DI-60 数字图像分析仪检测血小板凝集和获得准确血小板计数的可靠性。

IF 4 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY Annals of Laboratory Medicine Pub Date : 2024-04-19 DOI:10.3343/alm.2024.0003
Suhyeon Woo, Bohyun Kim, Nam Hun Heo, Min-Sun Kim, Young Ahn Yoon, Young-Jin Choi
{"title":"DI-60 数字图像分析仪检测血小板凝集和获得准确血小板计数的可靠性。","authors":"Suhyeon Woo, Bohyun Kim, Nam Hun Heo, Min-Sun Kim, Young Ahn Yoon, Young-Jin Choi","doi":"10.3343/alm.2024.0003","DOIUrl":null,"url":null,"abstract":"Pseudothrombocytopenia caused by platelet clumping (PC) can lead to unnecessary platelet transfusions or underdiagnosis of hematologic neoplasms. To overcome these limitations, we assessed the capacity of the Sysmex DI-60 digital morphology analyzer (Sysmex, Kobe, Japan) for detecting PC and determining an accurate platelet count in the presence of PC. For this purpose, 135 samples with or without PC (groups Y and N, respectively) were processed by an examiner (a hematologic specialist) using both the Sysmex XN-9000 and DI-60 analyzers. Although the platelet aggregate (PA) and giant platelet (GP) counts reported by the DI-60 and the examiner exhibited strong correlations, they proved inadequate as effective indicators for screening samples containing PC. Between the PA and GP counts and four platelet indices (the platelet distribution width [PDW], mean platelet volume [MPV], platelet large cell ratio [P_LCR], and plateletcrit [PCT]) reported by the XN-9000, we observed statistically significant correlations (both overall and with group Y), but they were relatively weak. The platelet counts determined using the DI-60 and light microscopy in group Y showed substantial variations. Although the performance of the DI-60 was reliable for detecting PA and GP in smear images, such fixed areas are not representative of whole samples. Further, in the presence of PC, the resulting platelet counts determined using the DI-60 were not sufficiently accurate to be accepted as the final count.","PeriodicalId":8421,"journal":{"name":"Annals of Laboratory Medicine","volume":"13 1","pages":""},"PeriodicalIF":4.0000,"publicationDate":"2024-04-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Reliability of the DI-60 Digital Image Analyzer for Detecting Platelet Clumping and Obtaining Accurate Platelet Counts.\",\"authors\":\"Suhyeon Woo, Bohyun Kim, Nam Hun Heo, Min-Sun Kim, Young Ahn Yoon, Young-Jin Choi\",\"doi\":\"10.3343/alm.2024.0003\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Pseudothrombocytopenia caused by platelet clumping (PC) can lead to unnecessary platelet transfusions or underdiagnosis of hematologic neoplasms. To overcome these limitations, we assessed the capacity of the Sysmex DI-60 digital morphology analyzer (Sysmex, Kobe, Japan) for detecting PC and determining an accurate platelet count in the presence of PC. For this purpose, 135 samples with or without PC (groups Y and N, respectively) were processed by an examiner (a hematologic specialist) using both the Sysmex XN-9000 and DI-60 analyzers. Although the platelet aggregate (PA) and giant platelet (GP) counts reported by the DI-60 and the examiner exhibited strong correlations, they proved inadequate as effective indicators for screening samples containing PC. Between the PA and GP counts and four platelet indices (the platelet distribution width [PDW], mean platelet volume [MPV], platelet large cell ratio [P_LCR], and plateletcrit [PCT]) reported by the XN-9000, we observed statistically significant correlations (both overall and with group Y), but they were relatively weak. The platelet counts determined using the DI-60 and light microscopy in group Y showed substantial variations. Although the performance of the DI-60 was reliable for detecting PA and GP in smear images, such fixed areas are not representative of whole samples. Further, in the presence of PC, the resulting platelet counts determined using the DI-60 were not sufficiently accurate to be accepted as the final count.\",\"PeriodicalId\":8421,\"journal\":{\"name\":\"Annals of Laboratory Medicine\",\"volume\":\"13 1\",\"pages\":\"\"},\"PeriodicalIF\":4.0000,\"publicationDate\":\"2024-04-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annals of Laboratory Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3343/alm.2024.0003\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3343/alm.2024.0003","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

