输液泵的流速偏差:输液装置缺陷导致输液泵功能失常和严重的精度偏差。

IF 2.9 Q1 NURSING Journal of Infusion Nursing Pub Date : 2024-01-01 DOI:10.1097/NAN.0000000000000530
Axel Hult, Irina Zholobova, Emelie Bäcklin, Per Nydert
{"title":"输液泵的流速偏差:输液装置缺陷导致输液泵功能失常和严重的精度偏差。","authors":"Axel Hult, Irina Zholobova, Emelie Bäcklin, Per Nydert","doi":"10.1097/NAN.0000000000000530","DOIUrl":null,"url":null,"abstract":"<p><p>Volumetric infusion pumps are used together with infusion sets to deliver medication to patients. Flow rate errors leading to overinfusion or underinfusion are known problems with these devices. Recently, numerous underinfusion flow rate errors were reported at a Swedish hospital. This experimental study reports on the investigation of these errors and specifically investigates the effect of operating the pump with a defective infusion set that has a visible elongation of the silicone segment of the set. Pump flow rate accuracy testing was performed using a gravimetric method. Experiments included a manipulated infusion set and a defective infusion set used in clinic. The use of a defective infusion set resulted in considerable accuracy deviations. The pump reported an infused amount greater than what was infused and did not provide any alarm or information indicating a reduced output. Using an elongated infusion set, the pump can be brought into an erroneous operating state where the infused amount delivered by the pump is considerably less than what has been programmed and what is shown on the pump display. This could put the patient at risk of not receiving the intended medication within the appropriate time.</p>","PeriodicalId":46291,"journal":{"name":"Journal of Infusion Nursing","volume":null,"pages":null},"PeriodicalIF":2.9000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Flow Rate Deviation in Infusion Pump: Infusion Set Defect Enables Pump Malfunction and Considerable Accuracy Deviation.\",\"authors\":\"Axel Hult, Irina Zholobova, Emelie Bäcklin, Per Nydert\",\"doi\":\"10.1097/NAN.0000000000000530\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Volumetric infusion pumps are used together with infusion sets to deliver medication to patients. Flow rate errors leading to overinfusion or underinfusion are known problems with these devices. Recently, numerous underinfusion flow rate errors were reported at a Swedish hospital. This experimental study reports on the investigation of these errors and specifically investigates the effect of operating the pump with a defective infusion set that has a visible elongation of the silicone segment of the set. Pump flow rate accuracy testing was performed using a gravimetric method. Experiments included a manipulated infusion set and a defective infusion set used in clinic. The use of a defective infusion set resulted in considerable accuracy deviations. The pump reported an infused amount greater than what was infused and did not provide any alarm or information indicating a reduced output. Using an elongated infusion set, the pump can be brought into an erroneous operating state where the infused amount delivered by the pump is considerably less than what has been programmed and what is shown on the pump display. This could put the patient at risk of not receiving the intended medication within the appropriate time.</p>\",\"PeriodicalId\":46291,\"journal\":{\"name\":\"Journal of Infusion Nursing\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Infusion Nursing\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1097/NAN.0000000000000530\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"NURSING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Infusion Nursing","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/NAN.0000000000000530","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"NURSING","Score":null,"Total":0}
引用次数: 0

摘要

容积式输液泵与输液器一起用于向患者输注药物。流量误差导致过度输液或输液不足是这些设备的已知问题。最近,一家瑞典医院报告了多起输液不足的流量误差。本实验研究报告对这些误差进行了调查,并特别研究了使用有缺陷的输液器操作泵的影响,该输液器的硅胶部分有明显的伸长。泵流量精度测试采用重力法进行。实验包括操作输液器和临床使用的有缺陷输液器。使用有缺陷的输液器导致了相当大的精度偏差。输液泵报告的输液量大于实际输液量,而且没有发出任何警报或提供任何表明输液量减少的信息。使用加长输液器可使输液泵进入错误的工作状态,即输液泵输出的输液量大大少于程序设定的输液量和输液泵显示屏上显示的输液量。这可能会导致患者无法在适当的时间内获得预期的药物。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Flow Rate Deviation in Infusion Pump: Infusion Set Defect Enables Pump Malfunction and Considerable Accuracy Deviation.

Volumetric infusion pumps are used together with infusion sets to deliver medication to patients. Flow rate errors leading to overinfusion or underinfusion are known problems with these devices. Recently, numerous underinfusion flow rate errors were reported at a Swedish hospital. This experimental study reports on the investigation of these errors and specifically investigates the effect of operating the pump with a defective infusion set that has a visible elongation of the silicone segment of the set. Pump flow rate accuracy testing was performed using a gravimetric method. Experiments included a manipulated infusion set and a defective infusion set used in clinic. The use of a defective infusion set resulted in considerable accuracy deviations. The pump reported an infused amount greater than what was infused and did not provide any alarm or information indicating a reduced output. Using an elongated infusion set, the pump can be brought into an erroneous operating state where the infused amount delivered by the pump is considerably less than what has been programmed and what is shown on the pump display. This could put the patient at risk of not receiving the intended medication within the appropriate time.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.00
自引率
15.00%
发文量
52
期刊介绍: Journal of Infusion Nursing, the official publication of the Infusion Nurses Society (INS), seeks to promote excellence in infusion nursing by presenting new research, clinical reviews, case studies, and professional development information relevant to the practice of infusion therapy. Articles selected for publication represent the broad scope of the infusion specialty and draw on the expertise of all healthcare providers who participate in the delivery of infusion therapy.
期刊最新文献
A Comparison of Pumping Infusion-Induced Contamination With Different Syringe Types. An Exemplary Publication. Complications Related to the Securement Device in Peripheral Intravenous Catheters: A Randomized Study. Midline Catheter-Associated Thrombosis (MCAT): Does Tip Location in the Axillary Vein Increase Risk? Parenteral Nutrition Safety.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1