{"title":"通过压力监测和回路冲洗量化儿科 ECMO 患者的潜在输液量","authors":"Steven Robertson, Katherine White","doi":"10.1051/ject/2024007","DOIUrl":null,"url":null,"abstract":"Background: Pressure monitoring on pediatric ECMO circuits is used to aid in the evaluation of patient hemodynamics and circuit health. ELSO recommends monitoring pressures on the venous line, pre, and post oxygenator. In order to keep pressure ports patent, crystalloid can be used as a flush. The fluid transfused to the patient through these lines can be challenging to quantify accurately due to variance in clinician practice. Currently, there is no published data or practice suggestions on this topic. Methods: This study reports on experiments used to quantify passively and actively infused crystalloid using an Edwards TruWave transducer and pressurized IV bag. Frequent negative and positive pressures measured during neonatal ECMO runs were simulated in order to quantify infused volumes. Results: When the pressure transducer and associated tubing is kept patent by using a pressurized IV bag, per the instructions for use, the daily volume transfused was found to be close to a typical neonates total blood volume. Conclusions: An alternative method for pressure transducer line patency is discussed which includes the use of an automated syringe pump. This allows for more precise infusion volumes within a 24 hour period. Further study is recommended to develop and publish best practices.","PeriodicalId":506828,"journal":{"name":"The Journal of ExtraCorporeal Technology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Quantifying Potential Fluid Transfused through Pressure Monitoring and Circuit Flushes in Pediatric ECMO Patients\",\"authors\":\"Steven Robertson, Katherine White\",\"doi\":\"10.1051/ject/2024007\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background: Pressure monitoring on pediatric ECMO circuits is used to aid in the evaluation of patient hemodynamics and circuit health. ELSO recommends monitoring pressures on the venous line, pre, and post oxygenator. In order to keep pressure ports patent, crystalloid can be used as a flush. The fluid transfused to the patient through these lines can be challenging to quantify accurately due to variance in clinician practice. Currently, there is no published data or practice suggestions on this topic. Methods: This study reports on experiments used to quantify passively and actively infused crystalloid using an Edwards TruWave transducer and pressurized IV bag. Frequent negative and positive pressures measured during neonatal ECMO runs were simulated in order to quantify infused volumes. Results: When the pressure transducer and associated tubing is kept patent by using a pressurized IV bag, per the instructions for use, the daily volume transfused was found to be close to a typical neonates total blood volume. Conclusions: An alternative method for pressure transducer line patency is discussed which includes the use of an automated syringe pump. This allows for more precise infusion volumes within a 24 hour period. Further study is recommended to develop and publish best practices.\",\"PeriodicalId\":506828,\"journal\":{\"name\":\"The Journal of ExtraCorporeal Technology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-04-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The Journal of ExtraCorporeal Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1051/ject/2024007\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of ExtraCorporeal Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1051/ject/2024007","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

背景:儿科 ECMO 循环的压力监测用于帮助评估患者血液动力学和循环健康状况。ELSO 建议监测静脉管路、氧合器前和氧合器后的压力。为保持压力端口通畅,可使用晶体液作为冲洗液。由于临床医生的做法各不相同,通过这些管路向患者输注的液体很难准确量化。目前,还没有这方面的公开数据或实践建议。方法:本研究报告了使用 Edwards TruWave 传感器和加压静脉输液袋对被动和主动输注的晶体液进行量化的实验。模拟新生儿 ECMO 运行期间测量到的频繁负压和正压,以量化输注量。结果:根据使用说明,使用加压静脉输液袋保持压力传感器和相关管道通畅时,发现每日输注量接近典型新生儿的总血量。结论讨论了压力传感器管路通畅的替代方法,包括使用自动注射泵。这种方法可在 24 小时内实现更精确的输液量。建议开展进一步研究,以制定和发布最佳实践方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Quantifying Potential Fluid Transfused through Pressure Monitoring and Circuit Flushes in Pediatric ECMO Patients
Background: Pressure monitoring on pediatric ECMO circuits is used to aid in the evaluation of patient hemodynamics and circuit health. ELSO recommends monitoring pressures on the venous line, pre, and post oxygenator. In order to keep pressure ports patent, crystalloid can be used as a flush. The fluid transfused to the patient through these lines can be challenging to quantify accurately due to variance in clinician practice. Currently, there is no published data or practice suggestions on this topic. Methods: This study reports on experiments used to quantify passively and actively infused crystalloid using an Edwards TruWave transducer and pressurized IV bag. Frequent negative and positive pressures measured during neonatal ECMO runs were simulated in order to quantify infused volumes. Results: When the pressure transducer and associated tubing is kept patent by using a pressurized IV bag, per the instructions for use, the daily volume transfused was found to be close to a typical neonates total blood volume. Conclusions: An alternative method for pressure transducer line patency is discussed which includes the use of an automated syringe pump. This allows for more precise infusion volumes within a 24 hour period. Further study is recommended to develop and publish best practices.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Increased white blood cell count is associated with an increased demand for unfractionated heparin during veno-arterial extracorporeal oxygenation in lung transplantation A case of the effective inhalation of nitric oxide therapy for caused severe pulmonary hypertension with protamine neutralization of systemic heparinization during totally endoscopic minimally invasive cardiac surgery Cardiopulmonary Bypass in a Pediatric Patient with Factor XII Deficiency High Pressure Excursion in a Radial Design Oxygenator Outcomes in patients who received ECMO and/or volatile anesthetics as rescue therapies for status asthmaticus
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1