{"title":"交联前和交联后猪角膜薄片的紫外线-A 吸收分析。","authors":"Renan Rodrigues, Paulo Schor, Bernardo Kaplan Moscovici, Priscila Guidini, Mario Antonio Stefani, Patrícia Alessandra Bersanetti","doi":"10.5935/0004-2749.2022-0273","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To determine the absorbance coefficient of the thin porcine cornea to ultraviolet-A radiation (365 nm) submitted for crosslinking.</p><p><strong>Methods: </strong>This in vitro, benchtop experiment using cadaver tissue study analyzed 12 porcine corneal lamellas, which were obtained using a microkeratome after mechanical de-epithelization and separated into three thickness groups: 180, 300, and 360 μm. The corneal thickness values were measured by anterior-segment optical coherence tomography. All lamellas had ultraviolet-A (365 nm) absorbance measured with a 96-well plate spectrophotometer using an ultraviolet transparent microplate before riboflavin instillation and preand post-crosslinking according to the Dresden protocol.</p><p><strong>Results: </strong>The ultraviolet absorbance profiles of the 180, 300, and 360 μm groups were obtained as α-coefficients of 12.85, 76.55, and 120.27, respectively. A theoretical formula was calculated though a statistical analysis that demonstrated the correlation between stromal lamellar thickness and ultraviolet absorbance.</p><p><strong>Conclusions: </strong>Corneal thickness and ultraviolet-A spectral absorbance of corneal lamellas showed linear correlation. These findings can potentially contribute to the optimization of ultraviolet-A application during crosslinking, making the treatment of corneas with thickness <400 μm safe and personalized energy delivery for each corneal thickness.</p>","PeriodicalId":8397,"journal":{"name":"Arquivos brasileiros de oftalmologia","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2024-03-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Ultraviolet-A absorbance analysis in thin porcine corneas pre-and post-crosslinking.\",\"authors\":\"Renan Rodrigues, Paulo Schor, Bernardo Kaplan Moscovici, Priscila Guidini, Mario Antonio Stefani, Patrícia Alessandra Bersanetti\",\"doi\":\"10.5935/0004-2749.2022-0273\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To determine the absorbance coefficient of the thin porcine cornea to ultraviolet-A radiation (365 nm) submitted for crosslinking.</p><p><strong>Methods: </strong>This in vitro, benchtop experiment using cadaver tissue study analyzed 12 porcine corneal lamellas, which were obtained using a microkeratome after mechanical de-epithelization and separated into three thickness groups: 180, 300, and 360 μm. The corneal thickness values were measured by anterior-segment optical coherence tomography. All lamellas had ultraviolet-A (365 nm) absorbance measured with a 96-well plate spectrophotometer using an ultraviolet transparent microplate before riboflavin instillation and preand post-crosslinking according to the Dresden protocol.</p><p><strong>Results: </strong>The ultraviolet absorbance profiles of the 180, 300, and 360 μm groups were obtained as α-coefficients of 12.85, 76.55, and 120.27, respectively. A theoretical formula was calculated though a statistical analysis that demonstrated the correlation between stromal lamellar thickness and ultraviolet absorbance.</p><p><strong>Conclusions: </strong>Corneal thickness and ultraviolet-A spectral absorbance of corneal lamellas showed linear correlation. These findings can potentially contribute to the optimization of ultraviolet-A application during crosslinking, making the treatment of corneas with thickness <400 μm safe and personalized energy delivery for each corneal thickness.</p>\",\"PeriodicalId\":8397,\"journal\":{\"name\":\"Arquivos brasileiros de oftalmologia\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-03-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Arquivos brasileiros de oftalmologia\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5935/0004-2749.2022-0273\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arquivos brasileiros de oftalmologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5935/0004-2749.2022-0273","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:测定用于交联的猪角膜薄片对紫外线-A 辐射(365 纳米)的吸收系数:这项体外台式实验使用尸体组织研究分析了 12 个猪角膜板层,这些板层是在机械去上皮后使用显微角膜刀获得的,并分为三个厚度组:180、300 和 360 μm。角膜厚度值是通过前段光学相干断层扫描测量的。根据德累斯顿方案,在核黄素灌注前和交联前后,所有板层均使用 96 孔板分光光度计和紫外透明微孔板测量紫外线-A(365 纳米)吸光度:结果:180、300 和 360 μm 组的紫外线吸光度曲线的 α 系数分别为 12.85、76.55 和 120.27。通过统计分析计算出的理论公式证明了基质板层厚度与紫外线吸收率之间的相关性:结论:角膜厚度与角膜板层的紫外线-A 光谱吸收率呈线性相关。这些发现可能有助于在交联过程中优化紫外线-A 的应用,使角膜厚度与紫外线-A 吸收率呈线性相关。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Ultraviolet-A absorbance analysis in thin porcine corneas pre-and post-crosslinking.

Purpose: To determine the absorbance coefficient of the thin porcine cornea to ultraviolet-A radiation (365 nm) submitted for crosslinking.

Methods: This in vitro, benchtop experiment using cadaver tissue study analyzed 12 porcine corneal lamellas, which were obtained using a microkeratome after mechanical de-epithelization and separated into three thickness groups: 180, 300, and 360 μm. The corneal thickness values were measured by anterior-segment optical coherence tomography. All lamellas had ultraviolet-A (365 nm) absorbance measured with a 96-well plate spectrophotometer using an ultraviolet transparent microplate before riboflavin instillation and preand post-crosslinking according to the Dresden protocol.

Results: The ultraviolet absorbance profiles of the 180, 300, and 360 μm groups were obtained as α-coefficients of 12.85, 76.55, and 120.27, respectively. A theoretical formula was calculated though a statistical analysis that demonstrated the correlation between stromal lamellar thickness and ultraviolet absorbance.

Conclusions: Corneal thickness and ultraviolet-A spectral absorbance of corneal lamellas showed linear correlation. These findings can potentially contribute to the optimization of ultraviolet-A application during crosslinking, making the treatment of corneas with thickness <400 μm safe and personalized energy delivery for each corneal thickness.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.60
自引率
0.00%
发文量
200
审稿时长
6-12 weeks
期刊介绍: The ABO-ARQUIVOS BRASILEIROS DE OFTALMOLOGIA (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - electronic version), the official bimonthly publication of the Brazilian Council of Ophthalmology (CBO), aims to disseminate scientific studies in Ophthalmology, Visual Science and Health public, by promoting research, improvement and updating of professionals related to the field.
期刊最新文献
Early visual intervention, visual function analysis, and grating visual acuity outcomes in children with congenital Zika syndrome. Prevalence of myopic eyes in private practice and public care services in Brazil: 11 years of retrospective analysis. A comparative analysis of the effectiveness of laser correction with a mechanical and laser microkeratome for myopia. A novel predictor of persistent ocular hypotony after pars plana vitrectomy for rhegmatogenous retinal detachment: The initial intraocular pressure difference between the eye with RRD and the fellow eye. Comparison of the refractive prediction errors of artificial intelligence formula with 3 conventional formulas and 11 combination methods in cataract surgery on eyes with short axial length.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1