{"title":"输液泵的流速偏差:输液装置缺陷导致输液泵功能失常和严重的精度偏差。","authors":"Axel Hult, Irina Zholobova, Emelie Bäcklin, Per Nydert","doi":"10.1097/NAN.0000000000000530","DOIUrl":null,"url":null,"abstract":"<p><p>Volumetric infusion pumps are used together with infusion sets to deliver medication to patients. Flow rate errors leading to overinfusion or underinfusion are known problems with these devices. Recently, numerous underinfusion flow rate errors were reported at a Swedish hospital. This experimental study reports on the investigation of these errors and specifically investigates the effect of operating the pump with a defective infusion set that has a visible elongation of the silicone segment of the set. Pump flow rate accuracy testing was performed using a gravimetric method. Experiments included a manipulated infusion set and a defective infusion set used in clinic. The use of a defective infusion set resulted in considerable accuracy deviations. The pump reported an infused amount greater than what was infused and did not provide any alarm or information indicating a reduced output. Using an elongated infusion set, the pump can be brought into an erroneous operating state where the infused amount delivered by the pump is considerably less than what has been programmed and what is shown on the pump display. This could put the patient at risk of not receiving the intended medication within the appropriate time.</p>","PeriodicalId":46291,"journal":{"name":"Journal of Infusion Nursing","volume":null,"pages":null},"PeriodicalIF":2.9000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Flow Rate Deviation in Infusion Pump: Infusion Set Defect Enables Pump Malfunction and Considerable Accuracy Deviation.\",\"authors\":\"Axel Hult, Irina Zholobova, Emelie Bäcklin, Per Nydert\",\"doi\":\"10.1097/NAN.0000000000000530\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Volumetric infusion pumps are used together with infusion sets to deliver medication to patients. Flow rate errors leading to overinfusion or underinfusion are known problems with these devices. Recently, numerous underinfusion flow rate errors were reported at a Swedish hospital. This experimental study reports on the investigation of these errors and specifically investigates the effect of operating the pump with a defective infusion set that has a visible elongation of the silicone segment of the set. Pump flow rate accuracy testing was performed using a gravimetric method. Experiments included a manipulated infusion set and a defective infusion set used in clinic. The use of a defective infusion set resulted in considerable accuracy deviations. The pump reported an infused amount greater than what was infused and did not provide any alarm or information indicating a reduced output. Using an elongated infusion set, the pump can be brought into an erroneous operating state where the infused amount delivered by the pump is considerably less than what has been programmed and what is shown on the pump display. This could put the patient at risk of not receiving the intended medication within the appropriate time.</p>\",\"PeriodicalId\":46291,\"journal\":{\"name\":\"Journal of Infusion Nursing\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Infusion Nursing\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1097/NAN.0000000000000530\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"NURSING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Infusion Nursing","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/NAN.0000000000000530","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"NURSING","Score":null,"Total":0}
Flow Rate Deviation in Infusion Pump: Infusion Set Defect Enables Pump Malfunction and Considerable Accuracy Deviation.
Volumetric infusion pumps are used together with infusion sets to deliver medication to patients. Flow rate errors leading to overinfusion or underinfusion are known problems with these devices. Recently, numerous underinfusion flow rate errors were reported at a Swedish hospital. This experimental study reports on the investigation of these errors and specifically investigates the effect of operating the pump with a defective infusion set that has a visible elongation of the silicone segment of the set. Pump flow rate accuracy testing was performed using a gravimetric method. Experiments included a manipulated infusion set and a defective infusion set used in clinic. The use of a defective infusion set resulted in considerable accuracy deviations. The pump reported an infused amount greater than what was infused and did not provide any alarm or information indicating a reduced output. Using an elongated infusion set, the pump can be brought into an erroneous operating state where the infused amount delivered by the pump is considerably less than what has been programmed and what is shown on the pump display. This could put the patient at risk of not receiving the intended medication within the appropriate time.
期刊介绍:
Journal of Infusion Nursing, the official publication of the Infusion Nurses Society (INS), seeks to promote excellence in infusion nursing by presenting new research, clinical reviews, case studies, and professional development information relevant to the practice of infusion therapy. Articles selected for publication represent the broad scope of the infusion specialty and draw on the expertise of all healthcare providers who participate in the delivery of infusion therapy.