{"title":"电外科设备中的热卡式息肉切除装置的可控电压","authors":"Saidi O. Olalere, JungHun Choi","doi":"10.1155/2023/5521294","DOIUrl":null,"url":null,"abstract":"The study aimed to understand the working procedure of the Olympus PSD-30 Electrosurgical Unit, in which a high-frequency alternating current measures the voltage and power output from the unit when used for a surgical operation to determine the extent of tissue damage. In examining this, power and voltage were analyzed using a stopwatch output; then with an Arduino time-based for 1, 2, and 3 seconds to understand the different modes of the cut and coagulation feedback with an RCC circuit used to mimic the human body. This shows a pattern in which the feedback power increases and the voltage decreases as the cut and coagulation mode increases. The percentage between the stopwatch and Arduino is 29% for the 1 and 2 seconds. With this information, Arduino Uno timing was used to experiment with the device for the different power settings for both the cut and coagulation modes from 2 W to 50 W at 5 W intervals. Based on each trial, the signal was measured for a magnitude of 1 Vpp, and the crest factor obtained was 1.5 with a voltage of 1.088 V and 1.0519 V for both the LabView and oscilloscope, respectively, for the electrosurgical unit of 350 kHz. The power control gives 0.4 W, 2.04 W, and 3.01 W for the power peak at 1, 2, and 3 seconds for the 50 W cut mode of the electrosurgical devices.","PeriodicalId":15716,"journal":{"name":"Journal of Engineering","volume":"53 51","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2023-12-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Controlled Voltage of Hot Snare Polypectomy Device in Electrosurgical Device\",\"authors\":\"Saidi O. Olalere, JungHun Choi\",\"doi\":\"10.1155/2023/5521294\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The study aimed to understand the working procedure of the Olympus PSD-30 Electrosurgical Unit, in which a high-frequency alternating current measures the voltage and power output from the unit when used for a surgical operation to determine the extent of tissue damage. In examining this, power and voltage were analyzed using a stopwatch output; then with an Arduino time-based for 1, 2, and 3 seconds to understand the different modes of the cut and coagulation feedback with an RCC circuit used to mimic the human body. This shows a pattern in which the feedback power increases and the voltage decreases as the cut and coagulation mode increases. The percentage between the stopwatch and Arduino is 29% for the 1 and 2 seconds. With this information, Arduino Uno timing was used to experiment with the device for the different power settings for both the cut and coagulation modes from 2 W to 50 W at 5 W intervals. Based on each trial, the signal was measured for a magnitude of 1 Vpp, and the crest factor obtained was 1.5 with a voltage of 1.088 V and 1.0519 V for both the LabView and oscilloscope, respectively, for the electrosurgical unit of 350 kHz. The power control gives 0.4 W, 2.04 W, and 3.01 W for the power peak at 1, 2, and 3 seconds for the 50 W cut mode of the electrosurgical devices.\",\"PeriodicalId\":15716,\"journal\":{\"name\":\"Journal of Engineering\",\"volume\":\"53 51\",\"pages\":\"\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2023-12-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1155/2023/5521294\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ENGINEERING, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2023/5521294","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, MULTIDISCIPLINARY","Score":null,"Total":0}
Controlled Voltage of Hot Snare Polypectomy Device in Electrosurgical Device
The study aimed to understand the working procedure of the Olympus PSD-30 Electrosurgical Unit, in which a high-frequency alternating current measures the voltage and power output from the unit when used for a surgical operation to determine the extent of tissue damage. In examining this, power and voltage were analyzed using a stopwatch output; then with an Arduino time-based for 1, 2, and 3 seconds to understand the different modes of the cut and coagulation feedback with an RCC circuit used to mimic the human body. This shows a pattern in which the feedback power increases and the voltage decreases as the cut and coagulation mode increases. The percentage between the stopwatch and Arduino is 29% for the 1 and 2 seconds. With this information, Arduino Uno timing was used to experiment with the device for the different power settings for both the cut and coagulation modes from 2 W to 50 W at 5 W intervals. Based on each trial, the signal was measured for a magnitude of 1 Vpp, and the crest factor obtained was 1.5 with a voltage of 1.088 V and 1.0519 V for both the LabView and oscilloscope, respectively, for the electrosurgical unit of 350 kHz. The power control gives 0.4 W, 2.04 W, and 3.01 W for the power peak at 1, 2, and 3 seconds for the 50 W cut mode of the electrosurgical devices.
期刊介绍:
Journal of Engineering is a peer-reviewed, Open Access journal that publishes original research articles as well as review articles in several areas of engineering. The subject areas covered by the journal are: - Chemical Engineering - Civil Engineering - Computer Engineering - Electrical Engineering - Industrial Engineering - Mechanical Engineering