评估放射治疗软件中的下垂校正校准误差对图像匹配的影响。

IF 2.4 4区 医学 Q3 ENGINEERING, BIOMEDICAL Physical and Engineering Sciences in Medicine Pub Date : 2024-06-01 Epub Date: 2024-02-19 DOI:10.1007/s13246-024-01388-y
Yumi Yamazawa, Akitane Osaka, Yasushi Fujii, Takahiro Nakayama, Kunio Nishioka, Yoshinori Tanabe
{"title":"评估放射治疗软件中的下垂校正校准误差对图像匹配的影响。","authors":"Yumi Yamazawa, Akitane Osaka, Yasushi Fujii, Takahiro Nakayama, Kunio Nishioka, Yoshinori Tanabe","doi":"10.1007/s13246-024-01388-y","DOIUrl":null,"url":null,"abstract":"<p><p>To investigate the impact of sagging correction calibration errors in radiotherapy software on image matching. Three software applications were used, with and without a polymethyl methacrylate rod supporting the ball bearings (BB). The calibration error for sagging correction across nine flex maps (FMs) was determined by shifting the BB positions along the Left-Right (LR), Gun-Target (GT), and Up-Down (UD) directions from the reference point. Lucy and pelvic phantom cone-beam computed tomography (CBCT) images underwent auto-matching after modifying each FM. Image deformation was assessed in orthogonal CBCT planes, and the correlations among BB shift magnitude, deformation vector value, and differences in auto-matching were analyzed. The average difference in analysis results among the three softwares for the Winston-Lutz test was within 0.1 mm. The determination coefficients (R<sup>2</sup>) between the BB shift amount and Lucy phantom matching error in each FM were 0.99, 0.99, and 1.00 in the LR-, GT-, and UD-directions, respectively. The pelvis phantom demonstrated no cross-correlation in the GT direction during auto-matching error evaluation using each FM. The correlation coefficient (r) between the BB shift and the deformation vector value was 0.95 on average for all image planes. Slight differences were observed among software in the evaluation of the Winston-Lutz test. The sagging correction calibration error in the radiotherapy imaging system was caused by an auto-matching error of the phantom and deformation of CBCT images.</p>","PeriodicalId":48490,"journal":{"name":"Physical and Engineering Sciences in Medicine","volume":null,"pages":null},"PeriodicalIF":2.4000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11166816/pdf/","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the effect of sagging correction calibration errors in radiotherapy software on image matching.\",\"authors\":\"Yumi Yamazawa, Akitane Osaka, Yasushi Fujii, Takahiro Nakayama, Kunio Nishioka, Yoshinori Tanabe\",\"doi\":\"10.1007/s13246-024-01388-y\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>To investigate the impact of sagging correction calibration errors in radiotherapy software on image matching. Three software applications were used, with and without a polymethyl methacrylate rod supporting the ball bearings (BB). The calibration error for sagging correction across nine flex maps (FMs) was determined by shifting the BB positions along the Left-Right (LR), Gun-Target (GT), and Up-Down (UD) directions from the reference point. Lucy and pelvic phantom cone-beam computed tomography (CBCT) images underwent auto-matching after modifying each FM. Image deformation was assessed in orthogonal CBCT planes, and the correlations among BB shift magnitude, deformation vector value, and differences in auto-matching were analyzed. The average difference in analysis results among the three softwares for the Winston-Lutz test was within 0.1 mm. The determination coefficients (R<sup>2</sup>) between the BB shift amount and Lucy phantom matching error in each FM were 0.99, 0.99, and 1.00 in the LR-, GT-, and UD-directions, respectively. The pelvis phantom demonstrated no cross-correlation in the GT direction during auto-matching error evaluation using each FM. The correlation coefficient (r) between the BB shift and the deformation vector value was 0.95 on average for all image planes. Slight differences were observed among software in the evaluation of the Winston-Lutz test. The sagging correction calibration error in the radiotherapy imaging system was caused by an auto-matching error of the phantom and deformation of CBCT images.</p>\",\"PeriodicalId\":48490,\"journal\":{\"name\":\"Physical and Engineering Sciences in Medicine\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11166816/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physical and Engineering Sciences in Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s13246-024-01388-y\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/19 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physical and Engineering Sciences in Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s13246-024-01388-y","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/19 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

研究放射治疗软件中的下垂校正校准误差对图像匹配的影响。使用了三种软件应用程序,分别使用和不使用支撑球轴承(BB)的聚甲基丙烯酸甲酯杆。通过沿参考点的左-右(LR)、枪-靶(GT)和上-下(UD)方向移动球轴承位置,确定了九张柔性图(FM)的下垂校正校准误差。修改每个调频后,对露西和骨盆模型锥束计算机断层扫描(CBCT)图像进行自动匹配。在正交 CBCT 平面上评估图像变形,并分析 BB 偏移幅度、变形矢量值和自动匹配差异之间的相关性。在 Winston-Lutz 测试中,三种软件分析结果的平均差异在 0.1 毫米以内。在 LR、GT 和 UD 方向上,每个调频中 BB 偏移量与露西模型匹配误差之间的判定系数 (R2) 分别为 0.99、0.99 和 1.00。在使用各调频进行自动匹配误差评估时,骨盆模型在 GT 方向上没有显示出交叉相关性。在所有图像平面上,BB 移位和变形矢量值之间的相关系数(r)平均为 0.95。在 Winston-Lutz 测试的评估中,不同软件之间存在细微差别。放射治疗成像系统的下垂校正校准误差是由模型的自动匹配误差和 CBCT 图像的变形造成的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Evaluation of the effect of sagging correction calibration errors in radiotherapy software on image matching.

To investigate the impact of sagging correction calibration errors in radiotherapy software on image matching. Three software applications were used, with and without a polymethyl methacrylate rod supporting the ball bearings (BB). The calibration error for sagging correction across nine flex maps (FMs) was determined by shifting the BB positions along the Left-Right (LR), Gun-Target (GT), and Up-Down (UD) directions from the reference point. Lucy and pelvic phantom cone-beam computed tomography (CBCT) images underwent auto-matching after modifying each FM. Image deformation was assessed in orthogonal CBCT planes, and the correlations among BB shift magnitude, deformation vector value, and differences in auto-matching were analyzed. The average difference in analysis results among the three softwares for the Winston-Lutz test was within 0.1 mm. The determination coefficients (R2) between the BB shift amount and Lucy phantom matching error in each FM were 0.99, 0.99, and 1.00 in the LR-, GT-, and UD-directions, respectively. The pelvis phantom demonstrated no cross-correlation in the GT direction during auto-matching error evaluation using each FM. The correlation coefficient (r) between the BB shift and the deformation vector value was 0.95 on average for all image planes. Slight differences were observed among software in the evaluation of the Winston-Lutz test. The sagging correction calibration error in the radiotherapy imaging system was caused by an auto-matching error of the phantom and deformation of CBCT images.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
8.40
自引率
4.50%
发文量
110
期刊最新文献
PET/CT-based 3D multi-class semantic segmentation of ovarian cancer and the stability of the extracted radiomics features. PPG2RespNet: a deep learning model for respirational signal synthesis and monitoring from photoplethysmography (PPG) signal Ecg signal watermarking using QR decomposition Effect of mirror system and scanner bed of a flatbed scanner on lateral response artefact in radiochromic film dosimetry A deep learning phase-based solution in 2D echocardiography motion estimation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1