{"title":"用超声弹性成像技术量化三维打印水凝胶模型中不同主动脉内移植物的平均压力归一化应变降低百分比","authors":"Dakota W. Gonring BA , Zachary R. Zottola BS , Adnan A. Hirad MD, PhD , Ronald Lakony BS , Michael S. Richards PhD , Grayson Pitcher MD , Michael C. Stoner MD , Doran S. Mix MD","doi":"10.1016/j.jvssci.2024.100198","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>Strain has become a viable index for evaluating abdominal aortic aneurysm stability after endovascular aneurysm repair (EVAR). In addition, literature has shown that healthy aortic tissue requires a degree of strain to maintain homeostasis. This has led to the hypothesis that too much strain reduction conferred by a high degree of graft oversizing is detrimental to the aneurysm neck in the seal zone of abdominal aortic aneurysms after EVAR. We investigated this in a laboratory experiment by examining the effects that graft oversizing has on the pressure-normalized strain (<span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/pulse pressure [PP]) reduction using four different infrarenal EVAR endografts and our ultrasound elastography technique. Approximate graft oversizing percentages were 20% (30 mm phantom-graft combinations), 30% (28 mm phantom-graft combinations), and 50% (24 mm phantom-graft combinations).</p></div><div><h3>Methods</h3><p>Axisymmetric, 10% by mass polyvinyl alcohol phantoms were connected to a flow simulator. Ultrasound elastography was performed before and after implantation with the four different endografts: (1) 36 mm polyester/stainless steel, (2) 36 mm polyester/electropolished nitinol, (3) 35 mm polytetrafluoroethylene (PTFE)/nitinol, and (4) 36 mm nitinol/polyester/platinum-iridium. Five ultrasound cine loops were taken of each phantom-graft combination. They were analyzed over two different cardiac cycles (end-diastole to end-diastole), yielding a total of 10 maximum mean principal strain (<span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>) values. <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span> was divided by pulse pressure to yield pressure-normalized strain (<span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP). An analysis of variance was performed for graft comparisons. We calculated the average percent <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP reduction by manufacturer and percent oversizing. These values were used for linear regression analysis.</p></div><div><h3>Results</h3><p>Results from one-way analysis of variance showed a significant difference in <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP between the empty phantom condition and all oversizing conditions for all graft manufacturers (<em>F</em>(3, 56) = 106.7 [graft A], 132.7 [graft B], 106.5 [graft C], 105.7 [graft D], <em>P</em> < .0001 for grafts A-D). There was a significant difference when comparing the 50% condition with the 30% and 20% conditions across all manufacturers by post hoc analysis (<em>P</em> < .0001). No significant difference was found when comparing the 20% and 30% oversizing conditions for any of the manufacturers or when comparing <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP values across the manufacturers according to percent oversize. Linear regression demonstrated a significant positive correlation between the percent graft oversize and the all-graft average percent <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP reduction (<span><math><mrow><msup><mi>R</mi><mn>2</mn></msup></mrow></math></span> = 0.84, <em>P</em> < .0001).</p></div><div><h3>Conclusions</h3><p>This brief report suggests that a 10% increase in graft oversizing leads to an approximate 5.9% reduction in <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP on average. Applied clinically, this increase may result in increased stiffness in axisymmetric vessels after EVAR. Further research is needed to determine if this is clinically significant.</p></div>","PeriodicalId":74035,"journal":{"name":"JVS-vascular science","volume":"5 ","pages":"Article 100198"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666350324000099/pdfft?md5=174462f0b3289da4a0f6e8331d8e0f44&pid=1-s2.0-S2666350324000099-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Ultrasound elastography to quantify average percent pressure-normalized strain reduction associated with different aortic endografts in 3D-printed hydrogel phantoms\",\"authors\":\"Dakota W. Gonring BA , Zachary R. Zottola BS , Adnan A. Hirad MD, PhD , Ronald Lakony BS , Michael S. Richards PhD , Grayson Pitcher MD , Michael C. Stoner MD , Doran S. Mix MD\",\"doi\":\"10.1016/j.jvssci.2024.100198\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><p>Strain has become a viable index for evaluating abdominal aortic aneurysm stability after endovascular aneurysm repair (EVAR). In addition, literature has shown that healthy aortic tissue requires a degree of strain to maintain homeostasis. This has led to the hypothesis that too much strain reduction conferred by a high degree of graft oversizing is detrimental to the aneurysm neck in the seal zone of abdominal aortic aneurysms after EVAR. We investigated this in a laboratory experiment by examining the effects that graft oversizing has on the pressure-normalized strain (<span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/pulse pressure [PP]) reduction using four different infrarenal EVAR endografts and our ultrasound elastography technique. Approximate graft oversizing percentages were 20% (30 mm phantom-graft combinations), 30% (28 mm phantom-graft combinations), and 50% (24 mm phantom-graft combinations).