particularly inside a sealed NMR tube. By analyzing the temperature-dependent spectral line shape changes of a chemical exchange process, we report that under standard experimental conditions, the sample temperature can take up to 2 to 3 h (instead of minutes) to stabilize. The time scale of the liquid–vapor equilibrium process is much slower, with a half-life exceeding 35 min, in contrast to the 2-min duration required to obtain each spectrum. This phenomenon is exclusively due to the liquid–vapor equilibrium process of the flame-sealed NMR tube and is not observable otherwise.
期刊介绍:
MRC is devoted to the rapid publication of papers which are concerned with the development of magnetic resonance techniques, or in which the application of such techniques plays a pivotal part. Contributions from scientists working in all areas of NMR, ESR and NQR are invited, and papers describing applications in all branches of chemistry, structural biology and materials chemistry are published.
The journal is of particular interest not only to scientists working in academic research, but also those working in commercial organisations who need to keep up-to-date with the latest practical applications of magnetic resonance techniques.