{"title":"[感染控制措施中消毒剂对医学影像和医疗设备影响的研究]。","authors":"Takahiro Araki, Toshiki Tateishi, Takayuki Kobayashi, Masaaki Taura, Natsue Matsuta, Nobuhiro Akagi, Tsutomu Kanazawa","doi":"10.6009/jjrt.2024-1425","DOIUrl":null,"url":null,"abstract":"<p><p>In this study, we investigated the effects of various disinfectants used to prevent infectious diseases on medical images and medical equipment. First, we investigated the effect of residual disinfectant on medical images in CT, mammography (MMG), and general imaging systems. Acrylic discs with various disinfectants attached were photographed using each imaging device, and visual evaluation and changes in image signal values were evaluated. We also conducted a questionnaire survey of each manufacturer regarding cleaning methods for medical devices. With CT/MMG, residual disinfectant could be visually confirmed on the image. Although this could not be confirmed with the general imaging system, a significant difference was confirmed in the image signal values of the general imaging system through statistical analysis. This is thought to be largely due to the influence of nonlinearity in the short-time imaging range of general imaging equipment. In addition, from the responses to a questionnaire survey of each medical device manufacturer, we were able to understand detailed cleaning methods that are not covered in medical device instruction manuals.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-06-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Research Study on the Effects of Disinfectants on Medical Images and Medical Equipment in Infection Control Measures].\",\"authors\":\"Takahiro Araki, Toshiki Tateishi, Takayuki Kobayashi, Masaaki Taura, Natsue Matsuta, Nobuhiro Akagi, Tsutomu Kanazawa\",\"doi\":\"10.6009/jjrt.2024-1425\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In this study, we investigated the effects of various disinfectants used to prevent infectious diseases on medical images and medical equipment. First, we investigated the effect of residual disinfectant on medical images in CT, mammography (MMG), and general imaging systems. Acrylic discs with various disinfectants attached were photographed using each imaging device, and visual evaluation and changes in image signal values were evaluated. We also conducted a questionnaire survey of each manufacturer regarding cleaning methods for medical devices. With CT/MMG, residual disinfectant could be visually confirmed on the image. Although this could not be confirmed with the general imaging system, a significant difference was confirmed in the image signal values of the general imaging system through statistical analysis. This is thought to be largely due to the influence of nonlinearity in the short-time imaging range of general imaging equipment. In addition, from the responses to a questionnaire survey of each medical device manufacturer, we were able to understand detailed cleaning methods that are not covered in medical device instruction manuals.</p>\",\"PeriodicalId\":74309,\"journal\":{\"name\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-06-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6009/jjrt.2024-1425\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/4/17 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1425","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/4/17 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
在这项研究中,我们调查了用于预防传染病的各种消毒剂对医学影像和医疗设备的影响。首先,我们调查了残留消毒剂对 CT、乳腺 X 射线(MMG)和普通成像系统中医学影像的影响。我们使用每种成像设备拍摄了附有各种消毒剂的丙烯酸盘,并对视觉评价和图像信号值的变化进行了评估。我们还就医疗设备的清洁方法对各制造商进行了问卷调查。通过 CT/MMG,可以在图像上直观地确认残留的消毒剂。虽然普通成像系统无法证实这一点,但通过统计分析证实,普通成像系统的图像信号值存在显著差异。这主要是由于普通成像设备短时间成像范围内非线性的影响。此外,通过对各医疗设备制造商的问卷调查,我们还了解到医疗设备使用手册中未涉及的详细清洁方法。
[Research Study on the Effects of Disinfectants on Medical Images and Medical Equipment in Infection Control Measures].
In this study, we investigated the effects of various disinfectants used to prevent infectious diseases on medical images and medical equipment. First, we investigated the effect of residual disinfectant on medical images in CT, mammography (MMG), and general imaging systems. Acrylic discs with various disinfectants attached were photographed using each imaging device, and visual evaluation and changes in image signal values were evaluated. We also conducted a questionnaire survey of each manufacturer regarding cleaning methods for medical devices. With CT/MMG, residual disinfectant could be visually confirmed on the image. Although this could not be confirmed with the general imaging system, a significant difference was confirmed in the image signal values of the general imaging system through statistical analysis. This is thought to be largely due to the influence of nonlinearity in the short-time imaging range of general imaging equipment. In addition, from the responses to a questionnaire survey of each medical device manufacturer, we were able to understand detailed cleaning methods that are not covered in medical device instruction manuals.