心内超声心动图能否减少半正常生理盐水灌注射频消融过程中的蒸汽爆裂?

IF 9.1 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS Circulation. Arrhythmia and electrophysiology Pub Date : 2024-06-01 Epub Date: 2024-05-28 DOI:10.1161/CIRCEP.123.012635
Kanae Hasegawa, Zachary T Yoneda, Jose R Martines-Parachini, Edward M Powers, Giovanni E Davogustto, Tiffany Y Hu, Kenichi Tokutake, Masaaki Kurata, Travis D Richardson, Jay A Montgomery, Sharon Shen, Juan C Estrada, Pablo J Saavedra, Amy Emerson, Marilyn L Walker, Harikrishna Tandri, Gregory F Michaud, Arvindh N Kanagasundram, William G Stevenson
{"title":"心内超声心动图能否减少半正常生理盐水灌注射频消融过程中的蒸汽爆裂?","authors":"Kanae Hasegawa, Zachary T Yoneda, Jose R Martines-Parachini, Edward M Powers, Giovanni E Davogustto, Tiffany Y Hu, Kenichi Tokutake, Masaaki Kurata, Travis D Richardson, Jay A Montgomery, Sharon Shen, Juan C Estrada, Pablo J Saavedra, Amy Emerson, Marilyn L Walker, Harikrishna Tandri, Gregory F Michaud, Arvindh N Kanagasundram, William G Stevenson","doi":"10.1161/CIRCEP.123.012635","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Irrigated radiofrequency ablation with half-normal saline can potentially increase lesion size but may increase the risk of steam pops with the risk of emboli or perforation. We hypothesized that pops would be preceded by intracardiac echocardiography (ICE) findings as well as a large impedance fall.</p><p><strong>Methods: </strong>In 100 consecutive patients undergoing endocardial ventricular arrhythmia radiofrequency ablation with half-normal saline, we attempted to observe the ablation site with ICE. Radiofrequency ablation power was titrated to a 15 to 20 Ohm impedance fall and could be adjusted for tissue whitening and increasing bubble formation on ICE. Steam pops were defined as audible or a sudden explosion of microbubbles on ICE.</p><p><strong>Results: </strong>Of 2190 ablation applications in 100 patients (82% cardiomyopathy, 50% sustained ventricular tachycardia), pops occurred during 43 (2.0%) applications. Sites with pops had greater impedance decreases of 18 [14, 21]% versus 13 [10, 17]% (<i>P</i><0.001). ICE visualized 1308 (59.7%) radiofrequency sites, and fewer pops occurred when ICE visualized the radiofrequency ablation site (1.4%) compared with without ICE visualization (2.8%; <i>P</i>=0.016). Of the 18 ICE-visible pops, 7 (39%) were silent but recognized as an explosion of bubbles on ICE. With ICE, 89% of pops were preceded by either tissue whitening or a sudden increase in bubbles. In a multivariable model, tissue whitening and a sudden increase in bubbles were associated with steam pops (odds ratio, 7.186; <i>P</i>=0.004, and odds ratio, 29.93; <i>P</i><0.001, respectively), independent of impedance fall and power. There were no pericardial effusions or embolic events with steam pops.</p><p><strong>Conclusions: </strong>Steam pops occurred in 2% of half-normal saline radiofrequency applications titrated to an impedance fall and are likely under-recognized without ICE. On ICE, steam pops are usually preceded by tissue whitening or a sudden increase in bubble formation, which can potentially be used to adjust radiofrequency application to help reduce pops.</p>","PeriodicalId":10319,"journal":{"name":"Circulation. Arrhythmia and electrophysiology","volume":" ","pages":"e012635"},"PeriodicalIF":9.1000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Can Intracardiac Echocardiography Reduce Steam Pops During Half-Normal Saline Irrigated Radiofrequency Ablation?\",\"authors\":\"Kanae Hasegawa, Zachary T Yoneda, Jose R Martines-Parachini, Edward M Powers, Giovanni E Davogustto, Tiffany Y Hu, Kenichi Tokutake, Masaaki Kurata, Travis D Richardson, Jay A Montgomery, Sharon Shen, Juan C Estrada, Pablo J Saavedra, Amy Emerson, Marilyn L Walker, Harikrishna Tandri, Gregory F Michaud, Arvindh N Kanagasundram, William G Stevenson\",\"doi\":\"10.1161/CIRCEP.123.012635\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Irrigated radiofrequency ablation with half-normal saline can potentially increase lesion size but may increase the risk of steam pops with the risk of emboli or perforation. We hypothesized that pops would be preceded by intracardiac echocardiography (ICE) findings as well as a large impedance fall.</p><p><strong>Methods: </strong>In 100 consecutive patients undergoing endocardial ventricular arrhythmia radiofrequency ablation with half-normal saline, we attempted to observe the ablation site with ICE. Radiofrequency ablation power was titrated to a 15 to 20 Ohm impedance fall and could be adjusted for tissue whitening and increasing bubble formation on ICE. Steam pops were defined as audible or a sudden explosion of microbubbles on ICE.</p><p><strong>Results: </strong>Of 2190 ablation applications in 100 patients (82% cardiomyopathy, 50% sustained ventricular tachycardia), pops occurred during 43 (2.0%) applications. Sites with pops had greater impedance decreases of 18 [14, 21]% versus 13 [10, 17]% (<i>P</i><0.001). ICE visualized 1308 (59.7%) radiofrequency sites, and fewer pops occurred when ICE visualized the radiofrequency ablation site (1.4%) compared with without ICE visualization (2.8%; <i>P</i>=0.016). Of the 18 ICE-visible pops, 7 (39%) were silent but recognized as an explosion of bubbles on ICE. With ICE, 89% of pops were preceded by either tissue whitening or a sudden increase in bubbles. In a multivariable model, tissue whitening and a sudden increase in bubbles were associated with steam pops (odds ratio, 7.186; <i>P</i>=0.004, and odds ratio, 29.93; <i>P</i><0.001, respectively), independent of impedance fall and power. There were no pericardial effusions or embolic events with steam pops.</p><p><strong>Conclusions: </strong>Steam pops occurred in 2% of half-normal saline radiofrequency applications titrated to an impedance fall and are likely under-recognized without ICE. On ICE, steam pops are usually preceded by tissue whitening or a sudden increase in bubble formation, which can potentially be used to adjust radiofrequency application to help reduce pops.</p>\",\"PeriodicalId\":10319,\"journal\":{\"name\":\"Circulation. Arrhythmia and electrophysiology\",\"volume\":\" \",\"pages\":\"e012635\"},\"PeriodicalIF\":9.1000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Circulation. Arrhythmia and electrophysiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1161/CIRCEP.123.012635\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/5/28 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation. Arrhythmia and electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1161/CIRCEP.123.012635","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/28 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

