{"title":"无创家用血糖监测仪测量血糖的准确性。","authors":"Himel Mondal, Sairavi Kiran Biri, Neha Pipil, Shaikat Mondal","doi":"10.4103/ijem.ijem_36_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Patients with diabetes mellitus monitor their blood glucose at home with monitors that require a drop of blood or use a continuous glucose monitoring device that implants a small needle in the body. However, both cause discomfort to the patients which may inhibit them for regular blood glucose checks. Photoplethysmogram (PPG) sensing technology is an approach for non-invasive blood glucose measurement and PPG sensors can be used to predict hypoglycaemic episodes. InChcek is a PPG-based non-invasive glucose monitor. However, its accuracy has not been checked yet. Hence, this study aimed to evaluate the accuracy of InCheck, a non-invasive glucose monitor for the estimation of blood glucose.</p><p><strong>Methods: </strong>In a tertiary care hospital, patients who came for blood glucose estimation were tested for blood glucose non-invasively on the InCheck device and then by the laboratory method (glucose oxidase-peroxidase). These two readings were compared. We used International Organization for Standardization (ISO) 15197:2013 (95% of values should be within ± 15 mg/dL of reference reading if reference glucose <100 mg/dL or within ± 15% of reference reading if reference glucose ≥100 mg/dL and 99% of the values should be within zones A and B in consensus error grid), and Surveillance Error Grid for analyzing the accuracy.</p><p><strong>Results: </strong>A total of 1223 samples were analyzed. There was a significant difference between the reference method glucose level (135 [Q1-Q3: 97 - 179] mg/dL) and monitor-measured glucose level (188.33 [Q1-Q3: 167.33-209.33] mg/dL) (P < 0.0001). A total of 18.5% of readings were following ISO 15197:2013 criteria and 67.25% of coordinates were within zone A and zone B of the consensus error grid. In the surveillance error grid analysis, about 29.4% of values were in the no-risk zone, 51.8% in slight risk, 18.6% in moderate risk, and 0.2% were in the severe risk zone.</p><p><strong>Conclusion: </strong>The accuracy of the InCheck device for the estimation of blood glucose by PPG signal is not following the recommended guidelines. Hence, further research is necessary for programming or redesigning the hardware and software for a better result from this optical sensor-based non-invasive home glucose monitor.</p>","PeriodicalId":13353,"journal":{"name":"Indian Journal of Endocrinology and Metabolism","volume":"28 1","pages":"60-64"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10962770/pdf/","citationCount":"0","resultStr":"{\"title\":\"Accuracy of a Non-Invasive Home Glucose Monitor for Measurement of Blood Glucose.\",\"authors\":\"Himel Mondal, Sairavi Kiran Biri, Neha Pipil, Shaikat Mondal\",\"doi\":\"10.4103/ijem.ijem_36_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Patients with diabetes mellitus monitor their blood glucose at home with monitors that require a drop of blood or use a continuous glucose monitoring device that implants a small needle in the body. However, both cause discomfort to the patients which may inhibit them for regular blood glucose checks. Photoplethysmogram (PPG) sensing technology is an approach for non-invasive blood glucose measurement and PPG sensors can be used to predict hypoglycaemic episodes. InChcek is a PPG-based non-invasive glucose monitor. However, its accuracy has not been checked yet. Hence, this study aimed to evaluate the accuracy of InCheck, a non-invasive glucose monitor for the estimation of blood glucose.</p><p><strong>Methods: </strong>In a tertiary care hospital, patients who came for blood glucose estimation were tested for blood glucose non-invasively on the InCheck device and then by the laboratory method (glucose oxidase-peroxidase). These two readings were compared. We used International Organization for Standardization (ISO) 15197:2013 (95% of values should be within ± 15 mg/dL of reference reading if reference glucose <100 mg/dL or within ± 15% of reference reading if reference glucose ≥100 mg/dL and 99% of the values should be within zones A and B in consensus error grid), and Surveillance Error Grid for analyzing the accuracy.</p><p><strong>Results: </strong>A total of 1223 samples were analyzed. There was a significant difference between the reference method glucose level (135 [Q1-Q3: 97 - 179] mg/dL) and monitor-measured glucose level (188.33 [Q1-Q3: 167.33-209.33] mg/dL) (P < 0.0001). A total of 18.5% of readings were following ISO 15197:2013 criteria and 67.25% of coordinates were within zone A and zone B of the consensus error grid. In the surveillance error grid analysis, about 29.4% of values were in the no-risk zone, 51.8% in slight risk, 18.6% in moderate risk, and 0.2% were in the severe risk zone.</p><p><strong>Conclusion: </strong>The accuracy of the InCheck device for the estimation of blood glucose by PPG signal is not following the recommended guidelines. Hence, further research is necessary for programming or redesigning the hardware and software for a better result from this optical sensor-based non-invasive home glucose monitor.</p>\",\"PeriodicalId\":13353,\"journal\":{\"name\":\"Indian Journal of Endocrinology and Metabolism\",\"volume\":\"28 1\",\"pages\":\"60-64\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10962770/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Indian Journal of Endocrinology and Metabolism\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/ijem.ijem_36_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/26 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Endocrinology and Metabolism","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/ijem.ijem_36_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/26 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

