单层和三层正畸矫治器混合物在热循环下的稳定性差异:对临床耐久性的影响。

IF 1.4 Q3 DENTISTRY, ORAL SURGERY & MEDICINE Acta Stomatologica Croatica Pub Date : 2023-12-01 DOI:10.15644/asc57/4/1
Luka Šimunović, Antonija Jurela, Karlo Sudarević, Ivana Bačić, Senka Meštrović
{"title":"单层和三层正畸矫治器混合物在热循环下的稳定性差异:对临床耐久性的影响。","authors":"Luka Šimunović, Antonija Jurela, Karlo Sudarević, Ivana Bačić, Senka Meštrović","doi":"10.15644/asc57/4/1","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To optimize the therapeutic usefulness of aligners, it is crucial to understand how their mechanical properties alter with time.</p><p><strong>Materials and methods: </strong>Specimens from four different brands, including Duran+, CA® Pro, Zendura A, and Zendura FLX, were produced for material testing of thermoplastic orthodontic aligners (TOA) using dimensions measuring 4mm x 10mm. Each brand's 24 samples were split into three groups as follows: G1 being thermoformed, G2 being thermoformed and underwent 500 thermocycles (simulating 7 days), and G3 being thermoformed and underwent 1000 thermocycles (simulating 14 days). Surface roughness, modulus of elasticity in bending, and spectrophotometry were used to assess the effect of aging on TOAs.</p><p><strong>Results: </strong>After 1000 thermocycles, Duran+ had the highest modulus of elasticity and differed statistically from all other groups. The intragroup comparison showed that only Duran+'s elastic modulus significantly changed after 1000 thermocycles in comparison with the control group. Surface roughness values (Ra), did not statistically differ among brands or thermocycling group measures. The change in chemical properties was not significant in any brand.</p><p><strong>Conclusion: </strong>One-layer PETG (Duran+) failed to demonstrate stability after <i>in vitro</i> aging, thus suggesting that clinicians should be aware of the change in mechanical properties when using one-layer PETG (Duran +) in a 2 weeks regime.</p>","PeriodicalId":7154,"journal":{"name":"Acta Stomatologica Croatica","volume":"57 4","pages":"286-299"},"PeriodicalIF":1.4000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10812913/pdf/","citationCount":"0","resultStr":"{\"title\":\"Differential Stability of One-layer and Three-layer Orthodontic Aligner Blends under Thermocycling: Implications for Clinical Durability.\",\"authors\":\"Luka Šimunović, Antonija Jurela, Karlo Sudarević, Ivana Bačić, Senka Meštrović\",\"doi\":\"10.15644/asc57/4/1\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>To optimize the therapeutic usefulness of aligners, it is crucial to understand how their mechanical properties alter with time.</p><p><strong>Materials and methods: </strong>Specimens from four different brands, including Duran+, CA® Pro, Zendura A, and Zendura FLX, were produced for material testing of thermoplastic orthodontic aligners (TOA) using dimensions measuring 4mm x 10mm. Each brand's 24 samples were split into three groups as follows: G1 being thermoformed, G2 being thermoformed and underwent 500 thermocycles (simulating 7 days), and G3 being thermoformed and underwent 1000 thermocycles (simulating 14 days). Surface roughness, modulus of elasticity in bending, and spectrophotometry were used to assess the effect of aging on TOAs.</p><p><strong>Results: </strong>After 1000 thermocycles, Duran+ had the highest modulus of elasticity and differed statistically from all other groups. The intragroup comparison showed that only Duran+'s elastic modulus significantly changed after 1000 thermocycles in comparison with the control group. Surface roughness values (Ra), did not statistically differ among brands or thermocycling group measures. The change in chemical properties was not significant in any brand.</p><p><strong>Conclusion: </strong>One-layer PETG (Duran+) failed to demonstrate stability after <i>in vitro</i> aging, thus suggesting that clinicians should be aware of the change in mechanical properties when using one-layer PETG (Duran +) in a 2 weeks regime.</p>\",\"PeriodicalId\":7154,\"journal\":{\"name\":\"Acta Stomatologica Croatica\",\"volume\":\"57 4\",\"pages\":\"286-299\"},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2023-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10812913/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta Stomatologica Croatica\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.15644/asc57/4/1\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Stomatologica Croatica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15644/asc57/4/1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