背景:使用半生理盐水冲洗射频消融术有可能增大病灶面积,但可能会增加蒸汽爆裂的风险以及栓塞或穿孔的风险。我们假设,爆裂之前会有心内超声心动图(ICE)检查结果以及阻抗大幅下降:方法:在 100 例连续接受心内膜室性心律失常射频消融术的患者中,我们尝试用 ICE 观察消融部位。射频消融功率被调节到 15 到 20 欧姆的阻抗值,并可根据组织变白和 ICE 上气泡形成增加的情况进行调整。蒸汽爆裂是指在 ICE 上可听到或突然爆发的微气泡:在 100 名患者(82% 患有心肌病,50% 患有持续性室性心动过速)的 2190 次消融治疗中,有 43 次(2.0%)发生了爆裂。出现爆裂的部位阻抗下降幅度更大,分别为 18 [14, 21]% 和 13 [10, 17]% (PP=0.016)。在 18 个 ICE 可见的爆裂声中,有 7 个(39%)是无声的,但在 ICE 上可识别为气泡爆炸。在使用 ICE 时,89% 的爆裂前会出现组织变白或气泡突然增加。在多变量模型中,组织变白和气泡突然增加与蒸汽爆裂有关(几率比为 7.186;P=0.004,几率比为 29.93;PC 结论):在阻抗下降的半正常生理盐水射频应用中,有 2% 出现了蒸汽爆裂,如果没有 ICE,这种情况很可能没有被充分认识到。在使用 ICE 时,蒸汽爆裂之前通常会出现组织变白或气泡形成突然增加的现象,这有可能用来调整射频应用以帮助减少爆裂。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Can Intracardiac Echocardiography Reduce Steam Pops During Half-Normal Saline Irrigated Radiofrequency Ablation?

