{"title":"用于矫正扭转性斜视的可穿戴设备。","authors":"Charles Malleson, Jean-Yves Guillemaut","doi":"10.1097/OPX.0000000000002122","DOIUrl":null,"url":null,"abstract":"<p><strong>Significance: </strong>A wearable optical apparatus that compensates for eye misalignment (strabismus) to correct for double vision (diplopia) is proposed. In contrast to prism lenses, commonly used to compensate for horizontal and/or vertical misalignment, the proposed approach is able to compensate for any combination of horizontal, vertical, and torsional misalignment.</p><p><strong>Purpose: </strong>If the action of the extraocular muscles is compromised (e.g., by nerve damage), a patient may lose their ability to maintain visual alignment, negatively affecting their binocular fusion and stereo depth perception capability. Torsional misalignment cannot be mitigated by standard Fresnel prism lenses. Surgical procedures intended to correct torsional misalignment may be unpredictable. A wearable device able to rectify visual alignment and restore stereo depth perception without surgical intervention could potentially be of great value to people with strabismus.</p><p><strong>Methods: </strong>We propose a novel lightweight wearable optical device for visual alignment correction. The device comprises two mirrors and a Fresnel prism, arranged in such a way that together they rotationally shift the view seen by the affected eye horizontally, vertically, and torsionally. The extent of the alignment correction on each axis can be arbitrarily adjusted according to the patient's particular misalignment characteristics.</p><p><strong>Results: </strong>The proposed approach was tested by computer simulation, and a prototype device was manufactured. The prototype device was tested by a strabismus patient exhibiting horizontal and torsional misalignment. In these tests, the device was found to function as intended, allowing the patient to enjoy binocular fusion and stereo depth perception while wearing the device for daily activities over a period of several months.</p><p><strong>Conclusions: </strong>The proposed device is effective in correcting arbitrary horizontal, vertical, and torsional misalignment of the eyes. The results of the initial testing performed are highly encouraging. Future study is warranted to formally assess the effectiveness of the device on multiple test patients.</p>","PeriodicalId":19649,"journal":{"name":"Optometry and Vision Science","volume":"101 4","pages":"204-210"},"PeriodicalIF":1.6000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Wearable apparatus for correction of visual alignment under torsional strabismus.\",\"authors\":\"Charles Malleson, Jean-Yves Guillemaut\",\"doi\":\"10.1097/OPX.0000000000002122\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Significance: </strong>A wearable optical apparatus that compensates for eye misalignment (strabismus) to correct for double vision (diplopia) is proposed. In contrast to prism lenses, commonly used to compensate for horizontal and/or vertical misalignment, the proposed approach is able to compensate for any combination of horizontal, vertical, and torsional misalignment.</p><p><strong>Purpose: </strong>If the action of the extraocular muscles is compromised (e.g., by nerve damage), a patient may lose their ability to maintain visual alignment, negatively affecting their binocular fusion and stereo depth perception capability. Torsional misalignment cannot be mitigated by standard Fresnel prism lenses. Surgical procedures intended to correct torsional misalignment may be unpredictable. A wearable device able to rectify visual alignment and restore stereo depth perception without surgical intervention could potentially be of great value to people with strabismus.</p><p><strong>Methods: </strong>We propose a novel lightweight wearable optical device for visual alignment correction. The device comprises two mirrors and a Fresnel prism, arranged in such a way that together they rotationally shift the view seen by the affected eye horizontally, vertically, and torsionally. The extent of the alignment correction on each axis can be arbitrarily adjusted according to the patient's particular misalignment characteristics.</p><p><strong>Results: </strong>The proposed approach was tested by computer simulation, and a prototype device was manufactured. The prototype device was tested by a strabismus patient exhibiting horizontal and torsional misalignment. In these tests, the device was found to function as intended, allowing the patient to enjoy binocular fusion and stereo depth perception while wearing the device for daily activities over a period of several months.</p><p><strong>Conclusions: </strong>The proposed device is effective in correcting arbitrary horizontal, vertical, and torsional misalignment of the eyes. The results of the initial testing performed are highly encouraging. Future study is warranted to formally assess the effectiveness of the device on multiple test patients.</p>\",\"PeriodicalId\":19649,\"journal\":{\"name\":\"Optometry and Vision Science\",\"volume\":\"101 4\",\"pages\":\"204-210\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Optometry and Vision Science\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/OPX.0000000000002122\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Optometry and Vision Science","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/OPX.0000000000002122","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
Wearable apparatus for correction of visual alignment under torsional strabismus.
Significance: A wearable optical apparatus that compensates for eye misalignment (strabismus) to correct for double vision (diplopia) is proposed. In contrast to prism lenses, commonly used to compensate for horizontal and/or vertical misalignment, the proposed approach is able to compensate for any combination of horizontal, vertical, and torsional misalignment.
Purpose: If the action of the extraocular muscles is compromised (e.g., by nerve damage), a patient may lose their ability to maintain visual alignment, negatively affecting their binocular fusion and stereo depth perception capability. Torsional misalignment cannot be mitigated by standard Fresnel prism lenses. Surgical procedures intended to correct torsional misalignment may be unpredictable. A wearable device able to rectify visual alignment and restore stereo depth perception without surgical intervention could potentially be of great value to people with strabismus.
Methods: We propose a novel lightweight wearable optical device for visual alignment correction. The device comprises two mirrors and a Fresnel prism, arranged in such a way that together they rotationally shift the view seen by the affected eye horizontally, vertically, and torsionally. The extent of the alignment correction on each axis can be arbitrarily adjusted according to the patient's particular misalignment characteristics.
Results: The proposed approach was tested by computer simulation, and a prototype device was manufactured. The prototype device was tested by a strabismus patient exhibiting horizontal and torsional misalignment. In these tests, the device was found to function as intended, allowing the patient to enjoy binocular fusion and stereo depth perception while wearing the device for daily activities over a period of several months.
Conclusions: The proposed device is effective in correcting arbitrary horizontal, vertical, and torsional misalignment of the eyes. The results of the initial testing performed are highly encouraging. Future study is warranted to formally assess the effectiveness of the device on multiple test patients.
期刊介绍:
Optometry and Vision Science is the monthly peer-reviewed scientific publication of the American Academy of Optometry, publishing original research since 1924. Optometry and Vision Science is an internationally recognized source for education and information on current discoveries in optometry, physiological optics, vision science, and related fields. The journal considers original contributions that advance clinical practice, vision science, and public health. Authors should remember that the journal reaches readers worldwide and their submissions should be relevant and of interest to a broad audience. Topical priorities include, but are not limited to: clinical and laboratory research, evidence-based reviews, contact lenses, ocular growth and refractive error development, eye movements, visual function and perception, biology of the eye and ocular disease, epidemiology and public health, biomedical optics and instrumentation, novel and important clinical observations and treatments, and optometric education.