目的:为了优化矫治器的治疗作用,了解其机械性能如何随时间变化至关重要:制作了四个不同品牌的样本,包括 Duran+、CA® Pro、Zendura A 和 Zendura FLX,用于热塑性正畸矫治器(TOA)的材料测试,尺寸为 4 毫米 x 10 毫米。每个品牌的 24 个样品被分成以下三组:G1 为热成型,G2 为热成型并经过 500 次热循环(模拟 7 天),G3 为热成型并经过 1000 次热循环(模拟 14 天)。采用表面粗糙度、弯曲弹性模量和分光光度法评估老化对 TOAs 的影响:结果:经过 1000 次热循环后,Duran+ 的弹性模量最高,与所有其他组别都有统计学差异。组内比较显示,与对照组相比,只有 Duran+ 的弹性模量在 1000 次热循环后发生了显著变化。各品牌或热循环组之间的表面粗糙度值(Ra)没有统计学差异。结论:单层 PETG(Duran+)在体外老化后未能表现出稳定性,因此建议临床医生在两周内使用单层 PETG(Duran+)时应注意机械性能的变化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Differential Stability of One-layer and Three-layer Orthodontic Aligner Blends under Thermocycling: Implications for Clinical Durability.

Objectives: To optimize the therapeutic usefulness of aligners, it is crucial to understand how their mechanical properties alter with time.

Materials and methods: Specimens from four different brands, including Duran+, CA® Pro, Zendura A, and Zendura FLX, were produced for material testing of thermoplastic orthodontic aligners (TOA) using dimensions measuring 4mm x 10mm. Each brand's 24 samples were split into three groups as follows: G1 being thermoformed, G2 being thermoformed and underwent 500 thermocycles (simulating 7 days), and G3 being thermoformed and underwent 1000 thermocycles (simulating 14 days). Surface roughness, modulus of elasticity in bending, and spectrophotometry were used to assess the effect of aging on TOAs.

Results: After 1000 thermocycles, Duran+ had the highest modulus of elasticity and differed statistically from all other groups. The intragroup comparison showed that only Duran+'s elastic modulus significantly changed after 1000 thermocycles in comparison with the control group. Surface roughness values (Ra), did not statistically differ among brands or thermocycling group measures. The change in chemical properties was not significant in any brand.

Conclusion: One-layer PETG (Duran+) failed to demonstrate stability after in vitro aging, thus suggesting that clinicians should be aware of the change in mechanical properties when using one-layer PETG (Duran +) in a 2 weeks regime.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Acta Stomatologica Croatica
Acta Stomatologica Croatica DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
2.40
自引率
28.60%
发文量
32
审稿时长
12 weeks
期刊介绍: The Acta Stomatologica Croatica (ASCRO) is a leading scientific non-profit journal in the field of dental, oral and cranio-facial sciences during the past 44 years in Croatia. ASCRO publishes original scientific and clinical papers, preliminary communications, case reports, book reviews, letters to the editor and news. Review articles are published by invitation from the Editor-in-Chief by acclaimed professionals in distinct fields of dental medicine. All manuscripts are subjected to peer review process.
期刊最新文献
Age Determination Based on Open Apex Measurement in the Developing Dentition: Comparing the Accuracy of the Belgrade Age Formula (BAF) with the European Formula on a Bosnian Children Population. Differences between Doctors of Dental Medicine and Doctors of Medicine Awareness of Their Online Image and Perception Concerns: a Quantitative Cross-Sectional Study. Evaluation of the Bond Strength between the Acrylic Teeth and Reinforced or Non-reinforced Complete Denture Base. Exploring the Root Canal Morphology of Third Molars in the Bosnia and Herzegovina Population. Legal Dental Age Thresholds Estimation in Croatian Children and Adolescents: Accuracy of the Olze, Haavikko, and Demirjian Methods.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1