{"title":"跖骨头下足底皮肤对生活压力刺激的动态微循环特征","authors":"Zhenming Zhang, Wen-Ming Chen, Xiong-Gang Yang, Xingyu Zhang, Xu Wang, Jiazhang Huang, Chao Zhang, Xiang Geng, Xin Ma","doi":"10.1111/micc.12860","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objective</h3>\n \n <p>Diabetic foot ulcer (DFU) is a severe complication with high mortality. High plantar pressure and poor microcirculation are considered main causes of DFU. The specific aims were to provide a novel technique for real-time measurement of plantar skin blood flow (SBF) under walking-like pressure stimulus and delineate the first plantar metatarsal head dynamic microcirculation characteristics because of life-like loading conditions in healthy individuals.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Twenty young healthy participants (14 male and 6 female) were recruited. The baseline (i.e., unloaded) SBF of soft tissue under the first metatarsal head were measured using laser Doppler flowmetry (LDF). A custom-made machine was utilized to replicate daily walking pressure exertion for 5 min. The exerted plantar force was adjusted from 10 N (127.3 kPa) to 40 N (509.3 kPa) at an increase of 5 N (63.7 kPa). Real-time SBF was acquired using the LDF. After each pressure exertion, postload SBF was measured for comparative purposes. Statistical analysis was performed using the R software.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>All levels of immediate-load and postload SBF increased significantly compared with baseline values. As the exerted load increased, the postload and immediate-load SBF tended to increase until the exerted load reached 35 N (445.6 kPa). However, in immediate-load data, the increasing trend tended to level off as the exerted pressure increased from 15 N (191.0 kPa) to 25 N (318.3 kPa). For postload and immediate-load SBF, they both peaked at 35 N (445.6 kPa). However, when the exerted force exceeds 35 N (445.6 kPa), both the immediate-load and postload SBF values started to decrease.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Our study offered a novel real-time plantar soft tissue microcirculation measurement technique under dynamic conditions. For the first metatarsal head of healthy people, 20 N (254.6 kPa)-plantar pressure has a fair microcirculation stimulus compared with higher pressure. There might be a pressure threshold at 35 N (445.6 kPa) for the first metatarsal head, and soft tissue microcirculation may decrease when local pressure exceeds it.</p>\n </section>\n </div>","PeriodicalId":18459,"journal":{"name":"Microcirculation","volume":"31 5","pages":""},"PeriodicalIF":1.9000,"publicationDate":"2024-06-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Dynamic Microcirculation Characteristics of Plantar Skin Under Metatarsal Head of Human Foot in Response to Life-Like Pressure Stimulus\",\"authors\":\"Zhenming Zhang, Wen-Ming Chen, Xiong-Gang Yang, Xingyu Zhang, Xu Wang, Jiazhang Huang, Chao Zhang, Xiang Geng, Xin Ma\",\"doi\":\"10.1111/micc.12860\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Objective</h3>\\n \\n <p>Diabetic foot ulcer (DFU) is a severe complication with high mortality. High plantar pressure and poor microcirculation are considered main causes of DFU. The specific aims were to provide a novel technique for real-time measurement of plantar skin blood flow (SBF) under walking-like pressure stimulus and delineate the first plantar metatarsal head dynamic microcirculation characteristics because of life-like loading conditions in healthy individuals.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>Twenty young healthy participants (14 male and 6 female) were recruited. The baseline (i.e., unloaded) SBF of soft tissue under the first metatarsal head were measured using laser Doppler flowmetry (LDF). A custom-made machine was utilized to replicate daily walking pressure exertion for 5 min. The exerted plantar force was adjusted from 10 N (127.3 kPa) to 40 N (509.3 kPa) at an increase of 5 N (63.7 kPa). Real-time SBF was acquired using the LDF. After each pressure exertion, postload SBF was measured for comparative purposes. Statistical analysis was performed using the R software.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>All levels of immediate-load and postload SBF increased significantly compared with baseline values. As the exerted load increased, the postload and immediate-load SBF tended to increase until the exerted load reached 35 N (445.6 kPa). However, in immediate-load data, the increasing trend tended to level off as the exerted pressure increased from 15 N (191.0 kPa) to 25 N (318.3 kPa). For postload and immediate-load SBF, they both peaked at 35 N (445.6 kPa). However, when the exerted force exceeds 35 N (445.6 kPa), both the immediate-load and postload SBF values started to decrease.