{"title":"将 CloudConnect 作为静脉注射疗法患者的输液监控器:一项探索性、非盲法、单中心试验。","authors":"Carlos Mejia-Chew MD , Brett Heuring PharmD , Jeffrey Salmons RN , Lori Weilmuenster PharmD , Joe Beggs BEng , Glen Kleinschmidt BEng , Jake Eshelman BEng , Sai Dodda PharmD","doi":"10.1016/j.curtheres.2024.100747","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Intravenous (IV) infusion therapy is commonly used in health care settings. However, IV therapy at home can be challenging because it relies on the patient's ability to manage multiple medications correctly, which may lead to decreased treatment adherence.</p></div><div><h3>Objective</h3><p>We aimed to assess the usability and accuracy of the IVsight monitor in capturing IV infusion data compared to manual recording.</p></div><div><h3>Methods</h3><p>A prospective, single-center, usability study involving patients receiving IV infusion therapy at one of the BJC's Home Infusion suites was conducted to evaluate the accuracy, performance, and acceptability of the device IVsight as a monitor for IV infusions.</p></div><div><h3>Results</h3><p>Of the 15 participants, the median (IQR) age was 46 years (36–55), 8 (53%) were female, and 13 (87%) were non-Hispanic white. Each participant received a median (IQR) of 4 (4–5) infusions during the study, and 68 infusions were observed overall. The intraclass correlation coefficient between the IVsight measurement and manual recording of infusion duration in seconds was excellent (ICC 0.97, 95% confidence interval 0.96–0.98). The Bland–Altman plot visually showed an acceptable limit of agreement for the 2 measurements, and the linear regression analysis revealed no significant proportional bias between the 2 methods for measuring the IV infusion time. None of the participants thought that IVsight was difficult to hold, use, clean, or store. Only one participant was concerned that the device could interfere with the IV infusion, and all participants felt comfortable with the device transmitting data to their health care providers.</p></div><div><h3>Conclusions</h3><p>The IVsight infusion monitoring device showed near-perfect agreement on the recorded IV infusion duration compared with manual recording, and good acceptability among the study participants.</p></div>","PeriodicalId":10920,"journal":{"name":"Current Therapeutic Research-clinical and Experimental","volume":"100 ","pages":"Article 100747"},"PeriodicalIF":1.6000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0011393X24000171/pdfft?md5=80fd87cb1f8920d6b08acb8349f56587&pid=1-s2.0-S0011393X24000171-main.pdf","citationCount":"0","resultStr":"{\"title\":\"IVsight as an Infusion Monitor for Patients Receiving Intravenous Therapy: An Exploratory, Unblinded, Single-Center Trial\",\"authors\":\"Carlos Mejia-Chew MD , Brett Heuring PharmD , Jeffrey Salmons RN , Lori Weilmuenster PharmD , Joe Beggs BEng , Glen Kleinschmidt BEng , Jake Eshelman BEng , Sai Dodda PharmD\",\"doi\":\"10.1016/j.curtheres.2024.100747\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>Intravenous (IV) infusion therapy is commonly used in health care settings. However, IV therapy at home can be challenging because it relies on the patient's ability to manage multiple medications correctly, which may lead to decreased treatment adherence.</p></div><div><h3>Objective</h3><p>We aimed to assess the usability and accuracy of the IVsight monitor in capturing IV infusion data compared to manual recording.</p></div><div><h3>Methods</h3><p>A prospective, single-center, usability study involving patients receiving IV infusion therapy at one of the BJC's Home Infusion suites was conducted to evaluate the accuracy, performance, and acceptability of the device IVsight as a monitor for IV infusions.</p></div><div><h3>Results</h3><p>Of the 15 participants, the median (IQR) age was 46 years (36–55), 8 (53%) were female, and 13 (87%) were non-Hispanic white. Each participant received a median (IQR) of 4 (4–5) infusions during the study, and 68 infusions were observed overall. The intraclass correlation coefficient between the IVsight measurement and manual recording of infusion duration in seconds was excellent (ICC 0.97, 95% confidence interval 0.96–0.98). The Bland–Altman plot visually showed an acceptable limit of agreement for the 2 measurements, and the linear regression analysis revealed no significant proportional bias between the 2 methods for measuring the IV infusion time. None of the participants thought that IVsight was difficult to hold, use, clean, or store. Only one participant was concerned that the device could interfere with the IV infusion, and all participants felt comfortable with the device transmitting data to their health care providers.</p></div><div><h3>Conclusions</h3><p>The IVsight infusion monitoring device showed near-perfect agreement on the recorded IV infusion duration compared with manual recording, and good acceptability among the study participants.</p></div>\",\"PeriodicalId\":10920,\"journal\":{\"name\":\"Current Therapeutic Research-clinical and Experimental\",\"volume\":\"100 \",\"pages\":\"Article 100747\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S0011393X24000171/pdfft?md5=80fd87cb1f8920d6b08acb8349f56587&pid=1-s2.0-S0011393X24000171-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Current Therapeutic Research-clinical and Experimental\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0011393X24000171\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICINE, RESEARCH & EXPERIMENTAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current Therapeutic Research-clinical and Experimental","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0011393X24000171","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
IVsight as an Infusion Monitor for Patients Receiving Intravenous Therapy: An Exploratory, Unblinded, Single-Center Trial
Background
Intravenous (IV) infusion therapy is commonly used in health care settings. However, IV therapy at home can be challenging because it relies on the patient's ability to manage multiple medications correctly, which may lead to decreased treatment adherence.
Objective
We aimed to assess the usability and accuracy of the IVsight monitor in capturing IV infusion data compared to manual recording.
Methods
A prospective, single-center, usability study involving patients receiving IV infusion therapy at one of the BJC's Home Infusion suites was conducted to evaluate the accuracy, performance, and acceptability of the device IVsight as a monitor for IV infusions.
Results
Of the 15 participants, the median (IQR) age was 46 years (36–55), 8 (53%) were female, and 13 (87%) were non-Hispanic white. Each participant received a median (IQR) of 4 (4–5) infusions during the study, and 68 infusions were observed overall. The intraclass correlation coefficient between the IVsight measurement and manual recording of infusion duration in seconds was excellent (ICC 0.97, 95% confidence interval 0.96–0.98). The Bland–Altman plot visually showed an acceptable limit of agreement for the 2 measurements, and the linear regression analysis revealed no significant proportional bias between the 2 methods for measuring the IV infusion time. None of the participants thought that IVsight was difficult to hold, use, clean, or store. Only one participant was concerned that the device could interfere with the IV infusion, and all participants felt comfortable with the device transmitting data to their health care providers.
Conclusions
The IVsight infusion monitoring device showed near-perfect agreement on the recorded IV infusion duration compared with manual recording, and good acceptability among the study participants.
期刊介绍:
We also encourage the submission of manuscripts presenting preclinical and very preliminary research that may stimulate further investigation of potentially relevant findings, as well as in-depth review articles on specific therapies or disease states, and applied health delivery or pharmacoeconomics.
CTR encourages and supports the submission of manuscripts describing:
• Interventions designed to understand or improve human health, disease treatment or disease prevention;
• Studies that focus on problems that are uncommon in resource-rich countries;
• Research that is "under-published" because of limited access to monetary resources such as English language support and Open Access fees (CTR offers deeply discounted English language editing);
• Republication of articles previously published in non-English journals (eg, evidence-based guidelines) which could be useful if translated into English;
• Preclinical and clinical product development studies that are not pursued for further investigation based upon early phase results.