Background: Irrigated radiofrequency ablation with half-normal saline can potentially increase lesion size but may increase the risk of steam pops with the risk of emboli or perforation. We hypothesized that pops would be preceded by intracardiac echocardiography (ICE) findings as well as a large impedance fall.

Methods: In 100 consecutive patients undergoing endocardial ventricular arrhythmia radiofrequency ablation with half-normal saline, we attempted to observe the ablation site with ICE. Radiofrequency ablation power was titrated to a 15 to 20 Ohm impedance fall and could be adjusted for tissue whitening and increasing bubble formation on ICE. Steam pops were defined as audible or a sudden explosion of microbubbles on ICE.

Results: Of 2190 ablation applications in 100 patients (82% cardiomyopathy, 50% sustained ventricular tachycardia), pops occurred during 43 (2.0%) applications. Sites with pops had greater impedance decreases of 18 [14, 21]% versus 13 [10, 17]% (P<0.001). ICE visualized 1308 (59.7%) radiofrequency sites, and fewer pops occurred when ICE visualized the radiofrequency ablation site (1.4%) compared with without ICE visualization (2.8%; P=0.016). Of the 18 ICE-visible pops, 7 (39%) were silent but recognized as an explosion of bubbles on ICE. With ICE, 89% of pops were preceded by either tissue whitening or a sudden increase in bubbles. In a multivariable model, tissue whitening and a sudden increase in bubbles were associated with steam pops (odds ratio, 7.186; P=0.004, and odds ratio, 29.93; P<0.001, respectively), independent of impedance fall and power. There were no pericardial effusions or embolic events with steam pops.

Conclusions: Steam pops occurred in 2% of half-normal saline radiofrequency applications titrated to an impedance fall and are likely under-recognized without ICE. On ICE, steam pops are usually preceded by tissue whitening or a sudden increase in bubble formation, which can potentially be used to adjust radiofrequency application to help reduce pops.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
13.70
自引率
4.80%
发文量
187
审稿时长
4-8 weeks
期刊介绍: Circulation: Arrhythmia and Electrophysiology is a journal dedicated to the study and application of clinical cardiac electrophysiology. It covers a wide range of topics including the diagnosis and treatment of cardiac arrhythmias, as well as research in this field. The journal accepts various types of studies, including observational research, clinical trials, epidemiological studies, and advancements in translational research.
期刊最新文献
Exercise Training Improves Cognitive Function and Neurovascular Control: A Secondary Analysis of the Exercise Training in Heart Failure With Reduced Ejection Fraction and Permanent Atrial Fibrillation: A Randomized Clinical Trial. Resuscitated Sudden Cardiac Arrest as the Initial Presentation of Hypertrophic Cardiomyopathy. Impact of Being Underweight on the Safety and Efficacy of Atrial Fibrillation Ablation: An Analysis From the Japanese Catheter Ablation Registry. Multicenter Hemodynamic Assessment of the LOT-CRT Strategy: When Does Combining Left Bundle Branch Pacing and Coronary Venous Pacing Enhance Resynchronization?: Primary Results of the CSPOT Study. Performance of Atrial Fibrillation Burden Trends for Stroke Risk Stratification.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1