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>Our study offered a novel real-time plantar soft tissue microcirculation measurement technique under dynamic conditions. For the first metatarsal head of healthy people, 20 N (254.6 kPa)-plantar pressure has a fair microcirculation stimulus compared with higher pressure. There might be a pressure threshold at 35 N (445.6 kPa) for the first metatarsal head, and soft tissue microcirculation may decrease when local pressure exceeds it.</p>\\n </section>\\n </div>\",\"PeriodicalId\":18459,\"journal\":{\"name\":\"Microcirculation\",\"volume\":\"31 5\",\"pages\":\"\"},\"PeriodicalIF\":1.9000,\"publicationDate\":\"2024-06-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Microcirculation\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/micc.12860\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"HEMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Microcirculation","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/micc.12860","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"HEMATOLOGY","Score":null,"Total":0}
Dynamic Microcirculation Characteristics of Plantar Skin Under Metatarsal Head of Human Foot in Response to Life-Like Pressure Stimulus
Objective
Diabetic foot ulcer (DFU) is a severe complication with high mortality. High plantar pressure and poor microcirculation are considered main causes of DFU. The specific aims were to provide a novel technique for real-time measurement of plantar skin blood flow (SBF) under walking-like pressure stimulus and delineate the first plantar metatarsal head dynamic microcirculation characteristics because of life-like loading conditions in healthy individuals.
Methods
Twenty young healthy participants (14 male and 6 female) were recruited. The baseline (i.e., unloaded) SBF of soft tissue under the first metatarsal head were measured using laser Doppler flowmetry (LDF). A custom-made machine was utilized to replicate daily walking pressure exertion for 5 min. The exerted plantar force was adjusted from 10 N (127.3 kPa) to 40 N (509.3 kPa) at an increase of 5 N (63.7 kPa). Real-time SBF was acquired using the LDF. After each pressure exertion, postload SBF was measured for comparative purposes. Statistical analysis was performed using the R software.
Results
All levels of immediate-load and postload SBF increased significantly compared with baseline values. As the exerted load increased, the postload and immediate-load SBF tended to increase until the exerted load reached 35 N (445.6 kPa). However, in immediate-load data, the increasing trend tended to level off as the exerted pressure increased from 15 N (191.0 kPa) to 25 N (318.3 kPa). For postload and immediate-load SBF, they both peaked at 35 N (445.6 kPa). However, when the exerted force exceeds 35 N (445.6 kPa), both the immediate-load and postload SBF values started to decrease.
Conclusions
Our study offered a novel real-time plantar soft tissue microcirculation measurement technique under dynamic conditions. For the first metatarsal head of healthy people, 20 N (254.6 kPa)-plantar pressure has a fair microcirculation stimulus compared with higher pressure. There might be a pressure threshold at 35 N (445.6 kPa) for the first metatarsal head, and soft tissue microcirculation may decrease when local pressure exceeds it.
期刊介绍:
The journal features original contributions that are the result of investigations contributing significant new information relating to the vascular and lymphatic microcirculation addressed at the intact animal, organ, cellular, or molecular level. Papers describe applications of the methods of physiology, biophysics, bioengineering, genetics, cell biology, biochemistry, and molecular biology to problems in microcirculation.
Microcirculation also publishes state-of-the-art reviews that address frontier areas or new advances in technology in the fields of microcirculatory disease and function. Specific areas of interest include: Angiogenesis, growth and remodeling; Transport and exchange of gasses and solutes; Rheology and biorheology; Endothelial cell biology and metabolism; Interactions between endothelium, smooth muscle, parenchymal cells, leukocytes and platelets; Regulation of vasomotor tone; and Microvascular structures, imaging and morphometry. Papers also describe innovations in experimental techniques and instrumentation for studying all aspects of microcirculatory structure and function.