{"title":"比较两种头颅立体定向放射治疗无框架固定系统的前瞻性随机研究","authors":"Dylan Callens ,&nbsp;Chahrazad Benazzouz ,&nbsp;Lise Stessens ,&nbsp;Wout Piot ,&nbsp;An Nulens ,&nbsp;Maarten Lambrecht ,&nbsp;Patrick Berkovic ,&nbsp;Jean-François Daisne","doi":"10.1016/j.tipsro.2024.100249","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>The Dual Shell Encompass Fibreplast™ System (DS-Encompass) by CQ Medical™ is validated for frameless immobilization in stereotactic brain radiotherapy. An alternative mask model has been proposed with the rear shell replaced by a Moldcare® cushion (M−Encompass). To validate the use of this model in our cranial stereotactic workflow method including HyperArc™, we performed a prospective randomized study comparing inter-and intrafractional motion and patients comfort between both masks.</p></div><div><h3>Materials &amp; Methods</h3><p>A prospective randomized study between DS-Encompass and M−Encompass was conducted involving 60 participants. Stratification between DS-Encompass and M−Encompass was carried out based on the fractionation scheme. Treatment plans were created with HyperArc™. During treatment, surface guidance was used for patient positioning and monitoring. A pre-treatment cone-beam CT (CBCT) was acquired to correct interfractional motion and a post-treatment CBCT was acquired to quantify the intrafractional motion. Patients reported comfort was analyzed with a Likert-scale at the end of the treatment. Unpaired t-tests were conducted to determine the level of significance.</p></div><div><h3>Results</h3><p>No significant difference in interfractional translations is present. A significant difference is revealed in roll-axis rotation, where DS-Encompass allows for smaller deviations. Since interfractional motion can be corrected through daily CBCT-scans and 6D-couch corrections, they are clinically irrelevant. Intrafractional motion does not differ significantly and remains below 0.5 mm and 0.5° for both systems. There is no statistical difference in patient-reported comfort.</p></div><div><h3>Conclusion</h3><p>We conclude that Encompass with Moldcare offers a safe alternative to Duall Shell Encompass for non-coplanar stereotactic brain radiotherapy. There is no significant difference in intrafractional motion nor difference in comfort levels.</p></div>","PeriodicalId":36328,"journal":{"name":"Technical Innovations and Patient Support in Radiation Oncology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-04-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2405632424000167/pdfft?md5=963922d30176313bcfa12c7f536cbb01&pid=1-s2.0-S2405632424000167-main.pdf","citationCount":"0","resultStr":"{\"title\":\"A prospective randomized study comparing two frameless immobilization systems for cranial stereotactic radiotherapy\",\"authors\":\"Dylan Callens ,&nbsp;Chahrazad Benazzouz ,&nbsp;Lise Stessens ,&nbsp;Wout Piot ,&nbsp;An Nulens ,&nbsp;Maarten Lambrecht ,&nbsp;Patrick Berkovic ,&nbsp;Jean-François Daisne\",\"doi\":\"10.1016/j.tipsro.2024.100249\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>The Dual Shell Encompass Fibreplast™ System (DS-Encompass) by CQ Medical™ is validated for frameless immobilization in stereotactic brain radiotherapy. An alternative mask model has been proposed with the rear shell replaced by a Moldcare® cushion (M−Encompass). To validate the use of this model in our cranial stereotactic workflow method including HyperArc™, we performed a prospective randomized study comparing inter-and intrafractional motion and patients comfort between both masks.</p></div><div><h3>Materials &amp; Methods</h3><p>A prospective randomized study between DS-Encompass and M−Encompass was conducted involving 60 participants. Stratification between DS-Encompass and M−Encompass was carried out based on the fractionation scheme. Treatment plans were created with HyperArc™. During treatment, surface guidance was used for patient positioning and monitoring. A pre-treatment cone-beam CT (CBCT) was acquired to correct interfractional motion and a post-treatment CBCT was acquired to quantify the intrafractional motion. Patients reported comfort was analyzed with a Likert-scale at the end of the treatment. Unpaired t-tests were conducted to determine the level of significance.</p></div><div><h3>Results</h3><p>No significant difference in interfractional translations is present. A significant difference is revealed in roll-axis rotation, where DS-Encompass allows for smaller deviations. Since interfractional motion can be corrected through daily CBCT-scans and 6D-couch corrections, they are clinically irrelevant. Intrafractional motion does not differ significantly and remains below 0.5 mm and 0.5° for both systems. There is no statistical difference in patient-reported comfort.</p></div><div><h3>Conclusion</h3><p>We conclude that Encompass with Moldcare offers a safe alternative to Duall Shell Encompass for non-coplanar stereotactic brain radiotherapy. There is no significant difference in intrafractional motion nor difference in comfort levels.</p></div>\",\"PeriodicalId\":36328,\"journal\":{\"name\":\"Technical Innovations and Patient Support in Radiation Oncology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-04-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2405632424000167/pdfft?md5=963922d30176313bcfa12c7f536cbb01&pid=1-s2.0-S2405632424000167-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Technical Innovations and Patient Support in Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405632424000167\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"Nursing\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Technical Innovations and Patient Support in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405632424000167","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"Nursing","Score":null,"Total":0}
引用次数: 0