血小板凝集(PC)引起的假性血小板减少症会导致不必要的血小板输注或血液肿瘤诊断不足。为了克服这些局限性,我们评估了 Sysmex DI-60 数字形态分析仪(Sysmex,日本神户)检测 PC 和确定 PC 存在时血小板准确计数的能力。为此,检查人员(血液学专家)使用 Sysmex XN-9000 和 DI-60 分析仪处理了 135 份有或没有 PC 的样本(分别为 Y 组和 N 组)。尽管 DI-60 和检验员报告的血小板聚集(PA)和巨血小板(GP)计数显示出很强的相关性,但它们不足以作为筛查含 PC 样品的有效指标。在 PA 和 GP 计数与 XN-9000 报告的四项血小板指数(血小板分布宽度 [PDW]、平均血小板体积 [MPV]、血小板大细胞比 [P_LCR] 和血小板比容 [PCT])之间,我们观察到了统计学上的显著相关性(包括总体相关性和与 Y 组的相关性),但相关性相对较弱。在 Y 组中,使用 DI-60 和光学显微镜测定的血小板计数差异很大。虽然 DI-60 在检测涂片图像中的 PA 和 GP 方面性能可靠,但这种固定区域并不能代表整个样本。此外,在存在 PC 的情况下,使用 DI-60 确定的血小板计数不够准确,不能作为最终计数。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Reliability of the DI-60 Digital Image Analyzer for Detecting Platelet Clumping and Obtaining Accurate Platelet Counts.
Pseudothrombocytopenia caused by platelet clumping (PC) can lead to unnecessary platelet transfusions or underdiagnosis of hematologic neoplasms. To overcome these limitations, we assessed the capacity of the Sysmex DI-60 digital morphology analyzer (Sysmex, Kobe, Japan) for detecting PC and determining an accurate platelet count in the presence of PC. For this purpose, 135 samples with or without PC (groups Y and N, respectively) were processed by an examiner (a hematologic specialist) using both the Sysmex XN-9000 and DI-60 analyzers. Although the platelet aggregate (PA) and giant platelet (GP) counts reported by the DI-60 and the examiner exhibited strong correlations, they proved inadequate as effective indicators for screening samples containing PC. Between the PA and GP counts and four platelet indices (the platelet distribution width [PDW], mean platelet volume [MPV], platelet large cell ratio [P_LCR], and plateletcrit [PCT]) reported by the XN-9000, we observed statistically significant correlations (both overall and with group Y), but they were relatively weak. The platelet counts determined using the DI-60 and light microscopy in group Y showed substantial variations. Although the performance of the DI-60 was reliable for detecting PA and GP in smear images, such fixed areas are not representative of whole samples. Further, in the presence of PC, the resulting platelet counts determined using the DI-60 were not sufficiently accurate to be accepted as the final count.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Annals of Laboratory Medicine
Annals of Laboratory Medicine MEDICAL LABORATORY TECHNOLOGY-
CiteScore
8.30
自引率
12.20%
发文量
100
审稿时长
6-12 weeks
期刊介绍: Annals of Laboratory Medicine is the official journal of Korean Society for Laboratory Medicine. The journal title has been recently changed from the Korean Journal of Laboratory Medicine (ISSN, 1598-6535) from the January issue of 2012. The JCR 2017 Impact factor of Ann Lab Med was 1.916.
期刊最新文献
Ability of the Capillary Electrophoresis-based HbA1c Assay to Detect Rare Hemoglobin Variants. Performance Evaluation of the LabGenius C-CT/NG-BMX Assay for Chlamydia trachomatis and Neisseria gonorrhoeae Detection. The First Korean Case of MAN1B1-Congenital Disorder of Glycosylation Diagnosed Using Whole-exome Sequencing and Matrix-assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry. Endogenous Thrombin Potential Level Helps Predict High Blood Loss in Patients Undergoing Cardiac Surgery. TP53 Mutation Status in Myelodysplastic Neoplasm and Acute Myeloid Leukemia: Impact of Reclassification Based on the 5th WHO and International Consensus Classification Criteria: A Korean Multicenter Study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1