导言:糖尿病患者在家中使用需要滴血的血糖监测仪或使用将小针植入体内的连续血糖监测仪来监测血糖。然而,这两种方法都会给患者带来不适,从而妨碍他们定期检查血糖。光敏血糖仪(PPG)传感技术是一种无创血糖测量方法,PPG 传感器可用于预测低血糖发作。InChcek 是一种基于 PPG 的无创血糖监测仪。然而,其准确性尚未得到检验。因此,本研究旨在评估无创血糖监测仪 InCheck 估算血糖的准确性:方法:在一家三级甲等医院,对前来估测血糖的患者先用 InCheck 设备进行无创血糖检测,然后再用实验室方法(葡萄糖氧化酶-过氧化物酶)进行检测。对这两种读数进行比较。我们采用了国际标准化组织 (ISO) 15197:2013 标准(如果参考血糖结果,95% 的数值应在参考读数的 ± 15 mg/dL 范围内):共分析了 1223 个样本。参考方法的血糖水平(135 [Q1-Q3: 97 - 179] mg/dL)与监测仪测量的血糖水平(188.33 [Q1-Q3: 167.33-209.33] mg/dL)之间存在明显差异(P < 0.0001)。共有 18.5% 的读数符合 ISO 15197:2013 标准,67.25% 的坐标在共识误差网格的 A 区和 B 区内。在监测误差网格分析中,约 29.4% 的数值处于无风险区,51.8% 处于轻微风险区,18.6% 处于中度风险区,0.2% 处于严重风险区:结论:InCheck 设备通过 PPG 信号估算血糖的准确性并不符合建议的指导方针。因此,有必要开展进一步研究,对硬件和软件进行编程或重新设计,使这种基于光学传感器的无创家用血糖监测仪取得更好的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Accuracy of a Non-Invasive Home Glucose Monitor for Measurement of Blood Glucose.

Introduction: Patients with diabetes mellitus monitor their blood glucose at home with monitors that require a drop of blood or use a continuous glucose monitoring device that implants a small needle in the body. However, both cause discomfort to the patients which may inhibit them for regular blood glucose checks. Photoplethysmogram (PPG) sensing technology is an approach for non-invasive blood glucose measurement and PPG sensors can be used to predict hypoglycaemic episodes. InChcek is a PPG-based non-invasive glucose monitor. However, its accuracy has not been checked yet. Hence, this study aimed to evaluate the accuracy of InCheck, a non-invasive glucose monitor for the estimation of blood glucose.

Methods: In a tertiary care hospital, patients who came for blood glucose estimation were tested for blood glucose non-invasively on the InCheck device and then by the laboratory method (glucose oxidase-peroxidase). These two readings were compared. We used International Organization for Standardization (ISO) 15197:2013 (95% of values should be within ± 15 mg/dL of reference reading if reference glucose <100 mg/dL or within ± 15% of reference reading if reference glucose ≥100 mg/dL and 99% of the values should be within zones A and B in consensus error grid), and Surveillance Error Grid for analyzing the accuracy.

Results: A total of 1223 samples were analyzed. There was a significant difference between the reference method glucose level (135 [Q1-Q3: 97 - 179] mg/dL) and monitor-measured glucose level (188.33 [Q1-Q3: 167.33-209.33] mg/dL) (P < 0.0001). A total of 18.5% of readings were following ISO 15197:2013 criteria and 67.25% of coordinates were within zone A and zone B of the consensus error grid. In the surveillance error grid analysis, about 29.4% of values were in the no-risk zone, 51.8% in slight risk, 18.6% in moderate risk, and 0.2% were in the severe risk zone.

Conclusion: The accuracy of the InCheck device for the estimation of blood glucose by PPG signal is not following the recommended guidelines. Hence, further research is necessary for programming or redesigning the hardware and software for a better result from this optical sensor-based non-invasive home glucose monitor.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Indian Journal of Endocrinology and Metabolism
Indian Journal of Endocrinology and Metabolism Medicine-Endocrinology, Diabetes and Metabolism
CiteScore
2.10
自引率
0.00%
发文量
75
期刊介绍: The Indian Journal of Endocrinology and Metabolism (IJEM) aims to function as the global face of Indian endocrinology research. It aims to act as a bridge between global and national advances in this field. The journal publishes thought-provoking editorials, comprehensive reviews, cutting-edge original research, focused brief communications and insightful letters to editor. The journal encourages authors to submit articles addressing aspects of science related to Endocrinology and Metabolism in particular Diabetology. Articles related to Clinical and Tropical endocrinology are especially encouraged. Sub-topic based Supplements are published regularly. This allows the journal to highlight issues relevant to Endocrine practitioners working in India as well as other countries. IJEM is free access in the true sense of the word, (it charges neither authors nor readers) and this enhances its global appeal.
期刊最新文献
The Burden of Diabetic Foot Ulcers in Urban India: A Community Healthcare Setup-Based Study. Approved and Emerging Hormone-Based Anti-Obesity Medications: A Review Article. Assessment of Correlation Between Glycaemic Variability Indices and Measures of Hypoglycaemia in Patients with Type 2 Diabetes Mellitus. Association Between Type 2 Diabetes and Hypogonadism in India: An Observational Study. Carotid Intima-Media Thickness and Sheehan's Syndrome.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1