</p></div><div><h3>Methods</h3><p>Axisymmetric, 10% by mass polyvinyl alcohol phantoms were connected to a flow simulator. Ultrasound elastography was performed before and after implantation with the four different endografts: (1) 36 mm polyester/stainless steel, (2) 36 mm polyester/electropolished nitinol, (3) 35 mm polytetrafluoroethylene (PTFE)/nitinol, and (4) 36 mm nitinol/polyester/platinum-iridium. Five ultrasound cine loops were taken of each phantom-graft combination. They were analyzed over two different cardiac cycles (end-diastole to end-diastole), yielding a total of 10 maximum mean principal strain (<span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>) values. <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span> was divided by pulse pressure to yield pressure-normalized strain (<span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP). An analysis of variance was performed for graft comparisons. We calculated the average percent <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP reduction by manufacturer and percent oversizing. These values were used for linear regression analysis.</p></div><div><h3>Results</h3><p>Results from one-way analysis of variance showed a significant difference in <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP between the empty phantom condition and all oversizing conditions for all graft manufacturers (<em>F</em>(3, 56) = 106.7 [graft A], 132.7 [graft B], 106.5 [graft C], 105.7 [graft D], <em>P</em> < .0001 for grafts A-D). There was a significant difference when comparing the 50% condition with the 30% and 20% conditions across all manufacturers by post hoc analysis (<em>P</em> < .0001). No significant difference was found when comparing the 20% and 30% oversizing conditions for any of the manufacturers or when comparing <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP values across the manufacturers according to percent oversize. Linear regression demonstrated a significant positive correlation between the percent graft oversize and the all-graft average percent <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP reduction (<span><math><mrow><msup><mi>R</mi><mn>2</mn></msup></mrow></math></span> = 0.84, <em>P</em> < .0001).</p></div><div><h3>Conclusions</h3><p>This brief report suggests that a 10% increase in graft oversizing leads to an approximate 5.9% reduction in <span><math><mrow><mover><msub><mi>ε</mi><mrow><mi>ρ</mi><mo>+</mo></mrow></msub><mo>¯</mo></mover></mrow></math></span>/PP on average. Applied clinically, this increase may result in increased stiffness in axisymmetric vessels after EVAR. Further research is needed to determine if this is clinically significant.</p></div>\",\"PeriodicalId\":74035,\"journal\":{\"name\":\"JVS-vascular science\",\"volume\":\"5 \",\"pages\":\"Article 100198\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2666350324000099/pdfft?md5=174462f0b3289da4a0f6e8331d8e0f44&pid=1-s2.0-S2666350324000099-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"JVS-vascular science\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2666350324000099\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"JVS-vascular science","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666350324000099","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
Ultrasound elastography to quantify average percent pressure-normalized strain reduction associated with different aortic endografts in 3D-printed hydrogel phantoms
Objective
Strain has become a viable index for evaluating abdominal aortic aneurysm stability after endovascular aneurysm repair (EVAR). In addition, literature has shown that healthy aortic tissue requires a degree of strain to maintain homeostasis. This has led to the hypothesis that too much strain reduction conferred by a high degree of graft oversizing is detrimental to the aneurysm neck in the seal zone of abdominal aortic aneurysms after EVAR. We investigated this in a laboratory experiment by examining the effects that graft oversizing has on the pressure-normalized strain (/pulse pressure [PP]) reduction using four different infrarenal EVAR endografts and our ultrasound elastography technique. Approximate graft oversizing percentages were 20% (30 mm phantom-graft combinations), 30% (28 mm phantom-graft combinations), and 50% (24 mm phantom-graft combinations).
Methods
Axisymmetric, 10% by mass polyvinyl alcohol phantoms were connected to a flow simulator. Ultrasound elastography was performed before and after implantation with the four different endografts: (1) 36 mm polyester/stainless steel, (2) 36 mm polyester/electropolished nitinol, (3) 35 mm polytetrafluoroethylene (PTFE)/nitinol, and (4) 36 mm nitinol/polyester/platinum-iridium. Five ultrasound cine loops were taken of each phantom-graft combination. They were analyzed over two different cardiac cycles (end-diastole to end-diastole), yielding a total of 10 maximum mean principal strain () values. was divided by pulse pressure to yield pressure-normalized strain (/PP). An analysis of variance was performed for graft comparisons. We calculated the average percent /PP reduction by manufacturer and percent oversizing. These values were used for linear regression analysis.
Results
Results from one-way analysis of variance showed a significant difference in /PP between the empty phantom condition and all oversizing conditions for all graft manufacturers (F(3, 56) = 106.7 [graft A], 132.7 [graft B], 106.5 [graft C], 105.7 [graft D], P < .0001 for grafts A-D). There was a significant difference when comparing the 50% condition with the 30% and 20% conditions across all manufacturers by post hoc analysis (P < .0001). No significant difference was found when comparing the 20% and 30% oversizing conditions for any of the manufacturers or when comparing /PP values across the manufacturers according to percent oversize. Linear regression demonstrated a significant positive correlation between the percent graft oversize and the all-graft average percent /PP reduction ( = 0.84, P < .0001).
Conclusions
This brief report suggests that a 10% increase in graft oversizing leads to an approximate 5.9% reduction in /PP on average. Applied clinically, this increase may result in increased stiffness in axisymmetric vessels after EVAR. Further research is needed to determine if this is clinically significant.