摘要

引言 CQ Medical™ 公司的双壳 Encompass Fibreplast™ 系统(DS-Encompass)已通过验证,可用于脑立体定向放射治疗中的无框架固定。还提出了另一种面罩模型,其后壳由 Moldcare® 软垫取代(M-Encompass)。为了验证该模型在包括 HyperArc™ 在内的头颅立体定向工作流程方法中的应用,我们进行了一项前瞻性随机研究,比较了两种面罩之间的点间和点内运动以及患者的舒适度。根据分层方案对 DS-Encompass 和 M-Encompass 进行了分层。使用 HyperArc™ 创建了治疗计划。在治疗过程中,表面引导用于患者定位和监测。采集治疗前锥形束 CT (CBCT) 以纠正点间移动,采集治疗后 CBCT 以量化点内移动。治疗结束后,采用李克特量表对患者报告的舒适度进行分析。进行非配对 t 检验以确定显著性水平。在滚动轴旋转方面存在明显差异,DS-Encompass 允许的偏差较小。由于切迹间运动可以通过日常的 CBCT 扫描和 6D 凹陷矫正进行校正,因此与临床无关。两个系统的切迹内移动差异不大,都保持在 0.5 毫米和 0.5°以下。结论我们得出结论,在非共面立体定向脑放射治疗中,带有 Moldcare 的 Encompass 可以安全地替代 Duall Shell Encompass。在点内运动和舒适度方面没有明显差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A prospective randomized study comparing two frameless immobilization systems for cranial stereotactic radiotherapy

Introduction

The Dual Shell Encompass Fibreplast™ System (DS-Encompass) by CQ Medical™ is validated for frameless immobilization in stereotactic brain radiotherapy. An alternative mask model has been proposed with the rear shell replaced by a Moldcare® cushion (M−Encompass). To validate the use of this model in our cranial stereotactic workflow method including HyperArc™, we performed a prospective randomized study comparing inter-and intrafractional motion and patients comfort between both masks.

Materials & Methods

A prospective randomized study between DS-Encompass and M−Encompass was conducted involving 60 participants. Stratification between DS-Encompass and M−Encompass was carried out based on the fractionation scheme. Treatment plans were created with HyperArc™. During treatment, surface guidance was used for patient positioning and monitoring. A pre-treatment cone-beam CT (CBCT) was acquired to correct interfractional motion and a post-treatment CBCT was acquired to quantify the intrafractional motion. Patients reported comfort was analyzed with a Likert-scale at the end of the treatment. Unpaired t-tests were conducted to determine the level of significance.

Results

No significant difference in interfractional translations is present. A significant difference is revealed in roll-axis rotation, where DS-Encompass allows for smaller deviations. Since interfractional motion can be corrected through daily CBCT-scans and 6D-couch corrections, they are clinically irrelevant. Intrafractional motion does not differ significantly and remains below 0.5 mm and 0.5° for both systems. There is no statistical difference in patient-reported comfort.

Conclusion

We conclude that Encompass with Moldcare offers a safe alternative to Duall Shell Encompass for non-coplanar stereotactic brain radiotherapy. There is no significant difference in intrafractional motion nor difference in comfort levels.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.10
自引率
0.00%
发文量
48
审稿时长
67 days
期刊最新文献
The status quo of global geriatric radiation oncology education: A scoping review A systematic review of prostate bed motion and anisotropic margins in post-prostatectomy external beam radiotherapy International virtual radiation therapy professional development: Reflections on a twinning collaboration between a low/middle and high income country A code orange for traffic-light-protocols as a communication mechanism in IGRT On the trail of CBCT-guided adaptive rectal boost radiotherapy, does daily delineation require a radiation oncologist?
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1